Gas chromatography and headspace

Partition coefficient in headspace analysis

The partition coefficient K in headspace GC: definition, K values of common solvents in water, and how temperature, salt, pH and phase ratio change sensitivity.

Definition

When a volatile compound is sealed in a vial with a sample and a gas space, it divides itself between the two. At equilibrium, the partition coefficient (K) is the ratio of its concentrations:

K = Cs / Cg

Cs is the concentration in the sample phase and Cg the concentration in the gas phase.

  • A low K means the compound readily leaves the sample for the gas: it is easy to measure by headspace.
  • A high K means the compound prefers to stay in the sample: its headspace concentration is low.

K is not a fixed property of the compound alone. It depends on the temperature and on the matrix the compound is dissolved in.

K values of common solvents in water

Approximate values for the air and water system at 40 °C, as widely quoted in the headspace literature:

CompoundK at 40 °C
Cyclohexane0.08
n-Hexane0.14
Tetrachloroethylene1.5
Toluene2.8
Benzene2.9
Dichloromethane5.7
Ethyl acetate62
Methyl ethyl ketone140
n-Butanol650
2-Propanol830
Ethanol1,355
1,4-Dioxane1,620

The pattern is clear. Non-polar compounds dislike water and escape into the gas (low K). Polar compounds that mix with water and form hydrogen bonds stay in it (high K).

Calculating the headspace concentration

The partition coefficient and the phase ratio (beta, gas volume divided by sample volume) together give the concentration in the gas:

Cg = C0 / (K + beta)

C0 is the original concentration in the sample. The smaller the sum of K and beta, the larger the signal. For a low-K compound, beta dominates and sample volume matters. For a high-K compound, K dominates, and the way to more signal is to lower K.

Lowering the partition coefficient

Raise the temperature

For most compounds K falls as temperature rises, because vapour pressure increases. The effect is strongest for high-K compounds: the K of ethanol in water falls several-fold between 40 and 80 °C. For low-K compounds temperature makes little difference.

The limits are practical: the pressure in the vial, the stability of sample and analytes, and the temperature rating of the septum. Keep the temperature constant to within a fraction of a degree, since for high-K compounds a small temperature error becomes a visible error in the result.

Add salt

Dissolving an inorganic salt in an aqueous sample makes the water a poorer solvent for organic compounds and pushes them into the gas. This "salting out" works best for polar compounds with a high K, and has little effect on non-polar ones. Sodium chloride, sodium sulfate, potassium carbonate and ammonium sulfate are commonly used, typically at or near saturation so that every vial has the same salt content.

Adjust the pH

Only the uncharged form of an acid or base is volatile. Acidify to measure volatile acids; make the sample basic to measure volatile amines.

Change the matrix

The same compound has a different K in a different liquid. Adding water to a water-miscible organic sample, or choosing a different diluent, can raise the headspace concentration of non-polar analytes considerably. Whatever is done to the samples must be done identically to the standards.

Combining the partition coefficient and the phase ratio

AnalyteBest ways to increase signal
Low K (non-polar compounds in water)More sample in the vial (lower beta); temperature and salt help little
High K (alcohols and other polar compounds in water)Higher temperature and salt; sample volume helps little
Intermediate KBoth approaches contribute

A note on other partition coefficients

The same term is used elsewhere in chemistry for the distribution of a compound between two liquids. The octanol-water partition coefficient (log P or log Kow) describes how a compound divides between octanol and water and is used to predict retention in reversed-phase chromatography and behaviour in solid phase extraction. It is a different quantity from the headspace K described here.

General guidance from Lancetech Limited. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.