Mobile phases in HPLC: preparation, buffers, filtration and degassing

How to prepare an HPLC mobile phase: measuring and mixing, choosing a buffer by pKa, adjusting pH, volatile buffers for LC-MS, filtration and degassing.

  • 3 min read
  • 1 figure
  • Reviewed 25 November 2026
On this page 7 sections
  1. Measure, then mix
  2. Choosing a buffer
  3. Adjusting the pH
  4. Buffer solubility
  5. Filtration
  6. Degassing
  7. Shelf life and housekeeping

Measure, then mix

Measure each component separately and then combine them. Water and organic solvents shrink in volume when mixed: 500 mL of methanol plus 500 mL of water gives noticeably less than 1,000 mL. Topping a measuring cylinder up to the mark with the second solvent therefore gives a different, and less reproducible, composition than the method intends.

Mixing also changes the temperature (methanol and water warm up; acetonitrile and water cool down). Let the mixture return to room temperature before use, because retention and dissolved gas both depend on it.

Choosing a buffer

A buffer is needed whenever an analyte can gain or lose a charge, because retention then depends on pH. A buffer holds the pH well only within about one unit either side of its pKa.

BufferpKa (25 °C)Useful pH rangeSuited to mass spectrometry
Phosphate2.1 / 7.2 / 12.31.1 to 3.1 and 6.2 to 8.2No (not volatile)
Formate3.82.8 to 4.8Yes
Acetate4.83.8 to 5.8Yes
Citrate3.1 / 4.8 / 6.42.1 to 7.4No
Ammonium (ammonia)9.28.2 to 10.2Yes
  • A concentration of 10 to 50 mmol/L is usual. Too little gives poor pH control and peak shape; too much risks precipitation and wears pump seals.
  • For LC-MS use volatile additives only: formic or acetic acid, ammonium formate or ammonium acetate, kept dilute.
  • For low-wavelength UV phosphate is the most transparent common buffer. Acetate, formate, citrate and trifluoroacetic acid absorb below roughly 210 to 230 nm.
  • Keep the mobile phase pH inside the range the column is rated for.
Useful pH ranges of common HPLC buffers1234567891011Phosphate (pKa 2.1)1.1 to 3.1Formate2.8 to 4.8Acetate3.8 to 5.8Citrate2.1 to 7.4Phosphate (pKa 7.2)6.2 to 8.2Ammonium (ammonia)8.2 to 10.2pH
Each buffer controls pH within about one unit either side of its pKa. Choose one whose range covers the pH of the method.

Adjusting the pH

Adjust and measure the pH of the aqueous part before adding the organic solvent. A pH electrode calibrated in aqueous buffers does not read true in a mixed solvent, and the method will only transfer between laboratories if everyone measures the same way. Calibrate the pH meter with fresh buffers that bracket the target value. See pH measurement.

Buffer solubility

Buffer salts are far less soluble in organic solvent than in water. Phosphate buffers can precipitate somewhere above roughly 70 to 80% organic solvent, earlier in acetonitrile than in methanol and earlier at higher concentration. Before running a gradient to high organic, mix the buffer with the strongest composition in a beaker and check that it stays clear.

Filtration

Filter mobile phases that contain dissolved solids through a membrane of 0.45 µm, or 0.2 µm for UHPLC and columns with particles below 2 µm. Choose a membrane compatible with the solvent: regenerated cellulose or nylon for aqueous and mixed phases, PTFE for organic solvents. Use bottle-top or vacuum filtration with clean glassware. High-purity solvents used straight from the bottle do not normally need filtering, and an unnecessary filtration step can add contamination.

Degassing

Dissolved air comes out of solution when solvents are mixed or the pressure drops, causing bubbles, unstable flow and baseline noise. An inline vacuum degasser is the standard remedy. Helium sparging is very effective but can change the composition of volatile mixtures over time. Sonication alone removes little gas. More in preventing air bubbles in the HPLC system.

Shelf life and housekeeping

  • Make aqueous buffers fresh and replace them within a few days. Near-neutral phosphate and acetate grow microorganisms quickly.
  • Never top up a reservoir. Rinse it and refill.
  • Label each bottle with contents, pH, date and initials.
  • At the end of the day flush buffer out of the system and the column with water-rich mobile phase without the salt before switching to high organic. Buffer left to dry out crystallises and damages seals, check valves and the column.

General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.