Lancetech Limited

Laboratory practice

Autoclave: how steam sterilisation works

How an autoclave sterilises: 121 °C and 134 °C cycles, why air must be removed, liquid and solid loads, what may be autoclaved, and how to verify a cycle.

What an autoclave is

An autoclave is a pressure vessel that sterilises with saturated steam under pressure. Raising the pressure raises the temperature at which water boils, so the chamber can be filled with steam well above 100 °C. Steam is far more effective than dry air at the same temperature: as it condenses on a cooler surface it releases a large amount of heat and delivers moisture, and together these rapidly destroy the proteins of microorganisms, including resistant bacterial spores.

Autoclaves are used to sterilise culture media, solutions, glassware and instruments before use, and to make microbiological waste safe before disposal.

Temperature, pressure and time

Sterilising temperatureApproximate chamber pressure (above atmospheric)Minimum holding time
121 °CAbout 1 bar (15 psi)15 minutes
134 °CAbout 2 bar (30 psi)3 minutes

The holding time starts only when every part of the load has reached the temperature. A large volume of liquid or a densely packed load takes much longer to heat through than the chamber does, so the total cycle is longer, often considerably. Validated cycles for each type of load are the only reliable guide.

Pressure alone does not sterilise: it is only the means of reaching the temperature.

Why air must be removed

Air trapped in the chamber or inside the load prevents steam from reaching the surfaces, and a mixture of air and steam is cooler than pure steam at the same pressure. Autoclaves remove air in one of two ways:

  • Gravity displacement: steam enters and pushes the heavier air out through a drain. Suitable for liquids and simple, unwrapped items.
  • Pre-vacuum (vacuum-assisted): a pump removes the air before steam is admitted, often in several pulses. Needed for wrapped goods, porous loads, tubing and hollow instruments.

The phases of a cycle

  1. Air removal and heating.
  2. Sterilisation: the holding time at temperature.
  3. Exhaust. For liquids the pressure must be released slowly, otherwise the superheated liquid boils violently and is lost or breaks its container.
  4. Drying and cooling, for wrapped and porous loads.

Always select the programme that matches the load: a liquid cycle for liquids, a solid or porous cycle for instruments and glassware.

An autoclave cycle: temperature against time through air removal, the sterilisation hold and coolingTimeTemperature121 °CAir removaland heatingSterilisation holdat least 15 minExhaustand coolingChamberLoad: heats up laterThe hold is timed from when the load, not only the chamber, reaches 121 °C. Illustrative.
A steam sterilisation cycle. The load reaches temperature after the chamber, which is why the hold is timed from the load.

Loading correctly

  • Loosen caps on bottles, or use vented closures. A sealed container can burst.
  • Fill bottles and flasks to no more than about two-thirds to three-quarters.
  • Stand containers in a tray that will hold a spill.
  • Do not overload: leave space for steam to circulate between items.
  • Place waste in autoclave bags left loosely closed, in a solid secondary container.
  • Put empty vessels on their side or upside down so that air can drain out.

What can and cannot be autoclaved

SuitableNot suitable
Borosilicate glasswareFlammable, volatile or reactive chemicals
Stainless steel instrumentsBleach and other chlorine-releasing or corrosive solutions
Polypropylene, polymethylpentene and fluoropolymer plasticsPolystyrene, polyethylene, PVC and other heat-sensitive plastics
Aqueous media, buffers and waterHeat-sensitive solutions (many vitamins, antibiotics, proteins)
Microbiological waste for decontaminationRadioactive material; sealed containers

For heat-sensitive solutions, sterilise by filtration through a sterile 0.2 µm membrane instead (syringe filter selection). The heat tolerance of each plastic is given in chemical resistance of plastics.

Proving that a cycle worked

CheckWhat it shows
Physical record of temperature, pressure and timeThe chamber reached the set conditions
Chemical indicators (tape, strips, integrators)The item was exposed to the process. Autoclave tape shows exposure only, not sterility
Biological indicators (spores of Geobacillus stearothermophilus)A highly resistant organism was killed: the direct test of sterilisation
Air-removal test (Bowie-Dick type)The vacuum system is removing air properly, in pre-vacuum autoclaves

Use indicators inside the load, at the position hardest for steam to reach, not only on the outside.

Safety

  • An autoclave is a pressure system. In the United Kingdom it must be maintained and examined periodically by a competent person under a written scheme, as required by the Pressure Systems Safety Regulations.
  • Open the door only when the pressure has returned to zero and the load has cooled. Liquids can remain superheated and boil over when moved: let them cool to below about 80 °C first.
  • Wear heat-resistant gloves, a face shield and a lab coat when unloading.
  • Stand clear of the door as it opens; steam escapes first from the top.
  • Only trained staff should operate the autoclave.

Common problems

ProblemLikely cause
Liquid lost or boiled overFast exhaust used for a liquid load; container too full
Wet packs after the cycleOverloading; insufficient drying time; cold spots
Indicator did not changeAir not removed; load too dense; temperature not reached
Glass brokenSealed container; thermal shock; soda-lime glass used
Media browned or precipitatedOver-long heating; ingredients that should have been filter-sterilised

General guidance from Lancetech Limited. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.