The harmonised chapter
The general chromatography chapters of the United States, European and Japanese pharmacopoeias have been harmonised. The harmonised text became official in USP general chapter <621> on 1 December 2022 and in Ph. Eur. chapter 2.2.46 on 1 January 2023. It sets out how far the conditions of a monograph method may be adjusted without the change counting as a modification that needs revalidation.
This page is a summary for orientation. The chapter in force, and anything the individual monograph says, always take precedence.
The principle
- Adjustments within the stated limits are permitted without revalidation. Changes beyond them need revalidation.
- The system suitability criteria of the monograph must still be met after the change.
- The chemistry of the stationary phase must not change: a C18 may not be replaced by a C8.
- When several adjustments are combined, their effect together should be assessed, and a verification of the adjusted method may be needed.
Isocratic elution
| Parameter | Permitted adjustment |
|---|---|
| Column length (L) and particle size (dp) | May be changed if the ratio L/dp stays the same or within -25% to +50% of the prescribed ratio |
| Totally porous to superficially porous particles | Other combinations of L and dp are allowed if the plate number stays within -25% to +50% of the prescribed column |
| Inner diameter | May be adjusted |
| Flow rate | Recalculated when the column dimensions change; in addition, plus or minus 50% |
| Column temperature | Plus or minus 10 °C where the operating temperature is specified |
| Mobile phase composition | A minor component by plus or minus 30% relative, but no component by more than 10% absolute |
| pH of the aqueous component | Plus or minus 0.2 pH units |
| Salt concentration of a buffer | Plus or minus 10% |
| Detector wavelength | No adjustment |
| Injection volume | May be scaled with the column dimensions. It may also be varied without a column change if system suitability is met: when it is reduced, check detection and repeatability; when it is increased, check linearity and resolution |
Gradient elution
Gradient methods are more sensitive to change, so the limits are tighter.
- Column length and particle size: may be changed if L/dp stays the same or within -25% to +50%, with the gradient times recalculated for the new column.
- Column temperature: plus or minus 5 °C.
- Mobile phase and gradient: adjustments are acceptable provided system suitability is met, the principal peaks elute within plus or minus 15% of the retention times obtained with the original conditions (this check does not apply when the column dimensions are changed), the first peaks are still sufficiently retained and the last peaks are eluted.
- Detector wavelength: no adjustment.
The calculations
When the column changes, flow rate, gradient time and injection volume are scaled so the separation stays equivalent. With dc the inner diameter, dp the particle size, L the length and F the flow rate (1 for the prescribed column, 2 for the new one):
- Flow rate: F2 = F1 x (dc2² x dp1) / (dc1² x dp2)
- Gradient segment time: tG2 = tG1 x (F1 / F2) x (L2 x dc2²) / (L1 x dc1²)
- Injection volume: V2 = V1 x (L2 x dc2²) / (L1 x dc1²)
Worked example
A monograph prescribes 250 x 4.6 mm, 5 µm (L/dp = 50,000). A 150 x 4.6 mm, 3 µm column has L/dp = 50,000: the same ratio, so the change is permitted. A 100 x 2.1 mm, 1.8 µm column has L/dp = 55,556, which is 11% higher and within the range. A 50 x 2.1 mm, 1.8 µm column has L/dp = 27,778, which is 44% lower and outside it.
What this means when you buy a column
Choose the same L class or Ph. Eur. description as the monograph (USP L listing), then pick dimensions inside the L/dp window. The method can move to a shorter column with smaller particles, saving time and solvent, and remain within the pharmacopoeia, provided system suitability passes on your instrument.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.