What shortens a column's life
Most columns do not wear out: they are blocked, contaminated or chemically damaged. The usual causes are particles on the inlet frit, strongly retained material from the sample, a mobile phase outside the column's pH range, pressure shocks, and buffer left to dry inside the column.
Daily working rules
- Stay inside the stated limits for pH, pressure and temperature. For silica-based phases, low pH slowly strips the bonded phase and high pH dissolves the silica itself. Both are irreversible, and both go faster at high temperature.
- Filter every sample, through 0.45 µm, or 0.2 µm for particles below 3 µm (sample filtration). Centrifuging helps but does not replace it.
- Clean up dirty samples. Proteins, fats and pigments that stick to the column are best removed beforehand, for example by solid phase extraction.
- Use a guard column or inline filter and replace it on schedule (inline filters). It is far cheaper than the column behind it.
- Change flow and composition gradually. Step up the flow; avoid sudden pressure jumps.
- Dissolve samples in the mobile phase or a weaker solvent, so nothing precipitates at the column inlet.
- Use the column for one type of method where you can, especially with ion-pair reagents.
- Keep a log: injections, pressure at a reference condition, and a regular test chromatogram.
End of the working day
If a run stops overnight in a mobile phase that is mild (no extreme pH, no high salt), the column can simply stay in it. If the mobile phase contains buffer or salt and the column will stand for longer, flush it out as below. Never leave a column standing in buffer for days: salt crystallises, bacteria grow and metal parts corrode.
Storing a reversed-phase column
- Flush with 10 to 20 column volumes of the same water and organic mixture without the buffer to remove salts. Do not switch straight from buffer to high organic solvent: the salt can precipitate inside the column.
- Flush with 10 to 20 column volumes of a storage solvent rich in organic solvent, such as acetonitrile and water or methanol and water with at least half organic, or the shipping solvent named on the certificate.
- Close both ends tightly with the original plugs. A column that dries out is hard to recover.
- Store it in its box at room temperature, away from vibration and heat, labelled with the storage solvent and the date.
- Flush out the bufferSame water and organic mixture, without the salt
- Change to storage solventRich in acetonitrile or methanol
- Close both endsTightly, with the original plugs
- Label and storeSolvent and date, at room temperature
Other types of column
| Column type | Typical storage |
|---|---|
| Normal phase (silica, cyano, diol, amino in non-polar solvents) | A non-polar solvent such as heptane or hexane, or 2-propanol |
| HILIC | A mixture high in acetonitrile without buffer, as the manufacturer states |
| Ion exchange and aqueous size exclusion | The aqueous solution the manufacturer specifies, usually with something to stop microbial growth |
| Polymer-based | As specified; many must never dry out or meet certain solvents |
When in doubt, the manufacturer's care sheet for that column decides.
Bringing a column back into use
Check which solvent it was stored in, flush with a miscible intermediate if needed, equilibrate (column equilibration) and run the test mixture. Compare with the earlier result before trusting it with samples.
When performance drops
Rising pressure, tailing or split peaks and lost retention do not always mean the column is finished. Work through regeneration and cleaning first, and check the rest of the system before blaming the column.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.