High performance liquid chromatography (HPLC) is liquid chromatography carried out on columns packed with very small particles, with the mobile phase driven through by a high-pressure pump. Small particles give sharp, well-separated peaks; the pressure is what it takes to move liquid through them at a useful speed.
HPLC is the most widely used technique for measuring non-volatile and heat-sensitive compounds: pharmaceuticals, food ingredients and contaminants, environmental pollutants, natural products, peptides and proteins.
The parts of an HPLC system
- Solvent reservoirs and degasser. Hold the mobile phase and remove dissolved gas.
- Pump. Delivers a constant, pulse-free flow, typically 0.1 to 2 mL/min on an analytical system, and mixes solvents for gradients.
- Injector or autosampler. Introduces a measured volume of sample, usually 1 to 50 µL, from a vial.
- Column, usually in a column oven. Where the separation happens. A guard column may sit in front of it.
- Detector. UV or diode array for compounds that absorb light; fluorescence, refractive index, evaporative light scattering, conductivity or a mass spectrometer for others.
- Data system. Records the chromatogram and integrates the peaks.
- Reservoirs and degasserHold the mobile phase and remove dissolved gas
- PumpConstant, pulse-free flow at high pressure
- InjectorIntroduces a measured volume of sample
- ColumnSeparation, usually in a column oven
- DetectorA signal for each compound
- Data systemRecords and integrates the chromatogram
Separation modes
| Mode | Stationary phase | Mobile phase | Used for |
|---|---|---|---|
| Reversed phase | Non-polar, for example C18 or C8 on silica | Water with acetonitrile or methanol | Most small molecules; the default mode |
| Normal phase | Polar, for example bare silica, cyano, amino, diol | Non-polar solvent with a polar modifier | Isomers, fat-soluble compounds |
| HILIC | Polar | Mostly acetonitrile with some water or buffer | Very polar compounds that reversed phase does not retain |
| Ion exchange | Charged groups | Aqueous buffer, salt or pH gradient | Ions, proteins, nucleic acids |
| Size exclusion | Inert porous particles | Aqueous buffer or organic solvent | Proteins, polymers by size |
| Chiral | A chiral selector | Depends on the phase | Enantiomers |
Reversed phase on a C18 column accounts for the large majority of methods. See how to select an HPLC column.
Isocratic and gradient elution
In an isocratic run the mobile phase composition stays constant. It is simple and needs no re-equilibration between injections, and suits samples whose components have similar retention.
In a gradient run the share of the strong solvent rises during the run. It elutes compounds of very different polarity in one run with sharp peaks throughout, at the cost of a re-equilibration step before the next injection.
HPLC and UHPLC
Conventional HPLC uses 3 to 5 µm particles at pressures up to about 400 bar. UHPLC uses particles below 2 µm in shorter, narrower columns at up to about 1,000 to 1,500 bar, depending on the instrument. The gain is speed and sharper peaks; the price is a system with very low internal volume and cleaner samples and solvents. Columns with solid-core particles of about 2.6 to 2.7 µm sit between the two. More in fast chromatography.
What makes a result reliable
- A column suited to the analytes and kept in good condition.
- A properly prepared, filtered and degassed mobile phase. See mobile phase preparation.
- Samples free of particles. See sample filtration for HPLC and UHPLC.
- Calibration against suitable reference standards.
- A system suitability check before the samples are run.