Why the crimp matters

A crimp cap is an aluminium or steel shell holding a septum. The crimper folds the skirt of the cap under the rim of the vial neck, pressing the septum onto the glass. Done properly, it is the tightest closure available for a vial, which is why it is used for volatile samples, for gas chromatography and for headspace work. Done badly, it leaks, and the leak shows up as drifting results, lost solvent or, in headspace analysis, low and erratic peak areas.

Types of crimper

Manual crimpers are hand tools like pliers. An adjusting screw or stop sets how far the jaws close, and so how tight the crimp is. They are inexpensive and reliable, but the result depends on the operator squeezing fully each time, and large numbers of vials are tiring.

Electronic crimpers are battery or mains powered. The crimping force is set and then applied identically at the press of a button. They give consistent results across operators, reduce strain, and are worth having when many vials are sealed every day.

Decappers remove crimp caps cleanly and safely. Do not lever caps off with other tools: the vial neck can break.

Each tool fits one cap size. The common sizes are 8 mm, 11 mm, 13 mm and 20 mm; an 11 mm crimper will not seal a 20 mm headspace cap. Browse crimpers and decappers.

How to crimp

  1. Place the filled vial on a flat, firm surface and set the cap on it, with the septum's PTFE face towards the sample.
  2. Lower the crimper squarely over the cap until it rests on it. Keep it vertical.
  3. Squeeze the handles fully in one smooth movement (or trigger the electronic tool), then release.
  4. Check the crimp, as below.

Correct crimping

A good crimp shows all of these:

  • The cap cannot be turned by hand, or only with real effort.
  • The skirt is folded evenly under the neck all the way round, with no gap and no sharp wrinkles.
  • The top of the cap is flat, and the septum is flat or only very slightly depressed, not bulging.
  • The cap sits level, not tilted.

Check the first vials of every batch, and again after changing the cap or vial type. Slight differences between makes of cap and vial are enough to need a new setting.

Three crimped vial caps in cross-section: correct, too loose and too tightCorrect✓Flat top, skirt tucked evenlyunder the rim, cap cannot turnToo loose✕Skirt not folded under,cap turns: the vial leaksToo tight✕Cap dished, septum bulging:poor seal, needle damageSeptumCap skirtVial neck
A crimp is right when the cap cannot be turned, the skirt is tucked evenly under the rim and the top stays flat.

Possible errors

ErrorWhat you seeCauseRemedy
1. Cap edges not grippingThe cap turns easily or lifts; the skirt stands away from the neckCrimp too looseTighten the adjustment in small steps
2. Curved or dished capThe top of the cap is bowed or pushed inCrimp too tightLoosen the adjustment
3. Deformed cap sidesThe sides are creased or buckledToo tight, or the tool was not held squareLoosen; keep the tool vertical
4. Bulging septumThe septum domes up through the hole, or is squeezed outMuch too tightLoosen; a bulging septum is easily cored by the needle
5. Uneven or rounded edgeThe skirt is folded on one side only, or the cap is tiltedTool tilted, worn jaws, or a cap and vial that do not matchHold square; service or replace the jaws; use matching parts

A crimp that is too tight is as bad as one that is too loose. Over-crimping deforms the septum so that it seals poorly around the needle, can cause needle damage and coring, and can crack the vial neck.

Adjusting the tool

Change the setting a little at a time and test after each change. On a manual crimper, set the stop so that the handles close fully at exactly the right tightness, then squeeze to the stop every time. Jaws wear with use: if a good crimp can no longer be achieved by adjustment, the jaws need replacing.

Headspace vials

Headspace vials are heated and pressurised, so the seal is tested harder than in any other use. Use 20 mm caps and septa rated for the method temperature, crimp so the cap cannot be turned, and never reuse a cap. Autosamplers that move vials magnetically need steel (magnetic) caps. Where the sample generates high pressure, caps with a pressure-release feature are available. See basic principles of headspace analysis.

Safety and housekeeping

  • Inspect vial necks: a chipped rim will not seal and may break in the crimper.
  • Wear eye protection and suitable gloves when crimping and decapping.
  • Dispose of decapped caps and broken glass in a sharps or glass container.
  • Keep the tool clean; glass fragments and aluminium dust in the jaws damage the next cap.

When a screw cap is the better choice

For non-volatile samples and routine HPLC, screw-cap vials seal well enough and need no tool. Crimp caps earn their place with volatile solvents, long storage and headspace analysis. See vial selection and septum selection.

General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.