The phase ratio
The amount of sample in a headspace vial sets the phase ratio (beta): the volume of gas divided by the volume of sample.
beta = Vg / Vs
For a 20 mL vial:
| Sample volume | Gas volume | Phase ratio (beta) |
|---|---|---|
| 1 mL | 19 mL | 19 |
| 2 mL | 18 mL | 9 |
| 5 mL | 15 mL | 3 |
| 10 mL | 10 mL | 1 |
Nominal vial sizes are approximate: the true internal volume of a "20 mL" vial is a little more. For precise work, use the actual volume given by the vial manufacturer.
How beta affects the result
The concentration of an analyte in the headspace at equilibrium is
Cg = C0 / (K + beta)
where C0 is the concentration in the sample and K is the partition coefficient. Whether sample volume matters depends on how K compares with beta.
When K is small (the compound prefers the gas phase, such as a hydrocarbon in water), beta dominates the sum. Taking hexane in water at 40 °C, with K of about 0.14:
- 2 mL of sample (beta = 9): Cg = C0 / 9.14
- 10 mL of sample (beta = 1): Cg = C0 / 1.14
The larger sample gives a peak about eight times bigger.
When K is large (the compound prefers the sample, such as ethanol in water at 40 °C, with K of about 1,355), beta hardly counts:
- 2 mL of sample: Cg = C0 / 1,364
- 10 mL of sample: Cg = C0 / 1,356
The difference is under 1%. For such compounds, more sample does not help; raising the temperature or adding salt does.
Practical consequences
- For low-K analytes, sample volume is a strong lever on sensitivity, and also a strong source of error: a small pipetting difference between vials changes the result. Pipette accurately.
- For high-K analytes, the result is robust against volume differences, so volume precision matters less.
- In every case, use the same volume for standards and samples.
How full can the vial be?
As a working rule, fill the vial no more than about half full. The headspace must be large enough for the sampling needle to stay clear of the liquid, for a representative aliquot to be taken, and to keep the pressure rise manageable when the vial is heated. Too little gas volume also means that withdrawing an aliquot disturbs the equilibrium noticeably.
Solid samples
For solids the sample "volume" is hard to define, and the analyte may be released only slowly from inside the particles. Keep the weight, particle size and filling pattern the same from vial to vial. Where possible, dissolve or disperse the solid in a suitable high-boiling solvent or in water, which turns it into a liquid sample with a defined phase ratio. Otherwise consider multiple headspace extraction (headspace analysis in GC).
Volume and equilibration time
A larger sample takes longer to reach equilibrium, because the analyte has further to travel to the surface. If you change the sample volume, check the equilibration time again. Agitation shortens it.
Choosing the vial size
A smaller vial with the same sample gives a lower phase ratio, which helps low-K analytes. But the vial must fit the autosampler and leave enough headspace for sampling. Most methods use 20 mL vials with 1 to 10 mL of sample, or 10 mL vials with proportionally less. See vial selection.
General guidance from Lancetech Limited. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.