Solvents for HPLC: grades, UV cut-off and handling

HPLC solvents explained: HPLC, gradient and LC-MS grade, UV cut-off of acetonitrile, methanol and others, viscosity and pressure, water quality and storage.

  • 2 min read
  • 1 figure
  • Reviewed 25 November 2026
On this page 5 sections
  1. Why the grade matters
  2. UV cut-off
  3. Viscosity and back pressure
  4. Water
  5. Handling and storage

Why the grade matters

An HPLC detector sees whatever is in the mobile phase. Impurities in a solvent raise the baseline, produce ghost peaks in gradients, shorten column life and, with mass spectrometry, suppress the signal or form adducts. Solvent grades are defined by what they are tested for.

GradeTested forUse it for
HPLC (isocratic) gradeUV transmission, low particles and residueIsocratic methods with UV detection
Gradient gradeAs above, plus a gradient run showing no interfering peaksGradient methods; low-wavelength UV
LC-MS gradeAs above, plus low metal ions and ionisable impurities, checked by mass spectrometryAny work with a mass spectrometer

A solvent of a lower grade can be fine for extraction or cleaning but should not go on the column.

UV cut-off

The UV cut-off is the wavelength below which the solvent itself absorbs too strongly to work with (an absorbance of 1 in a 1 cm cell). The detection wavelength should sit well above it. Typical values for high-purity solvents:

SolventApproximate UV cut-off
Waterbelow 190 nm
Acetonitrile190 nm
n-Hexane195 nm
Methanol205 nm
2-Propanol205 nm
Ethanol210 nm
Tetrahydrofuran (unstabilised)212 nm
Dichloromethane233 nm
Ethyl acetate256 nm
Acetone330 nm

Values vary a little by supplier and grade: the certificate for the bottle in your hand is the figure to rely on. Acetonitrile's low cut-off is one reason it is preferred for detection near 200 to 220 nm.

Approximate UV cut-off of common HPLC solvents200220240260280300320340Acetonitrile190n-Hexane195Methanol2052-Propanol205Ethanol210Tetrahydrofuran212Dichloromethane233Ethyl acetate256Acetone330Wavelength, nmThe solvent absorbs too strongly below the end of its bar
Approximate UV cut-off of common solvents. Detection has to work at a wavelength well above the cut-off of the mobile phase.

Viscosity and back pressure

Back pressure rises with the viscosity of the mobile phase. At room temperature acetonitrile (about 0.37 mPa·s) is thinner than methanol (about 0.6 mPa·s) and water (about 1.0 mPa·s).

Mixtures do not behave like the average of their parts. Methanol and water mixed in roughly equal parts are considerably more viscous than either alone, so the pressure of a methanol gradient peaks mid-run. Acetonitrile and water mixtures stay close to the viscosity of water or below. Warming the column lowers viscosity and pressure.

Water

Water is the largest part of most reversed-phase mobile phases and the most common source of contamination. Use freshly drawn ultrapure water (resistivity 18.2 megohm·cm at 25 °C, low total organic carbon) or bottled water of the appropriate grade. Water stored in an open container picks up organic contaminants and grows microorganisms within days.

Handling and storage

  • Keep bottles closed, away from light and heat. Use solvent bottle caps with filtered vents on the instrument.
  • Do not top up a reservoir: empty, rinse and refill it, so contamination does not build up.
  • Replace aqueous mobile phases and buffers frequently. Algae and bacteria block frits and filters.
  • Tetrahydrofuran and other ethers form peroxides. Buy small quantities, note the opening date and do not use old stock.
  • Stabilisers matter: butylated hydroxytoluene in stabilised tetrahydrofuran absorbs in the UV; ethanol or amylene in chlorinated solvents changes normal-phase retention.
  • Check that the solvent suits the system. Some polymer tubing and seals do not tolerate tetrahydrofuran or chlorinated solvents. See chemical resistance of plastics.

Preparing the mixture correctly is covered in mobile phase preparation.

General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.