How to use this page
Plastics differ widely in what they tolerate. A bottle that stores dilute acid for years may soften, crack or dissolve in an organic solvent within minutes. The tables below give a general orientation at room temperature. They are not a guarantee: resistance also depends on concentration, temperature, contact time and mechanical stress, and differs between grades of the same plastic. For any critical or long-term use, check the manufacturer's data for the exact product and chemical, and test under your own conditions.
The plastics
| Abbreviation | Plastic | Typical laboratory use |
|---|---|---|
| LDPE | Low-density polyethylene | Wash bottles, dropping bottles, flexible containers |
| HDPE | High-density polyethylene | Bottles, carboys, waste containers |
| PP | Polypropylene | Tubes, pipette tips, beakers, bottles, syringe filter housings, vials |
| PMP | Polymethylpentene | Clear, heat-resistant measuring cylinders and beakers |
| PC | Polycarbonate | Clear, tough vessels, centrifuge ware, desiccators |
| PS | Polystyrene | Clear disposable ware: Petri dishes, culture plates, cuvettes |
| PVC | Polyvinyl chloride | Tubing, some bottles |
| PET | Polyethylene terephthalate | Clear bottles for media and samples |
| PMMA | Polymethyl methacrylate (acrylic) | Cuvettes, shields, tanks |
| PTFE, FEP, PFA | Fluoropolymers | Stirrer bars, tubing, liners, bottles for aggressive chemicals |
| PEEK | Polyether ether ketone | HPLC tubing and fittings |
Resistance by chemical class at room temperature
Good = suitable for continuous contact. Limited = short contact only; some swelling, softening or loss of strength. Poor = not suitable.
| Chemical class | LDPE | HDPE | PP | PMP | PC | PS | PVC | PET | PTFE / FEP / PFA |
|---|---|---|---|---|---|---|---|---|---|
| Dilute acids | Good | Good | Good | Good | Good | Good | Good | Good | Good |
| Concentrated non-oxidising acids | Good | Good | Good | Good | Poor | Limited | Limited | Poor | Good |
| Oxidising acids (concentrated nitric, sulfuric) | Poor | Limited | Poor | Poor | Poor | Poor | Poor | Poor | Good |
| Bases (alkalis) | Good | Good | Good | Good | Poor | Good | Good | Poor | Good |
| Alcohols | Good | Good | Good | Good | Limited | Limited | Good | Limited | Good |
| Aliphatic hydrocarbons (hexane, heptane) | Poor | Limited | Limited | Limited | Limited | Poor | Limited | Good | Good |
| Aromatic hydrocarbons (toluene, xylene) | Poor | Limited | Limited | Poor | Poor | Poor | Poor | Poor | Good |
| Halogenated solvents (dichloromethane, chloroform) | Poor | Limited | Limited | Poor | Poor | Poor | Poor | Poor | Good |
| Ketones (acetone, methyl ethyl ketone) | Limited | Limited | Limited | Limited | Poor | Poor | Poor | Poor | Good |
| Esters (ethyl acetate) | Limited | Good | Limited | Limited | Poor | Poor | Poor | Poor | Good |
| Ethers (diethyl ether, tetrahydrofuran) | Poor | Limited | Limited | Poor | Poor | Poor | Poor | Poor | Good |
| Aqueous salt solutions and buffers | Good | Good | Good | Good | Good | Good | Good | Good | Good |
Other materials in brief:
- PMMA (acrylic) tolerates dilute acids, dilute alkalis and aliphatic hydrocarbons, but is attacked by alcohols (which cause fine cracking), ketones, esters, aromatic and halogenated solvents.
- PEEK resists almost all common solvents, acids and bases. The exceptions are concentrated nitric and sulfuric acids, and swelling in tetrahydrofuran, dichloromethane and dimethyl sulfoxide (HPLC tubing).
- Fluoropolymers resist essentially everything in routine laboratory use.
Patterns worth remembering
- Polyolefins (PE, PP, PMP) are the workhorses for aqueous solutions, acids and bases. Their weakness is non-polar and chlorinated organic solvents, which make them swell, and strong oxidising agents, which embrittle them over time.
- Clear, rigid plastics (PC, PS, PMMA) are the most easily damaged by organic solvents. Polycarbonate is also attacked by alkalis, including alkaline cleaning agents, and by ammonia.
- Strong oxidising acids belong in glass or a fluoropolymer, not in other plastics.
- Ketones differ. Polyolefins tolerate acetone better than other ketones, which is why acetone wash bottles are commonly polyethylene; check the chart for the ketone you use.
- Higher temperature lowers resistance. A "limited" rating at 20 °C is often "poor" at 50 °C.
- Stress matters. A part under load, such as a centrifuge tube or a tightened cap, can crack in a chemical that an unstressed piece of the same plastic tolerates.
- Permeation is not attack. Solvents and gases can pass slowly through a polyethylene wall without visibly damaging it, so volatile solvents stored in plastic lose volume, and samples can pick up or lose trace compounds.
- Leaching works the other way. Plastics release additives into solvents, which matters in trace analysis and mass spectrometry.
Temperature limits and autoclaving
| Plastic | Approximate maximum use temperature | Autoclavable at 121 °C |
|---|---|---|
| LDPE | 80 °C | No |
| HDPE | 110 °C | No |
| PP | 135 °C | Yes |
| PMP | 150 °C | Yes |
| PC | 125 °C | Yes, but repeated cycles weaken it |
| PS | 70 °C | No |
| PVC | 70 °C | No |
| PET | 70 °C | No |
| PMMA | 70 °C | No |
| FEP | 205 °C | Yes |
| PTFE, PFA | 260 °C | Yes |
Before autoclaving, loosen or remove caps, and make sure no chemical residue is present that could attack the plastic at high temperature. See autoclave.
Signs of chemical attack
Stop using an item that shows swelling, softening, stickiness, clouding or discolouration, fine surface cracks (crazing), brittleness, or a change in weight or dimensions. In volumetric ware, any swelling means the calibration is lost.
Choosing in practice
- Identify the most aggressive chemical the item will meet, its concentration and the temperature.
- Check the table for orientation, then the manufacturer's resistance chart.
- For organic solvents, default to glass or a fluoropolymer unless the chart confirms otherwise.
- Consider the closure, liner and any tubing as well as the vessel (sample bottles and caps).
- Test before committing a valuable sample or a long storage period.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.