What SPE does

Solid phase extraction is a sample preparation technique that uses a small bed of sorbent to separate the compounds you want from the rest of the sample. It works by the same interactions as liquid chromatography, in a simple on-off form: conditions are chosen so that a compound is either held on the sorbent or passes straight through.

It is used to:

  • clean up a sample, removing matrix that would interfere with the analysis or foul the instrument;
  • concentrate trace analytes from a large volume into a small one;
  • change the solvent to one suited to the instrument.

Compared with liquid-liquid extraction it uses less solvent, forms no emulsions, is easier to automate and offers many more selectivities.

Two strategies

  • Retain the analyte (bind and elute). The analyte is held on the sorbent while the matrix passes through or is washed away; the analyte is then released with a small volume of a stronger solvent. This both cleans and concentrates.
  • Retain the interferences (pass-through). The sorbent holds the unwanted matrix and the analyte passes through and is collected. This is quick, and is the principle behind dispersive clean-up in QuEChERS.

The four steps

  1. Condition. Wet the sorbent with a solvent such as methanol, then bring it to the conditions of the sample, for example water or buffer. Conditioning activates the bonded phase and removes air. With silica-based reversed-phase sorbents, do not let the bed run dry before the sample is loaded, or recovery suffers.
  2. Load. Pass the sample through slowly, at a steady flow of roughly one drop a second. The analyte needs time to interact with the sorbent.
  3. Wash. Rinse with a solvent strong enough to remove weakly held interferences but too weak to move the analyte. Drying the bed under vacuum at this point removes water before elution with an organic solvent.
  4. Elute. Release the analyte with a small volume of a strong solvent. Two small portions, each allowed to soak for a moment, recover more than one large portion.

The eluate is then injected directly, or evaporated and taken up in a solvent that suits the analysis.

Retention modes

ModeSorbent examplesSample isRetainsEluted with
Reversed phaseC18, C8, phenyl, polymeric sorbentsAqueousNon-polar to moderately polar compoundsMethanol, acetonitrile or another organic solvent
Normal phaseSilica, Florisil, alumina, amino, cyano, diolA non-polar organic solventPolar compoundsA more polar solvent
Cation exchangeSulfonic acid (strong), carboxylic acid (weak)Aqueous, low ionic strengthPositively charged compounds (bases)A change of pH or high salt
Anion exchangeQuaternary amine (strong), amino (weak)Aqueous, low ionic strengthNegatively charged compounds (acids)A change of pH or high salt
Mixed modeReversed phase combined with ion exchangeAqueousCharged compounds, very selectivelyOrganic solvent with acid or base

How to pick among them is covered in selecting the SPE sorbent.

Formats

  • Cartridges (columns): syringe-barrel tubes of 1 to 60 mL holding from tens of milligrams to several grams of sorbent. The general-purpose format. See choosing the cartridge size.
  • 96-well plates: for high sample numbers and automation.
  • Disks: for fast extraction of large water volumes.
  • Dispersive SPE: loose sorbent shaken with an extract and then centrifuged, as in QuEChERS for pesticide residues in food.

Liquid is drawn through with a vacuum manifold, pushed through with positive pressure, or spun through in a centrifuge.

When recovery is low

FindingLikely causeRemedy
Analyte found in the liquid that passed through during loadingSample solvent too strong; wrong pH; flow too fast; too little sorbentDilute the sample with water or buffer, adjust the pH, slow the flow, use more sorbent
Analyte found in the washWash solvent too strongWeaken the wash
Analyte found nowhereStill on the sorbentStronger or larger elution volume; add acid or base; soak
Results vary between cartridgesBed ran dry, or flow variedKeep the bed wet during conditioning; control the flow

Collecting and analysing each fraction (load, wash and eluate) during method development shows exactly where the analyte goes.

General guidance from Lancetech Limited. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.

New products and price changes, once a month

No daily emails. Unsubscribe in one click.

  • New arrivals
  • Price changes
  • Brand offers

About once a month. Unsubscribe any time. See our privacy policy.

Solid phase extraction (SPE) explained: steps, modes and formats | Lancetech Limited