Sample bottles and caps: choosing the right container

How to choose sample bottles and caps: glass or plastic by analysis, clear or amber, neck and thread sizes such as GL 45, cap liners, filling and transport.

  • 3 min read
  • 1 figure
  • Reviewed 25 November 2026
On this page 9 sections
  1. The container is part of the analysis
  2. Glass or plastic
  3. Clear or amber
  4. Neck, thread and cap
  5. Liners and seals
  6. Matching the bottle to the analysis
  7. Filling
  8. Cleanliness
  9. Labelling and transport

The container is part of the analysis

A sample can only be as good as the bottle it travels and waits in. The container can add to the sample (ions from glass, additives from plastic, residues from earlier use), remove from it (adsorption on the wall, loss of volatiles through the closure) or let it change (light, air, microbial growth). Choosing the bottle and cap is therefore part of the method. Browse labware.

Glass or plastic

MaterialGood forAvoid for
Borosilicate glassOrganic compounds, solvents, most reagents; heating and autoclavingTrace metals, sodium, boron and silica; hydrofluoric acid; strong hot alkali
Soda-lime glassGeneral storage of solids and non-critical liquidsTrace analysis; temperature shock
HDPE (high-density polyethylene)Aqueous samples, acids and bases, trace metalsMany organic solvents; trace organics
LDPE (low-density polyethylene)Wash bottles, dropping bottles, aqueous samplesOrganic solvents; heating
PolypropyleneAqueous samples, trace metals, autoclavingStrong oxidising acids; some solvents
PETClear, light bottles for aqueous and biological samplesHeating; strong acids and bases
Fluoropolymers (FEP, PFA)Ultra-trace metals; aggressive chemicalsPFAS analysis

The usual rule: organic analytes in glass, inorganic analytes in plastic. For the detail on what each plastic tolerates, see chemical resistance of plastics.

Clear or amber

Amber glass and opaque or amber plastic protect light-sensitive samples and reagents. Clear containers let you inspect the contents. Where a light-sensitive sample must be in clear glass, keep it in the dark or wrap it.

Neck, thread and cap

Laboratory bottles commonly use the GL thread series, named by the nominal outer diameter of the thread in millimetres: GL 25, GL 32, GL 45 and GL 80 are frequent. GL 45 is the standard neck for laboratory bottles from 100 mL to several litres. A cap of the same GL size fits bottles from any manufacturer following the standard.

  • Narrow necks pour cleanly and limit evaporation: for liquids.
  • Wide necks take spatulas, solids, pastes and viscous samples.

Cap materials matter when the bottle is heated. Standard polypropylene screw caps and pouring rings tolerate autoclaving at 121 °C; caps made of higher-temperature plastics are available for dry heat. The cap's limit is usually lower than the bottle's.

A laboratory bottle with a GL 45 screw cap, pouring ring and fill limit for autoclavingScrew cap, GL 45: the thread is45 mm acrossPouring ring: drip-free pouringBorosilicate glass bodyFor autoclaving, fill no more than aboutthree-quarters and loosen the capGraduations are approximate
A laboratory bottle with a GL 45 neck. A cap of the same GL size fits any bottle made to the standard.

Liners and seals

The part of the cap that touches the sample is the liner.

  • PTFE-faced liners are the general choice for organic and trace work: inert and clean.
  • Polyethylene cone or foam liners seal well for aqueous samples.
  • Septum caps with a hole allow sampling by needle without opening, as with volatile organic compounds in water (septum selection).
  • For PFAS, avoid PTFE liners altogether (PFAS analysis).

Matching the bottle to the analysis

AnalysisTypical container
Trace metalsAcid-cleaned HDPE, polypropylene or fluoropolymer
Pesticides, hydrocarbons and other organicsAmber glass with PTFE-lined cap
Volatile organic compounds in waterGlass vial with PTFE-faced silicone septum cap, filled without any air space
Nutrients, anions, general water chemistryPolyethylene or glass, as the method states
MicrobiologySterile bottle; with a dechlorinating agent for chlorinated water
Clinical specimensSterile specimen containers

The governing method or standard normally names the container, the preservative and the maximum holding time. Follow it. See sampling.

Filling

  • For volatile analytes, fill completely so there is no air space for them to escape into.
  • For samples that will be frozen, leave room for expansion. Water expands by about 9% on freezing and will crack a full glass bottle.
  • For samples that will be shaken or mixed, leave some headspace.
  • For autoclaving, loosen the cap and do not fill above about three-quarters.

Cleanliness

Use new or properly cleaned containers, and for trace work pre-cleaned, certified ones. Rinsing with the sample itself is common for water sampling, except where the bottle contains a preservative or the method forbids it. Never reuse a container of unknown history for trace analysis.

Labelling and transport

Label the bottle, not the cap, with a marker or label that survives water, cold and solvent. Tighten caps to a firm, even seal, and pack bottles upright with secondary containment and absorbent. Hazardous and biological samples are subject to transport regulations that set the packaging required; see waste disposal and sample transport.

General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.