  1. Start with the instrument
  2. Needle tip styles
  3. Needle gauge
  4. Needle length
  5. Fixed or removable needle
  6. Plunger types
  7. Choosing the volume
  8. Injection technique for manual work
  9. Care

Start with the instrument

A syringe for an autosampler must match that autosampler: barrel diameter, flange, overall length and needle length are fixed by the injector tower and the inlet. Check the instrument's requirements first, then choose among the options below. For manual injection you have more freedom.

Needle tip styles

Tip styleShapeUse
BevelledCut at an angle to a sharp pointPiercing septa; the usual tip for manual GC injection
Cone (conical)Symmetrical point with the opening at the tipAutosampler injection; parts the septum instead of cutting it, so the septum lasts longer
Side holeClosed, rounded or conical tip with the opening in the sideHeadspace and gas sampling, and where septum coring is a problem; the hole cannot be plugged by a septum fragment
Blunt (square cut)Flat endHPLC injection valves and sample handling; not for piercing septa

A blunt needle pushed through a septum cuts a core out of it. The fragment ends up in the liner or the vial and causes ghost peaks and leaks.

Four syringe needle tip styles: bevelled, cone, side hole and bluntNeedle tips, magnified. The opening is shown in white.BevelledPiercing septa by handConeAutosamplersSide holeHeadspace and gasesBluntHPLC valves, not septa
Needle tip styles. Only pointed tips are meant to pierce a septum; a blunt needle cuts a core out of it.

Needle gauge

The gauge states the needle's outer diameter: the higher the number, the thinner the needle.

  • 26s gauge (about 0.47 mm) is the common general-purpose size for GC.
  • 23 gauge (about 0.64 mm) is sturdier and bends less, and is required by some inlet seals and septumless inlets.
  • Tapered needles combine a thicker, stronger shaft with a thinner tip, giving durability with less septum wear.
  • On-column injection needs a very fine needle that fits inside the column bore.

A thinner needle is kinder to the septum; a thicker one survives more injections without bending. The inlet determines which sizes are acceptable.

Needle length

The needle must reach the correct depth in the liner, so that the sample is released in the hot zone and not on the septum or at the base of the inlet. Use the length specified for your inlet and autosampler.

Fixed or removable needle

  • Fixed (cemented) needles have the lowest internal volume and suit small volumes. The cement limits the temperature and some solvents.
  • Removable needles can be replaced when bent or blocked, which saves the barrel, and can be exchanged for another tip style.

Plunger types

  • Standard liquid syringes have a metal plunger hand-fitted to its glass barrel. Plungers are not interchangeable between syringes.
  • Gas-tight syringes have a PTFE-tipped plunger that seals against the barrel. Use them for gases, headspace and volatile liquids. The plunger tip can be replaced.

Choosing the volume

Pick a syringe whose capacity is not far above the volume you inject. Accuracy and repeatability fall away when only a small fraction of the barrel is used. A 10 µL syringe is the usual choice for 1 µL injections, and 5 µL or smaller syringes for sub-microlitre volumes.

Injection technique for manual work

  1. Rinse the syringe several times with solvent, then with sample.
  2. Pump the plunger with the needle in the sample to expel air bubbles.
  3. Draw the sample, then a little air, so the sample sits in the barrel and not in the needle.
  4. Pierce the septum, inject smoothly and quickly, and withdraw.

Do the same thing each time. Variation in technique is the main source of scatter in manual injections.

Care

  • Rinse immediately after use with a solvent that dissolves the sample, then with a volatile solvent. Dried residue seizes plungers.
  • Never force a stuck plunger, and avoid touching it with bare fingers.
  • Do not heat cemented-needle syringes above their stated limit.
  • Replace a syringe or needle that is bent, barbed or gives poor repeatability, and replace the inlet septum regularly (septum selection).
  • Store syringes in their case.

The syringe, the septum and the liner work together: a damaged needle tip ruins septa, and septum fragments in the liner spoil the chromatography.

General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.