The idea
In static headspace analysis the sample is sealed in a vial and warmed. Volatile compounds leave the sample and collect in the gas above it, the headspace. Once the distribution has settled, a portion of that gas is transferred to the gas chromatograph. The non-volatile bulk of the sample (water, salts, polymer, biological matrix) never enters the instrument.
The two phases in the vial
- The sample phase (also called the condensed phase): the liquid or solid sample, with any diluent or matrix modifier.
- The gas phase: the headspace above it.
At a constant temperature each volatile compound distributes itself between the two until the rate of leaving the sample equals the rate of returning. This is equilibrium. From then on, the concentration in the gas is in a fixed proportion to the concentration in the sample, which is what makes the measurement quantitative.
The equation that governs sensitivity
Two quantities describe the vial:
- The partition coefficient K = concentration in the sample phase / concentration in the gas phase, at equilibrium. A high K means the compound prefers the sample. See partition coefficient.
- The phase ratio beta = volume of gas phase / volume of sample phase. See headspace sample volume.
The concentration in the headspace (Cg) follows from the original concentration in the sample (C0):
Cg = C0 / (K + beta)
To get more analyte into the gas, and so a larger peak, make K smaller, beta smaller, or both.
What you can adjust
| Setting | Effect |
|---|---|
| Temperature | Higher temperature lowers K for most compounds, especially polar compounds in water |
| Equilibration time | Must be long enough to reach equilibrium, and the same for every vial |
| Shaking or agitation | Shortens the time to equilibrium |
| Sample volume | Changes beta; matters most for compounds with a low K |
| Salt addition | "Salting out" lowers K for polar compounds in aqueous samples |
| pH | Acids and bases must be in their uncharged form to be volatile |
| Matrix or diluent | A high-boiling solvent that dissolves the sample completely gives a uniform matrix |
Find the equilibration time by experiment: analyse vials after increasing times and use a time safely on the plateau.
Vial and septum issues
The vial is the measuring vessel, so its integrity decides the result.
- Vial size. 10 and 20 mL are the common sizes. The vial must match the autosampler.
- Vial quality. Use headspace vials, which have a thicker, uniform wall, and often a rounded bottom, to withstand pressure and heat. Do not reuse them.
- Closure. Aluminium crimp caps, magnetic crimp caps or screw caps for headspace. A loose cap leaks; an over-crimped one deforms the septum. See correct crimping.
- Septum material. Choose by temperature and by the cleanliness required. PTFE-faced silicone tolerates higher temperatures than PTFE-faced butyl rubber; butyl is less permeable but softens earlier. See selecting a septum.
- Septum bleed. Volatile compounds from the septum itself appear as background peaks, more so at higher temperature. Run blanks with empty sealed vials.
- Pressure. Heating raises the pressure in the vial, especially with water or volatile solvents. Stay below the solvent's boiling point where possible, and use pressure-release caps if the manufacturer recommends them.
- One puncture only. After the needle has pierced the septum the seal is no longer reliable, so a vial should not be stored and re-analysed, unless the method is validated for it.
Static and dynamic headspace
This page describes static headspace: one aliquot from a closed vial at equilibrium. In dynamic headspace (purge and trap) a stream of gas continuously sweeps volatiles out of the sample onto an adsorbent trap, which is then heated to release them. It is more sensitive, and more complex. Solid phase microextraction from the headspace is a third option, using a coated fibre to collect the analytes.
How the gas reaches the GC
Three sampling techniques are in use: the balanced pressure system, the pressurised loop system and the syringe injection system. Applications are covered in headspace analysis in GC.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.