Two questions decide the sorbent
- What is the sample dissolved in? Water, or an organic solvent?
- What is the analyte like? Non-polar, polar, or carrying a charge?
The sorbent must attract the analyte more strongly than the sample liquid does.
A decision table
| Sample matrix | Analyte | Mode | Typical sorbents |
|---|---|---|---|
| Aqueous | Non-polar to moderately polar, uncharged | Reversed phase | C18, C8, phenyl; polymeric reversed-phase sorbents |
| Aqueous | Polar and uncharged, poorly held on C18 | Reversed phase with a more retentive sorbent | Hydrophilic-lipophilic polymeric sorbents; graphitised carbon |
| Aqueous | Basic (positively charged) | Cation exchange or mixed mode | Strong or weak cation exchangers; mixed-mode cation exchange |
| Aqueous | Acidic (negatively charged) | Anion exchange or mixed mode | Strong or weak anion exchangers; mixed-mode anion exchange |
| Non-polar organic solvent | Polar | Normal phase | Silica, Florisil, alumina, diol, amino, cyano |
Browse solid phase extraction products and sorbents for SPE.
Reversed-phase sorbents
- C18 is the most retentive silica-based phase and the usual starting point for extracting organic compounds from water, drinks and biological fluids.
- C8, C4 and phenyl hold compounds less strongly, which makes very non-polar analytes easier to elute.
- End-capped and non-end-capped versions differ. Without end-capping, the remaining silanol groups add polar and ion-exchange interactions, which helps retain basic compounds but can make them hard to elute.
- Polymeric sorbents (based on polystyrene-divinylbenzene, often modified to be water-wettable) offer higher capacity, stability across the whole pH range and tolerance of running dry. They retain polar compounds better than C18.
For reversed-phase retention the analyte should be uncharged: adjust the sample to at least two pH units below the pKa for an acid, or above it for a base.
Normal-phase sorbents
Polar sorbents retain polar compounds from non-polar solvents such as hexane. Silica is the general choice. Florisil (magnesium silicate) and alumina are established for cleaning up pesticide and environmental extracts. Diol, amino and cyano bonded phases are milder and less sensitive to water. Elute with a more polar solvent.
Ion-exchange sorbents
Ion exchangers hold compounds by charge, which is highly selective. Both the sorbent and the analyte must be charged while loading, and one of them must be neutralised (or swamped with salt) to elute.
| Sorbent | Functional group | Charged | Best for |
|---|---|---|---|
| Strong cation exchanger | Sulfonic acid | At all pH values | Weak bases, which can be neutralised to elute |
| Weak cation exchanger | Carboxylic acid | Fully only above about pH 7 | Strong bases and quaternary amines, which are always charged |
| Strong anion exchanger | Quaternary amine | At all pH values | Weak acids, which can be neutralised to elute |
| Weak anion exchanger | Primary, secondary or tertiary amine | Only below about pH 8 | Strong acids, which are always charged |
The rule to remember: pair a strong exchanger with a weak analyte, and a weak exchanger with a strong analyte. If both are permanently charged, nothing can be switched off and the analyte is very hard to recover. Ion exchange also needs a sample of low ionic strength, since salts compete for the sites.
Mixed-mode sorbents
These combine a reversed-phase backbone with ion-exchange groups. The analyte is held by charge, so the bed can be washed with pure organic solvent to remove neutral interferences, before the analyte is released with an organic solvent containing acid or base. They give the cleanest extracts for basic or acidic drugs in biological fluids.
Sorbents that remove interferences
In pass-through and dispersive clean-up, as in QuEChERS, the sorbent is chosen for what it takes out:
| Sorbent | Removes |
|---|---|
| Primary secondary amine (PSA) | Sugars, organic acids, fatty acids |
| C18 | Fats, waxes and other non-polar matrix |
| Graphitised carbon black | Pigments such as chlorophyll and carotenoids; note it also holds planar analytes |
| Anhydrous magnesium sulfate | Residual water |
Special cases
- Large molecules need wide-pore sorbents.
- Very volatile analytes are easily lost when the eluate is evaporated: choose a sorbent and an elution volume that avoid the evaporation step.
- Regulated methods often name the sorbent type, as in PFAS analysis, where weak anion-exchange sorbents are standard.
Once the sorbent is chosen, the amount is next: see packing weight and cartridge size. The procedure itself is in solid phase extraction.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.