The idea
A separation needs a certain number of theoretical plates. Plates can come from a long column of large particles or a short column of small particles. The short column delivers them in a fraction of the time and solvent. Fast chromatography is the practice of moving a method to a shorter column with smaller particles while keeping the separation the same.
Keep L/dp constant
The plate number a column can deliver is roughly proportional to its length divided by its particle size (L/dp). Columns with a similar ratio have similar separating power:
| Column | L/dp | Run time at the same linear velocity, relative to the 250 mm column |
|---|---|---|
| 250 mm, 5 µm | 50,000 | 100% |
| 150 mm, 3 µm | 50,000 | 60% |
| 100 mm, 2 µm | 50,000 | 40% |
| 150 mm, 5 µm | 30,000 | 60% |
| 100 mm, 3 µm | 33,333 | 40% |
| 50 mm, 1.7 µm | 29,412 | 20% |
Smaller particles also stay efficient at higher flow, so the flow can often be raised as well, which shortens the run further.
The cost: pressure
At the same length and linear velocity, pressure rises with the inverse square of the particle size. Going from 5 µm to 1.7 µm multiplies the pressure by nearly nine. Shortening the column gives some of that back, but sub-2 µm columns still need a UHPLC system.
Solid-core (superficially porous) particles of about 2.6 to 2.7 µm are the common compromise. Their short diffusion path gives efficiency close to sub-2 µm fully porous particles at roughly half the pressure, within reach of many conventional instruments.
Raising the column temperature lowers the viscosity of the mobile phase and so the pressure. See temperature and retention time.
Scaling the method
When the column changes, scale the other settings so the chromatography stays equivalent (dc = inner diameter, L = length):
- Flow rate for the same linear velocity: F2 = F1 x (dc2 / dc1)². From 4.6 mm to 2.1 mm, 1.0 mL/min becomes 0.21 mL/min, a saving of about 79% in solvent.
- Injection volume in proportion to column volume: V2 = V1 x (L2 x dc2²) / (L1 x dc1²).
- Gradient times so that each step delivers the same number of column volumes: tG2 = tG1 x (F1 / F2) x (L2 x dc2²) / (L1 x dc1²).
For pharmacopoeial methods the permitted window is set out in USP 621 and Ph. Eur. 2.2.46.
The instrument has to keep up
Narrow peaks from small columns are easily broadened outside the column.
- Extra-column volume: use short capillaries of narrow bore, zero-dead-volume fittings made up correctly, and a low-volume detector cell. See preventing dead volume.
- Detector sampling rate: collect at least about 20 data points across the narrowest peak, and shorten the response time to match.
- Dwell volume: a large gradient delay volume distorts fast gradients on narrow columns. Know your system's value when transferring a method.
- Sample solvent and volume: small columns tolerate less strong solvent and less volume.
Cleaner samples and solvents
Small particles mean frits with small pores, which block easily. Filter samples through 0.2 µm for sub-2 µm columns (sample filtration), filter buffered mobile phases, and use a guard column or inline filter of matching low volume (inline filters).
Is it worth it?
Productivity is samples per hour, including re-equilibration. A method moved from 250 x 4.6 mm, 5 µm to 100 x 2.1 mm, sub-2 µm can run several times faster on about a tenth of the solvent. The gain is real only if the instrument's extra-column volume and pressure limit suit the new column; otherwise a 3 µm or solid-core column on the existing system is the better step.