The principle
In the syringe injection system a heated, gas-tight syringe takes a portion of gas from the headspace vial and injects it into the GC inlet, much as a liquid autosampler injects a liquid. It is the simplest of the three headspace sampling techniques; the others are the balanced pressure system and the pressurised loop system.
The three steps
Step 1: equilibration
The sealed vial is heated in an incubator, usually with agitation, until the volatile compounds have distributed between the sample and the gas phase. Temperature and time are the same for every vial.
Step 2: sampling
The syringe, held at a temperature above that of the vial, pierces the septum and draws up a set volume of headspace gas. Some systems first pressurise the vial or pump the plunger to mix the gas.
Step 3: injection
The syringe moves to the GC inlet and injects the gas through the inlet septum, normally in split mode with a narrow liner. Afterwards the syringe is flushed with inert gas to remove any trace of the sample before the next vial.
- EquilibrateVial incubated with agitation
- SampleHeated syringe draws a set volume of headspace
- InjectInto the GC inlet; syringe is then flushed
Advantages of a syringe injection system
- Simplicity. There is no transfer line, valve or loop to heat, maintain or keep inert.
- Flexibility. The injection volume is changed in the method, without changing hardware. The same autosampler can usually perform liquid injection, headspace and solid phase microextraction by exchanging the tool.
- Low carry-over, because the syringe is purged with gas after each injection and has little surface.
- Standard GC inlet. The sample enters through the normal split/splitless inlet, so no special interface is needed.
Points to watch
- Condensation in the syringe. If the syringe is cooler than the vial, less volatile compounds and water condense on the barrel and needle. Set the syringe temperature above the vial temperature, by a margin the manufacturer recommends, and keep within the syringe's own temperature limit.
- Pressure loss on withdrawal. The heated vial is above atmospheric pressure. When the needle leaves the septum, the gas in the syringe is briefly open to the atmosphere and can expand out of the needle, so some sample is lost. The loss is small and repeatable if the timing is constant, but it makes the technique slightly less precise for very volatile compounds than the closed systems. Keeping the needle in the vial for a short, fixed time after filling, and moving quickly to the inlet, limits it.
- The syringe is a wearing part. The plunger seal and needle wear with use. Falling peak areas or worsening repeatability are the signs. Replace the plunger tip or the syringe (GC syringes and needle tips).
- Two septa are pierced: the vial septum and the inlet septum. A side-hole needle reduces coring. Replace the inlet septum regularly and use vial septa that seal well around the needle (septum selection).
- Inlet settings. Use a narrow-bore liner for gas injections, a suitable split ratio, and an injection speed fast enough to give a sharp band (choosing a liner).
Method settings that matter
| Setting | What it controls |
|---|---|
| Incubation temperature, time and agitation | The headspace concentration at equilibrium |
| Syringe temperature | Freedom from condensation |
| Sample volume and fill speed | The amount taken, and how representative it is |
| Delay before withdrawal and injection speed | Pressure loss and band width |
| Syringe flush time | Carry-over |
| Split ratio and liner | Peak shape and sensitivity |
When to choose it
Syringe sampling is a good fit for laboratories that need one autosampler for several techniques, for method development where injection volume is varied, and for routine analyses of moderately volatile compounds. For the highest precision with very volatile analytes, or for compounds that condense easily, a loop or balanced pressure sampler has the advantage. Whichever system is used, the underlying behaviour of the vial is the same: see basic principles of headspace analysis.