Why filter your sample?
The inlet frit of an HPLC column has pores of about 2 µm, and about 0.5 µm or less on columns packed with very small particles. Anything larger collects there. The result is rising back pressure, split or tailing peaks and a shortened column life. Particles also wear injector seals and block capillaries and needles. Filtering each sample costs little; a blocked column costs a great deal more.
Filter the mobile phase too, whenever it contains dissolved salts or additives (mobile phase preparation).
Choosing the right pore size
| Column packing | Filter pore size |
|---|---|
| Particles of 3 µm and larger | 0.45 µm |
| Particles below 3 µm; UHPLC; LC-MS | 0.2 µm (also sold as 0.22 µm) |
| Samples heavy with particles | A glass-fibre prefilter first, then 0.45 or 0.2 µm |
A finer filter than necessary blocks sooner and needs more force, so match the pore size to the column.
The filter membrane
Choose the membrane first by what the sample is dissolved in, then by what must not be lost from the sample.
| Membrane | Suits | Take care with |
|---|---|---|
| Regenerated cellulose (RC) | Aqueous and most organic solvents; a good general choice; low protein binding | Strong acids and bases |
| Nylon (polyamide) | Aqueous and organic solvents; naturally hydrophilic | Binds proteins and some acidic compounds; not for strong acids |
| PTFE (hydrophobic) | Organic solvents, strong acids and bases | Does not wet with water: pre-wet with alcohol for aqueous samples, or use hydrophilic PTFE |
| Hydrophilic PTFE | Aqueous and organic solvents; very broad compatibility | |
| PVDF | Aqueous and mild organic solvents; low protein binding | Strong bases; some ketones and amides |
| Polyethersulfone (PES) | Aqueous samples, buffers, biological fluids; fast flow; low protein binding | Many organic solvents |
| Cellulose acetate | Aqueous biological samples; very low protein binding | Organic solvents |
| Cellulose mixed esters | Aqueous samples | Organic solvents |
| Glass fibre | Prefiltration of dirty or viscous samples | Not a final filter: no defined small pore size |
When in doubt, check the manufacturer's compatibility chart for the exact solvent, and see chemical resistance of plastics for the housing.
Filter size
Pick the diameter by the volume to be filtered. Small filters hold back less sample; large filters cope with more volume and more particles. See the syringe filter selection guide for the sizes, and use 47 mm membrane discs in a vacuum filtration unit for mobile phases.
Two things a filter can do to your result
Add something (extractables). Membranes and housings can release traces of compounds that appear as extra peaks, especially with strong solvents, low-wavelength UV detection or mass spectrometry. Use filters sold for chromatography, discard the first portion of filtrate (about 1 mL for a 25 mm filter), and run a filtered blank.
Remove something (adsorption). The analyte itself can bind to the membrane, which lowers the result, most noticeably at low concentrations and with proteins, peptides and some basic or hydrophobic compounds. Check during method development by comparing a filtered standard with an unfiltered or centrifuged one. If the filtered result is low, try another membrane or discard a larger first portion so the binding sites are saturated.
Ion chromatography
In ion chromatography the analytes are common ions such as chloride, nitrate, sulfate and sodium, at low levels. Ordinary filters can release these same ions. Use filters tested for low ionic extractables, rinse the filter with ultrapure water before use where the method allows, and always run a filtered blank. Plastic vials are generally preferred to glass for the same reason.
Good practice
- Use each filter once.
- Filter into a clean vial and close it straight away.
- Do not force a blocking filter: the membrane can rupture and release what it had held. Change to a new one, with a prefilter.
- Record the filter type in the method, so the result can be repeated.
Filtration protects the column from particles; an inline filter catches what the instrument itself sheds.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.