Selection of sample vials in the laboratory
A vial has to fit the instrument, keep the sample unchanged and seal reliably. Choose in this order: the size and neck the autosampler takes, the closure type, the material, and any insert for small volumes. Browse vials and caps and septa.
Sizes and neck finishes
Vial necks are commonly named by their nominal diameter in millimetres, often written with an N (N 9, N 11, N 20).
| Vial | Typical dimensions | Neck finish | Use |
|---|---|---|---|
| 1.5 to 2 mL autosampler vial | 12 x 32 mm (11.6 x 32 mm) | 9 mm short screw thread; 8 mm or 10 mm screw; 11 mm crimp; 11 mm snap ring | The standard vial for HPLC and GC autosamplers |
| 4 mL vial | 15 x 45 mm | 13 mm screw; 13 mm crimp | Larger sample volumes; some autosamplers; storage |
| 10 mL headspace vial | 22.5 x 46 mm | 20 mm crimp; 18 mm screw | Headspace GC |
| 20 mL headspace vial | 22.5 x 75 mm | 20 mm crimp; 18 mm screw | Headspace GC; the most common headspace size |
| 20, 40 and 60 mL storage vials | 27.5 mm diameter, various heights | 24 mm screw | Sample storage; environmental water samples |
Autosamplers are built for particular dimensions. Check the instrument's requirements, especially vial height and whether the gripper needs a specific cap or a magnetic one.
Screw, crimp or snap
| Closure | Advantages | Limits |
|---|---|---|
| Screw cap | Opened and closed by hand; no tool; good seal | Seal depends on how firmly it is tightened |
| Crimp cap | The tightest seal; best for volatile samples and long storage | Needs a crimper and a decapper, set correctly |
| Snap ring cap | Fastest to apply; no tool | The least tight; not for volatile solvents or long storage |
For crimp caps, see correct crimping of sample vials.
Glass of the first hydrolytic class
Laboratory vials are normally made of borosilicate glass of hydrolytic class 1 (Type I), which releases very little alkali into the sample and resists chemicals and temperature changes.
- Clear glass lets you see the sample.
- Amber glass protects light-sensitive samples.
Silanised sample vials
The surface of glass carries silanol groups that can adsorb polar and basic compounds. At trace concentrations a noticeable share of the analyte can be lost to the wall, giving low and variable results. Silanised (deactivated) vials have a treated, water-repellent surface that greatly reduces this. Consider them for trace analysis of amines, basic drugs, pesticides and other compounds that stick to glass.
Plastic vials
Polypropylene vials are the choice when glass itself is the problem:
- Ion chromatography and metals: glass releases sodium and other ions.
- Proteins, peptides and some other biomolecules that adsorb to glass. Suitability depends on the molecule, so test recovery.
- PFAS analysis, together with caps that contain no PTFE (PFAS analysis).
- Alkaline samples and hydrofluoric acid, which attack glass.
- LC-MS methods troubled by sodium adducts.
Check that the plastic resists the sample solvent (chemical resistance of plastics). Some organic solvents extract additives from plastics.
Small volumes: inserts and micro-vials
For samples of less than about 0.3 mL, a micro-insert placed in a standard vial raises the liquid so the needle can reach it. Conical inserts allow almost complete withdrawal; inserts with a spring or polymer feet cushion and centre themselves. Vials with a built-in conical or tapered interior serve the same purpose without a separate part. Match the insert to the vial's opening: wide-opening vials take wider inserts.
HPLC and GC applications with an autosampler
- Fill a vial to about the shoulder, not to the brim. The needle displaces liquid, and a completely full, tightly sealed vial can form a vacuum as sample is withdrawn, giving poor injection repeatability.
- Check the needle depth setting when using inserts.
- For repeat injections from one vial, a pre-slit septum prevents vacuum; for volatile samples and storage, a closed septum prevents evaporation (septum selection).
- Use vials and caps from one system, so that the cap, septum and neck match.
Headspace vials and other applications
Headspace vials have a thicker, uniform wall to withstand heat and pressure, with a flat or rounded bottom, and a bevelled or flat neck made for a pressure-tight seal. Do not substitute ordinary storage vials. See basic principles of headspace analysis.
For storing and shipping samples, larger screw-neck vials and bottles are covered in sample bottles and caps.
Clean and certified
Vials are single-use items for analytical work: washing rarely removes everything and adds its own residues. For trace analysis and mass spectrometry, choose vials and caps sold as cleaned or certified for low background.
General guidance from Lancetech Limited. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.