Lancetech Limited

Vials, caps and consumables

Selection of the correct sample septum

How to choose a vial septum: PTFE/silicone, PTFE/rubber, butyl and PTFE-only compared, pre-slit or closed, hardness and thickness, temperature and bleed.

What a septum has to do

The septum is the disc in the cap that the autosampler needle pierces. It has to seal the vial against evaporation and contamination, let the needle through without damaging it, close again afterwards if more injections follow, and add nothing to the sample. No single material does all of this best, so the choice depends on the solvent, the temperature and how often the vial is pierced. Browse caps and septa.

How a septum is built

Most septa are a layer of rubber or silicone, which gives the seal and the ability to close again, faced with a thin layer of PTFE on the side towards the sample. The PTFE is the chemical barrier. Once the needle has pierced it, the sample or its vapour can reach the elastic layer behind, which matters for the second and later injections.

Cross-section of a vial cap with a PTFE-faced silicone septum, closed and pre-slitClosed septumBest seal: volatile samples,GC, headspace, storagePre-slit septumNo vacuum on repeat injections;more evaporationSilicone: resealsPTFE face,towards the sampleVial neck
A PTFE-faced silicone septum in its cap. The PTFE faces the sample; the silicone closes again after the needle is withdrawn.

The common materials

SeptumReseals after piercingTypical temperature rangeUse
PTFE / siliconeVery wellAbout -40 to 200 °CThe general-purpose choice for HPLC and GC; multiple injections; storage
PTFE / silicone / PTFEWellAbout -40 to 200 °CCoated on both sides for the most demanding trace work and for internal standards
PTFE / natural or synthetic red rubberModeratelyAbout -40 to 110 °CEconomical routine GC and HPLC with a single injection
PTFE / butyl rubberModeratelyAbout -40 to 120 °CLow gas permeability; headspace at moderate temperature
PTFE only (thin disc)Not at allWideThe cleanest, most inert option; single injection soon after sealing
Silicone / aluminium foilPoorlyHighHeadspace at high temperature
FluoroelastomerPoorlyHighChlorinated solvents and other aggressive samples

Temperature limits vary by manufacturer and by cap. Use the figures given for the product.

Pre-slit or closed

A pre-slit septum has a cut through the centre.

  • It lets air into the vial as sample is drawn out, so no vacuum forms and injection volumes stay constant over repeated injections.
  • It offers little resistance, protecting fine or blunt needles from bending and reducing the risk of coring.
  • But it seals less well: solvent evaporates faster, so it is not for volatile samples, long sequences or storage.

A closed (non-slit) septum seals best and is the choice for volatile solvents, GC, headspace and storage.

Hardness and thickness

Septum hardness is given in Shore A: the lower the number, the softer the material. Product descriptions often state both, for example 45° Shore A and 1.0 mm.

  • Softer and thinner septa are easier to pierce, and are kinder to thin needles.
  • Harder and thicker septa seal more firmly and stand up to repeated piercing and higher pressure.

Match them to the autosampler: a needle that bends, or a septum pushed into the vial, means the septum is too hard or too thick for that needle.

Bonded septa

In a bonded cap the septum is fixed to the cap. It cannot be pushed into the vial by the needle or fall out during handling, which is valuable with automated cap handling and with blunt HPLC needles.

Contamination from the septum

  • Silicone bleed. Silicone releases cyclic siloxanes, seen in GC-MS as regularly spaced peaks. They increase with temperature and when solvent reaches the silicone after piercing.
  • Extractables. Solvents can extract plasticisers and other compounds from rubber. Chlorinated solvents and other strong solvents swell silicone and rubber.
  • Fragments. A damaged needle cuts cores out of the septum, which fall into the sample.

To limit all three: choose a PTFE-faced septum suited to the solvent, do not leave pierced vials standing for long, do not reuse septa, and run blanks from a sealed vial of pure solvent.

Quick guide

SituationChoose
Routine HPLC, several injections per vialPTFE/silicone, pre-slit
Volatile solvents; GC; long sequencesPTFE/silicone, closed
Trace analysis by GC-MS or LC-MSPTFE/silicone/PTFE or PTFE-only, certified low-bleed
Headspace GCPTFE/silicone or PTFE/butyl rated for the oven temperature
Chlorinated or very aggressive solventsPTFE-only, or fluoroelastomer
PFAS analysisPolyethylene or polypropylene caps without PTFE
Storage of samples and standardsPTFE/silicone, closed, in a screw or crimp cap

The septum on the GC inlet

The septum that closes the GC inlet is a different part: a thick, high-temperature, low-bleed disc pierced at every injection. It wears and must be replaced regularly, before it leaks or sheds particles into the liner. A leaking inlet septum shows as shifting retention times, poor repeatability and ghost peaks.

The seal also depends on the cap being applied properly: see correct crimping and vial selection.

General guidance from Lancetech Limited. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.