What an inline filter does
An inline filter is a small holder with a replaceable frit, fitted in the flow path just before the column. It catches particles before they reach the inlet frit of the column, which is the usual place a column blocks. When the pressure climbs, you change an inexpensive frit instead of the column.
- Pump and injectorShed seal wear particles
- Inline filterA frit: stops particles
- Guard columnPacking: also traps strongly retained compounds
- Analytical columnThe part being protected
- Detector
Where the particles come from
- The sample: undissolved material, or material that precipitates when the sample meets the mobile phase.
- The instrument: wear from pump piston seals and from the rotor seal of the injection valve.
- The mobile phase: precipitated buffer, microbial growth in aqueous reservoirs, dust.
Filtering samples and mobile phases deals with the first and third. Only a filter between the injector and the column deals with the second.
Which frit porosity
Match the filter frit to the inlet frit of the column it protects:
| Column packing | Typical inline filter frit |
|---|---|
| 5 µm particles and larger | 2 µm |
| 3 to 3.5 µm particles | 0.5 µm (some makers fit 2 µm inlet frits: check the column's care sheet) |
| Below 3 µm, including sub-2 µm and solid-core particles | 0.5 µm (or 0.2 µm for UHPLC) |
Follow the column manufacturer's recommendation where one is given.
Inline filter, guard column or both
| Inline filter | Guard column | |
|---|---|---|
| Contains | A frit only | A short bed of packing between frits |
| Stops particles | Yes | Yes |
| Traps strongly retained compounds | No | Yes |
| Effect on the separation | Very small | Small, if the packing matches the column |
| Typical use | Clean samples; UHPLC | Dirty samples and complex matrices |
A guard column should carry the same phase as the analytical column, or it changes the separation. For dirty samples on an expensive column, an inline filter followed by a guard column is a sensible pairing.
Solvent inlet filters are something else
The filters on the ends of the solvent lines in the reservoirs (often called sinkers) are coarse, typically 10 to 20 µm. They protect the pump and check valves from large particles and are not a substitute for an inline filter. A partly blocked inlet filter starves the pump and causes bubbles and unstable flow, so clean or replace these as well (air bubbles in HPLC).
Fitting it without losing efficiency
An inline filter adds a little volume to the flow path. On columns of 4.6 mm inner diameter this is negligible. On 2.1 mm columns and in UHPLC, use a filter designed for low dispersion, connect it directly to the column or with the shortest narrow-bore capillary, and make the fittings correctly (preventing dead volume).
When to replace the frit
- Note the system pressure with a new frit under a fixed condition.
- Replace the frit when the pressure has risen clearly above that value, or on a fixed schedule based on the number of injections.
- If frits block quickly, look for the source: unfiltered samples, a sample solvent stronger than the mobile phase, buffer precipitating in the gradient, or a worn pump or rotor seal.
Do not try to clean and reuse frits for regulated work; the saving is small and the risk of carry-over is real.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.