What is PFAS?
PFAS stands for per- and polyfluoroalkyl substances: a group of thousands of man-made chemicals that contain chains of carbon atoms bonded to fluorine. The carbon-fluorine bond is one of the strongest in chemistry. It gives PFAS their useful properties, repelling water, oil and grease and resisting heat, and it is also why they barely break down in the environment, earning the name "forever chemicals".
They have been used in non-stick coatings, water- and stain-repellent textiles, food packaging, firefighting foams and many industrial processes. Well-known examples are perfluorooctanoic acid (PFOA), perfluorooctane sulfonic acid (PFOS), perfluorohexane sulfonic acid (PFHxS) and perfluorononanoic acid (PFNA).
PFAS are now found in water, soil, food, wildlife and human blood worldwide, and some are linked to harmful effects on health. Regulators have responded with very low limits. The recast EU Drinking Water Directive, for example, sets 0.10 µg/L for the sum of 20 named PFAS and 0.50 µg/L for total PFAS. Measuring at these levels, in the low nanogram per litre range for individual compounds, is demanding.
How PFAS are measured
Targeted PFAS analysis uses liquid chromatography with tandem mass spectrometry (LC-MS/MS), with negative electrospray ionisation and isotope-labelled internal standards. Water samples are first concentrated and cleaned by solid phase extraction.
| Method | Scope | Extraction |
|---|---|---|
| US EPA 537.1 | Drinking water | SPE on a styrene-divinylbenzene polymer |
| US EPA 533 | Drinking water, including shorter-chain PFAS | SPE on a weak anion exchanger; isotope dilution |
| US EPA 1633 | Waste water, surface and ground water, soil, sediment, biosolids, tissue | SPE on a weak anion exchanger, with carbon clean-up; isotope dilution |
| ISO 21675 | Drinking, surface and waste water | SPE on a weak anion exchanger |
| EN 17892 | Drinking water; the 20 PFAS of the EU directive | Direct injection or SPE |
The method you must follow is the one named by your regulator or client. Read its current version: the lists of compounds and the quality control requirements are revised.
Which products are suitable for PFAS analysis?
Solid phase extraction
Weak anion-exchange (WAX) polymeric sorbents are the standard for PFAS. Most target PFAS are acids and are negatively charged in water. A mixed-mode weak anion exchanger holds the long-chain compounds by its reversed-phase backbone and the short-chain, more water-soluble ones by ion exchange, then releases all of them with a basic organic solvent such as methanol containing ammonia. Cartridges sold for PFAS are tested by the manufacturer for low PFAS background. Browse solid phase extraction products; see selecting the sorbent and cartridge size.
HPLC columns
- An analytical column, typically C18, separates the PFAS, including branched and linear isomers.
- A delay column is fitted between the pump's mixer and the injector. PFAS leaching from the instrument's own tubing and solvents are held back on it and elute later than the same compounds from the sample, so the two can be told apart.
Vials, caps and containers
- Use polypropylene or polyethylene (HDPE) bottles and vials. PFAS adsorb onto glass, which lowers results.
- Use caps with polyethylene or polypropylene seals. Avoid PTFE-lined septa and caps: PTFE is itself a fluoropolymer and a source of contamination.
- See sample bottles and caps and vial selection.
Filters
Filtration can both add PFAS and remove them by adsorption. Avoid PTFE and other fluoropolymer membranes. Where a method permits filtration, use a filter type verified for it, and check recovery and blanks yourself (sample filtration).
Reference standards
Quantitation relies on certified native reference standards and on isotope-labelled analogues added to every sample before extraction.
Avoiding contamination
PFAS are in many everyday and laboratory materials, and the target levels are extremely low. Contamination control is the heart of the analysis.
- Remove fluoropolymers from the sample path wherever possible: PTFE tubing, solvent line frits, degasser parts, seals, septa, and PTFE-coated stirrer bars. Replace with PEEK, stainless steel or polyolefin parts where the instrument allows.
- Check solvents and reagents by batch, including the water, methanol and ammonium salts.
- Avoid common sources at sampling and in the laboratory: water-repellent clothing, some food packaging, certain personal care products, and paper or labels with stain-resistant coatings.
- Run blanks at every stage: field blanks, laboratory reagent blanks and instrument blanks. A method is only as good as its blank.
- Rinse sampling and laboratory equipment with PFAS-free solvent, and dedicate glassware and pipettes to the analysis.
Beyond targeted analysis
Targeted methods measure a defined list of a few dozen compounds. To estimate the PFAS that a list misses, laboratories also use sum parameters such as adsorbable organic fluorine, the total oxidisable precursor assay, and non-targeted screening by high-resolution mass spectrometry.
General guidance from Lancetech Limited. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.