One of three ways to sample the headspace
Once a headspace vial has reached equilibrium, a portion of its gas has to be moved to the GC column without changing its composition. Three techniques are in common use: the balanced pressure system described here, the pressurised loop system and the syringe injection system.
The balanced pressure system uses neither a syringe nor a sample loop. The vial is pressurised with carrier gas, and its own pressure then pushes headspace gas straight onto the column for a set time.
The sequence, step by step
- Equilibration. The sealed vial is held at a constant temperature in the oven until the volatile compounds have distributed between sample and gas.
- Pressurisation. A heated needle pierces the septum. Carrier gas flows into the vial and raises its pressure to the pressure at the head of the GC column, or to a set value above it. The vial is given a short time to stabilise.
- Sampling (transfer). A valve shuts off the normal carrier gas supply to the column. With the supply closed, the pressurised vial becomes the source of gas: headspace flows out through the needle and the heated transfer line directly onto the column. This continues for a precisely timed interval, usually a few seconds.
- Withdrawal. The valve reopens, the normal carrier flow resumes and the needle is withdrawn. The needle is purged before the next vial.
The quantity transferred is set by time and by the pressure and flow conditions, not by a fixed volume.
- EquilibrateVial at constant temperature
- PressuriseCarrier gas raises the vial to column head pressure
- TransferCarrier supply closed: headspace flows to the column for a timed interval
- WithdrawCarrier flow resumes; needle is purged
Strengths
- A short, simple sample path. The headspace goes from vial to column through one heated needle and transfer line, with no loop, valve rotor or syringe barrel for compounds to adsorb on or condense in.
- Good repeatability, because the transfer is controlled by time and pressure, both of which an instrument regulates very precisely.
- Low carry-over, as the needle and line are continuously purged with carrier gas between samples.
- No pressure release before transfer. The vial is never vented to atmosphere, so the headspace composition is not disturbed by a sudden expansion.
Points to watch
- The injected amount is not a fixed volume. It depends on sampling time, vial pressure and column flow. Changing the column, its dimensions or the head pressure changes the amount transferred, so those settings are part of the method.
- Pressure in the vial must be reproducible. A leaking septum or a badly crimped cap gives a lower vial pressure and a smaller peak. This technique is therefore particularly dependent on good crimping and on a septum that seals reliably around the needle at the method temperature.
- Vials with a high natural pressure. If the sample generates a pressure above the set pressurisation value, for example a volatile solvent at high temperature, gas flows the wrong way during pressurisation. Lower the temperature or raise the pressurisation pressure.
- Needle and transfer line temperature must be at least as high as the vial temperature, to prevent condensation.
Method settings that matter
| Setting | What it controls |
|---|---|
| Vial temperature and equilibration time | The headspace concentration at equilibrium |
| Pressurisation pressure and time | The starting pressure for transfer |
| Sampling (injection) time | The amount transferred to the column |
| Needle and transfer line temperature | Freedom from condensation and carry-over |
| Column head pressure | The flow during transfer, and so the amount |
Keep all of them identical for standards and samples.
Consumables
The technique places the same demands on the vial as any headspace method: pressure-rated headspace vials, caps applied with the correct force, and septa rated for the temperature. The background is in basic principles of headspace analysis.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.