What an autoclave is
An autoclave is a pressure vessel that sterilises with saturated steam under pressure. Raising the pressure raises the temperature at which water boils, so the chamber can be filled with steam well above 100 °C. Steam is far more effective than dry air at the same temperature: as it condenses on a cooler surface it releases a large amount of heat and delivers moisture, and together these rapidly destroy the proteins of microorganisms, including resistant bacterial spores.
Autoclaves are used to sterilise culture media, solutions, glassware and instruments before use, and to make microbiological waste safe before disposal.
Temperature, pressure and time
| Sterilising temperature | Approximate chamber pressure (above atmospheric) | Minimum holding time |
|---|---|---|
| 121 °C | About 1 bar (15 psi) | 15 minutes |
| 134 °C | About 2 bar (30 psi) | 3 minutes |
The holding time starts only when every part of the load has reached the temperature. A large volume of liquid or a densely packed load takes much longer to heat through than the chamber does, so the total cycle is longer, often considerably. Validated cycles for each type of load are the only reliable guide.
Pressure alone does not sterilise: it is only the means of reaching the temperature.
Why air must be removed
Air trapped in the chamber or inside the load prevents steam from reaching the surfaces, and a mixture of air and steam is cooler than pure steam at the same pressure. Autoclaves remove air in one of two ways:
- Gravity displacement: steam enters and pushes the heavier air out through a drain. Suitable for liquids and simple, unwrapped items.
- Pre-vacuum (vacuum-assisted): a pump removes the air before steam is admitted, often in several pulses. Needed for wrapped goods, porous loads, tubing and hollow instruments.
The phases of a cycle
- Air removal and heating.
- Sterilisation: the holding time at temperature.
- Exhaust. For liquids the pressure must be released slowly, otherwise the superheated liquid boils violently and is lost or breaks its container.
- Drying and cooling, for wrapped and porous loads.
Always select the programme that matches the load: a liquid cycle for liquids, a solid or porous cycle for instruments and glassware.
Loading correctly
- Loosen caps on bottles, or use vented closures. A sealed container can burst.
- Fill bottles and flasks to no more than about two-thirds to three-quarters.
- Stand containers in a tray that will hold a spill.
- Do not overload: leave space for steam to circulate between items.
- Place waste in autoclave bags left loosely closed, in a solid secondary container.
- Put empty vessels on their side or upside down so that air can drain out.
What can and cannot be autoclaved
| Suitable | Not suitable |
|---|---|
| Borosilicate glassware | Flammable, volatile or reactive chemicals |
| Stainless steel instruments | Bleach and other chlorine-releasing or corrosive solutions |
| Polypropylene, polymethylpentene and fluoropolymer plastics | Polystyrene, polyethylene, PVC and other heat-sensitive plastics |
| Aqueous media, buffers and water | Heat-sensitive solutions (many vitamins, antibiotics, proteins) |
| Microbiological waste for decontamination | Radioactive material; sealed containers |
For heat-sensitive solutions, sterilise by filtration through a sterile 0.2 µm membrane instead (syringe filter selection). The heat tolerance of each plastic is given in chemical resistance of plastics.
Proving that a cycle worked
| Check | What it shows |
|---|---|
| Physical record of temperature, pressure and time | The chamber reached the set conditions |
| Chemical indicators (tape, strips, integrators) | The item was exposed to the process. Autoclave tape shows exposure only, not sterility |
| Biological indicators (spores of Geobacillus stearothermophilus) | A highly resistant organism was killed: the direct test of sterilisation |
| Air-removal test (Bowie-Dick type) | The vacuum system is removing air properly, in pre-vacuum autoclaves |
Use indicators inside the load, at the position hardest for steam to reach, not only on the outside.
Safety
- An autoclave is a pressure system. In the United Kingdom it must be maintained and examined periodically by a competent person under a written scheme, as required by the Pressure Systems Safety Regulations.
- Open the door only when the pressure has returned to zero and the load has cooled. Liquids can remain superheated and boil over when moved: let them cool to below about 80 °C first.
- Wear heat-resistant gloves, a face shield and a lab coat when unloading.
- Stand clear of the door as it opens; steam escapes first from the top.
- Only trained staff should operate the autoclave.
Common problems
| Problem | Likely cause |
|---|---|
| Liquid lost or boiled over | Fast exhaust used for a liquid load; container too full |
| Wet packs after the cycle | Overloading; insufficient drying time; cold spots |
| Indicator did not change | Air not removed; load too dense; temperature not reached |
| Glass broken | Sealed container; thermal shock; soda-lime glass used |
| Media browned or precipitated | Over-long heating; ingredients that should have been filter-sterilised |
General guidance from Lancetech Limited. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.