Why the grade matters
An HPLC detector sees whatever is in the mobile phase. Impurities in a solvent raise the baseline, produce ghost peaks in gradients, shorten column life and, with mass spectrometry, suppress the signal or form adducts. Solvent grades are defined by what they are tested for.
| Grade | Tested for | Use it for |
|---|---|---|
| HPLC (isocratic) grade | UV transmission, low particles and residue | Isocratic methods with UV detection |
| Gradient grade | As above, plus a gradient run showing no interfering peaks | Gradient methods; low-wavelength UV |
| LC-MS grade | As above, plus low metal ions and ionisable impurities, checked by mass spectrometry | Any work with a mass spectrometer |
A solvent of a lower grade can be fine for extraction or cleaning but should not go on the column.
UV cut-off
The UV cut-off is the wavelength below which the solvent itself absorbs too strongly to work with (an absorbance of 1 in a 1 cm cell). The detection wavelength should sit well above it. Typical values for high-purity solvents:
| Solvent | Approximate UV cut-off |
|---|---|
| Water | below 190 nm |
| Acetonitrile | 190 nm |
| n-Hexane | 195 nm |
| Methanol | 205 nm |
| 2-Propanol | 205 nm |
| Ethanol | 210 nm |
| Tetrahydrofuran (unstabilised) | 212 nm |
| Dichloromethane | 233 nm |
| Ethyl acetate | 256 nm |
| Acetone | 330 nm |
Values vary a little by supplier and grade: the certificate for the bottle in your hand is the figure to rely on. Acetonitrile's low cut-off is one reason it is preferred for detection near 200 to 220 nm.
Viscosity and back pressure
Back pressure rises with the viscosity of the mobile phase. At room temperature acetonitrile (about 0.37 mPa·s) is thinner than methanol (about 0.6 mPa·s) and water (about 1.0 mPa·s).
Mixtures do not behave like the average of their parts. Methanol and water mixed in roughly equal parts are considerably more viscous than either alone, so the pressure of a methanol gradient peaks mid-run. Acetonitrile and water mixtures stay close to the viscosity of water or below. Warming the column lowers viscosity and pressure.
Water
Water is the largest part of most reversed-phase mobile phases and the most common source of contamination. Use freshly drawn ultrapure water (resistivity 18.2 megohm·cm at 25 °C, low total organic carbon) or bottled water of the appropriate grade. Water stored in an open container picks up organic contaminants and grows microorganisms within days.
Handling and storage
- Keep bottles closed, away from light and heat. Use solvent bottle caps with filtered vents on the instrument.
- Do not top up a reservoir: empty, rinse and refill it, so contamination does not build up.
- Replace aqueous mobile phases and buffers frequently. Algae and bacteria block frits and filters.
- Tetrahydrofuran and other ethers form peroxides. Buy small quantities, note the opening date and do not use old stock.
- Stabilisers matter: butylated hydroxytoluene in stabilised tetrahydrofuran absorbs in the UV; ethanol or amylene in chlorinated solvents changes normal-phase retention.
- Check that the solvent suits the system. Some polymer tubing and seals do not tolerate tetrahydrofuran or chlorinated solvents. See chemical resistance of plastics.
Preparing the mixture correctly is covered in mobile phase preparation.