Solid phase extraction (SPE): selection of the sorbent

How to choose an SPE sorbent from the sample matrix and the analyte: reversed phase, normal phase, strong and weak ion exchangers, mixed mode and clean-up.

  • 3 min read
  • 1 figure
  • Reviewed 25 November 2026
On this page 8 sections
  1. Two questions decide the sorbent
  2. A decision table
  3. Reversed-phase sorbents
  4. Normal-phase sorbents
  5. Ion-exchange sorbents
  6. Mixed-mode sorbents
  7. Sorbents that remove interferences
  8. Special cases

Two questions decide the sorbent

  1. What is the sample dissolved in? Water, or an organic solvent?
  2. What is the analyte like? Non-polar, polar, or carrying a charge?

The sorbent must attract the analyte more strongly than the sample liquid does.

A decision table

Sample matrixAnalyteModeTypical sorbents
AqueousNon-polar to moderately polar, unchargedReversed phaseC18, C8, phenyl; polymeric reversed-phase sorbents
AqueousPolar and uncharged, poorly held on C18Reversed phase with a more retentive sorbentHydrophilic-lipophilic polymeric sorbents; graphitised carbon
AqueousBasic (positively charged)Cation exchange or mixed modeStrong or weak cation exchangers; mixed-mode cation exchange
AqueousAcidic (negatively charged)Anion exchange or mixed modeStrong or weak anion exchangers; mixed-mode anion exchange
Non-polar organic solventPolarNormal phaseSilica, Florisil, alumina, diol, amino, cyano

Browse solid phase extraction products and sorbents for SPE.

Choosing an SPE sorbent from the sample liquid and the analyteWhat is the sample in?WaterNon-polar organic solventIs the analyte charged?Normal phaseSilica, Florisil, diol, aminoNoYes: a base (+)Yes: an acid (-)Reversed phaseC18, C8, polymerCation exchangeor mixed modeAnion exchangeor mixed modeWeak base: strong exchangerStrong base: weak exchangerWeak acid: strong exchangerStrong acid: weak exchangerKeep the analyte unchargedby adjusting the pH
A first choice of sorbent from two questions: what the sample is dissolved in, and whether the analyte carries a charge.

Reversed-phase sorbents

  • C18 is the most retentive silica-based phase and the usual starting point for extracting organic compounds from water, drinks and biological fluids.
  • C8, C4 and phenyl hold compounds less strongly, which makes very non-polar analytes easier to elute.
  • End-capped and non-end-capped versions differ. Without end-capping, the remaining silanol groups add polar and ion-exchange interactions, which helps retain basic compounds but can make them hard to elute.
  • Polymeric sorbents (based on polystyrene-divinylbenzene, often modified to be water-wettable) offer higher capacity, stability across the whole pH range and tolerance of running dry. They retain polar compounds better than C18.

For reversed-phase retention the analyte should be uncharged: adjust the sample to at least two pH units below the pKa for an acid, or above it for a base.

Normal-phase sorbents

Polar sorbents retain polar compounds from non-polar solvents such as hexane. Silica is the general choice. Florisil (magnesium silicate) and alumina are established for cleaning up pesticide and environmental extracts. Diol, amino and cyano bonded phases are milder and less sensitive to water. Elute with a more polar solvent.

Ion-exchange sorbents

Ion exchangers hold compounds by charge, which is highly selective. Both the sorbent and the analyte must be charged while loading, and one of them must be neutralised (or swamped with salt) to elute.

SorbentFunctional groupChargedBest for
Strong cation exchangerSulfonic acidAt all pH valuesWeak bases, which can be neutralised to elute
Weak cation exchangerCarboxylic acidFully only above about pH 7Strong bases and quaternary amines, which are always charged
Strong anion exchangerQuaternary amineAt all pH valuesWeak acids, which can be neutralised to elute
Weak anion exchangerPrimary, secondary or tertiary amineOnly below about pH 8Strong acids, which are always charged

The rule to remember: pair a strong exchanger with a weak analyte, and a weak exchanger with a strong analyte. If both are permanently charged, nothing can be switched off and the analyte is very hard to recover. Ion exchange also needs a sample of low ionic strength, since salts compete for the sites.

Mixed-mode sorbents

These combine a reversed-phase backbone with ion-exchange groups. The analyte is held by charge, so the bed can be washed with pure organic solvent to remove neutral interferences, before the analyte is released with an organic solvent containing acid or base. They give the cleanest extracts for basic or acidic drugs in biological fluids.

Sorbents that remove interferences

In pass-through and dispersive clean-up, as in QuEChERS, the sorbent is chosen for what it takes out:

SorbentRemoves
Primary secondary amine (PSA)Sugars, organic acids, fatty acids
C18Fats, waxes and other non-polar matrix
Graphitised carbon blackPigments such as chlorophyll and carotenoids; note it also holds planar analytes
Anhydrous magnesium sulfateResidual water

Special cases

  • Large molecules need wide-pore sorbents.
  • Very volatile analytes are easily lost when the eluate is evaporated: choose a sorbent and an elution volume that avoid the evaporation step.
  • Regulated methods often name the sorbent type, as in PFAS analysis, where weak anion-exchange sorbents are standard.

Once the sorbent is chosen, the amount is next: see packing weight and cartridge size. The procedure itself is in solid phase extraction.

General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.