Why filter your sample?

The inlet frit of an HPLC column has pores of about 2 µm, and about 0.5 µm or less on columns packed with very small particles. Anything larger collects there. The result is rising back pressure, split or tailing peaks and a shortened column life. Particles also wear injector seals and block capillaries and needles. Filtering each sample costs little; a blocked column costs a great deal more.

Filter the mobile phase too, whenever it contains dissolved salts or additives (mobile phase preparation).

Choosing the right pore size

Column packingFilter pore size
Particles of 3 µm and larger0.45 µm
Particles below 3 µm; UHPLC; LC-MS0.2 µm (also sold as 0.22 µm)
Samples heavy with particlesA glass-fibre prefilter first, then 0.45 or 0.2 µm

A finer filter than necessary blocks sooner and needs more force, so match the pore size to the column.

Filter pore sizes compared with the inlet frits of HPLC columns on a micrometre scale0.1 µm1 µm10 µm0.2 µm filter:UHPLC, small particles0.45 µm filter:general HPLC0.5 µm frit:columns below 3 µm2 µm frit:5 µm columns5 µmpacking particleRound: filters and particles. Square: column inlet frits. Logarithmic scale.
Why the filter is matched to the column: the pore size has to be smaller than the inlet frit it protects.

The filter membrane

Choose the membrane first by what the sample is dissolved in, then by what must not be lost from the sample.

MembraneSuitsTake care with
Regenerated cellulose (RC)Aqueous and most organic solvents; a good general choice; low protein bindingStrong acids and bases
Nylon (polyamide)Aqueous and organic solvents; naturally hydrophilicBinds proteins and some acidic compounds; not for strong acids
PTFE (hydrophobic)Organic solvents, strong acids and basesDoes not wet with water: pre-wet with alcohol for aqueous samples, or use hydrophilic PTFE
Hydrophilic PTFEAqueous and organic solvents; very broad compatibility
PVDFAqueous and mild organic solvents; low protein bindingStrong bases; some ketones and amides
Polyethersulfone (PES)Aqueous samples, buffers, biological fluids; fast flow; low protein bindingMany organic solvents
Cellulose acetateAqueous biological samples; very low protein bindingOrganic solvents
Cellulose mixed estersAqueous samplesOrganic solvents
Glass fibrePrefiltration of dirty or viscous samplesNot a final filter: no defined small pore size

When in doubt, check the manufacturer's compatibility chart for the exact solvent, and see chemical resistance of plastics for the housing.

Filter size

Pick the diameter by the volume to be filtered. Small filters hold back less sample; large filters cope with more volume and more particles. See the syringe filter selection guide for the sizes, and use 47 mm membrane discs in a vacuum filtration unit for mobile phases.

Two things a filter can do to your result

Add something (extractables). Membranes and housings can release traces of compounds that appear as extra peaks, especially with strong solvents, low-wavelength UV detection or mass spectrometry. Use filters sold for chromatography, discard the first portion of filtrate (about 1 mL for a 25 mm filter), and run a filtered blank.

Remove something (adsorption). The analyte itself can bind to the membrane, which lowers the result, most noticeably at low concentrations and with proteins, peptides and some basic or hydrophobic compounds. Check during method development by comparing a filtered standard with an unfiltered or centrifuged one. If the filtered result is low, try another membrane or discard a larger first portion so the binding sites are saturated.

Ion chromatography

In ion chromatography the analytes are common ions such as chloride, nitrate, sulfate and sodium, at low levels. Ordinary filters can release these same ions. Use filters tested for low ionic extractables, rinse the filter with ultrapure water before use where the method allows, and always run a filtered blank. Plastic vials are generally preferred to glass for the same reason.

Good practice

  1. Use each filter once.
  2. Filter into a clean vial and close it straight away.
  3. Do not force a blocking filter: the membrane can rupture and release what it had held. Change to a new one, with a prefilter.
  4. Record the filter type in the method, so the result can be repeated.

Filtration protects the column from particles; an inline filter catches what the instrument itself sheds.

General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.