What a syringe filter is
A syringe filter is a single-use membrane sealed in a plastic housing with a Luer inlet. Fitted to a syringe, it removes particles from a small volume of liquid in seconds. It is the standard way to prepare samples for HPLC, UHPLC, ion chromatography and dissolution testing, and with a sterile 0.2 µm membrane it is also used to sterilise small volumes of aqueous solution.
Four questions settle the choice. Browse syringe filters.
1) Sample volume: which diameter?
| Filter diameter | Typical sample volume | Liquid held back in the filter |
|---|---|---|
| 3 to 4 mm | Up to about 1 mL | Very little |
| 13 to 15 mm | About 1 to 10 mL | Small |
| 25 mm | About 10 to 100 mL | More |
| 30 to 33 mm | Up to roughly 150 mL | Most |
The figures are a guide: the volume a filter can handle falls sharply with dirty samples. Use the smallest diameter that copes with the volume, because a larger filter keeps back more of a small, precious sample. Exact hold-up volumes are given by each manufacturer.
2) Size and amount of particulates in the sample
A sample that looks cloudy, or is viscous, blocks a fine membrane almost at once. Use a filter with a built-in glass-fibre prefilter, which catches the bulk of the particles before the membrane, or centrifuge first and filter the clear liquid. For very dirty samples filter in two stages, coarse then fine.
3) Membrane compatibility with the sample and solvent
The membrane must be chemically resistant to the solvent and must not bind the analyte.
| Sample | Usual membrane |
|---|---|
| Aqueous solutions and buffers | Regenerated cellulose, nylon, PES, hydrophilic PTFE, cellulose acetate |
| Mixtures of water and organic solvent (typical HPLC samples) | Regenerated cellulose, nylon, hydrophilic PTFE |
| Organic solvents | PTFE, regenerated cellulose, nylon |
| Strong acids and bases | PTFE |
| Protein and peptide solutions | PES, PVDF, cellulose acetate, regenerated cellulose (low binding) |
| Samples heavy with particles | Glass fibre, alone or as a prefilter |
Hydrophobic PTFE does not let water through unless it is first wetted with a little alcohol. The membranes are compared in more detail in selecting filters for HPLC and UHPLC.
4) Pore size for the application
| Pore size | Use |
|---|---|
| 0.2 µm (0.22 µm) | UHPLC and columns with particles below 3 µm; LC-MS; sterilising filtration of aqueous solutions with a sterile filter |
| 0.45 µm | General HPLC with particles of 3 µm and larger; dissolution samples; general clarification |
| 1 to 5 µm | Coarse clarification; prefiltration |
Housing and connections
Most housings are polypropylene, which resists the usual HPLC solvents. A Luer lock inlet screws onto the syringe and will not blow off under pressure; a Luer slip inlet simply pushes on. The outlet is a Luer slip tip that fits into the neck of a vial.
How to use one properly
- Draw the sample into the syringe, with a little air behind it to help empty the filter.
- Attach the filter firmly.
- Push gently and steadily. Discard the first portion (a few drops for small filters, about 1 mL for a 25 mm filter) to waste: it carries most of any extractables and has lost most analyte to adsorption.
- Collect the rest in a clean vial and cap it.
- If the resistance rises, stop and take a new filter. Forcing it can rupture the membrane.
A small syringe generates far more pressure than a large one for the same thumb force. With 25 mm filters, use a syringe of 10 mL or larger where possible, and do not exceed the pressure rating of the filter.
Sterile or non-sterile
Non-sterile filters are used for analytical sample preparation. Individually wrapped sterile filters are for sterilising solutions and for microbiological work. Sterility has no bearing on chromatographic cleanliness: for trace analysis look instead for filters certified for low extractables.
Converting sizes
Filter diameters are metric. If a document gives inches, 13 mm is 0.51 inch and 25 mm is 0.98 inch; see inches to mm.
General guidance from Lancetech Limited. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.