What a septum has to do
The septum is the disc in the cap that the autosampler needle pierces. It has to seal the vial against evaporation and contamination, let the needle through without damaging it, close again afterwards if more injections follow, and add nothing to the sample. No single material does all of this best, so the choice depends on the solvent, the temperature and how often the vial is pierced. Browse caps and septa.
How a septum is built
Most septa are a layer of rubber or silicone, which gives the seal and the ability to close again, faced with a thin layer of PTFE on the side towards the sample. The PTFE is the chemical barrier. Once the needle has pierced it, the sample or its vapour can reach the elastic layer behind, which matters for the second and later injections.
The common materials
| Septum | Reseals after piercing | Typical temperature range | Use |
|---|---|---|---|
| PTFE / silicone | Very well | About -40 to 200 °C | The general-purpose choice for HPLC and GC; multiple injections; storage |
| PTFE / silicone / PTFE | Well | About -40 to 200 °C | Coated on both sides for the most demanding trace work and for internal standards |
| PTFE / natural or synthetic red rubber | Moderately | About -40 to 110 °C | Economical routine GC and HPLC with a single injection |
| PTFE / butyl rubber | Moderately | About -40 to 120 °C | Low gas permeability; headspace at moderate temperature |
| PTFE only (thin disc) | Not at all | Wide | The cleanest, most inert option; single injection soon after sealing |
| Silicone / aluminium foil | Poorly | High | Headspace at high temperature |
| Fluoroelastomer | Poorly | High | Chlorinated solvents and other aggressive samples |
Temperature limits vary by manufacturer and by cap. Use the figures given for the product.
Pre-slit or closed
A pre-slit septum has a cut through the centre.
- It lets air into the vial as sample is drawn out, so no vacuum forms and injection volumes stay constant over repeated injections.
- It offers little resistance, protecting fine or blunt needles from bending and reducing the risk of coring.
- But it seals less well: solvent evaporates faster, so it is not for volatile samples, long sequences or storage.
A closed (non-slit) septum seals best and is the choice for volatile solvents, GC, headspace and storage.
Hardness and thickness
Septum hardness is given in Shore A: the lower the number, the softer the material. Product descriptions often state both, for example 45° Shore A and 1.0 mm.
- Softer and thinner septa are easier to pierce, and are kinder to thin needles.
- Harder and thicker septa seal more firmly and stand up to repeated piercing and higher pressure.
Match them to the autosampler: a needle that bends, or a septum pushed into the vial, means the septum is too hard or too thick for that needle.
Bonded septa
In a bonded cap the septum is fixed to the cap. It cannot be pushed into the vial by the needle or fall out during handling, which is valuable with automated cap handling and with blunt HPLC needles.
Contamination from the septum
- Silicone bleed. Silicone releases cyclic siloxanes, seen in GC-MS as regularly spaced peaks. They increase with temperature and when solvent reaches the silicone after piercing.
- Extractables. Solvents can extract plasticisers and other compounds from rubber. Chlorinated solvents and other strong solvents swell silicone and rubber.
- Fragments. A damaged needle cuts cores out of the septum, which fall into the sample.
To limit all three: choose a PTFE-faced septum suited to the solvent, do not leave pierced vials standing for long, do not reuse septa, and run blanks from a sealed vial of pure solvent.
Quick guide
| Situation | Choose |
|---|---|
| Routine HPLC, several injections per vial | PTFE/silicone, pre-slit |
| Volatile solvents; GC; long sequences | PTFE/silicone, closed |
| Trace analysis by GC-MS or LC-MS | PTFE/silicone/PTFE or PTFE-only, certified low-bleed |
| Headspace GC | PTFE/silicone or PTFE/butyl rated for the oven temperature |
| Chlorinated or very aggressive solvents | PTFE-only, or fluoroelastomer |
| PFAS analysis | Polyethylene or polypropylene caps without PTFE |
| Storage of samples and standards | PTFE/silicone, closed, in a screw or crimp cap |
The septum on the GC inlet
The septum that closes the GC inlet is a different part: a thick, high-temperature, low-bleed disc pierced at every injection. It wears and must be replaced regularly, before it leaks or sheds particles into the liner. A leaking inlet septum shows as shifting retention times, poor repeatability and ghost peaks.
The seal also depends on the cap being applied properly: see correct crimping and vial selection.
General guidance from Lancetech Limited. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.