Lancetech Limited
Laboratory practice

Hygiene monitoring in pharmacies and pharmaceutical settings

Hygiene monitoring in pharmacy aseptic units and cleanrooms: settle plates, contact plates, air samples, glove prints, limits by grade and trending.

  • 4 min read
  • 1 figure
  • Reviewed 25 November 2026
On this page 7 sections
  1. Why monitor
  2. What is monitored
  3. Cleanroom grades and limits
  4. Culture media and incubation
  5. Doing it properly
  6. Reading the results
  7. Supporting measures

Why monitor

Medicines prepared aseptically, such as injections, infusions, parenteral nutrition and eye drops, cannot be sterilised in their final container. Their safety depends entirely on the cleanliness of the environment and the technique of the people preparing them. Hygiene monitoring, more formally environmental and personnel monitoring, measures microbial and particle contamination in the preparation area, to show that the conditions are under control and to give early warning when they are not.

It applies to hospital pharmacy aseptic units, to compounding pharmacies and to pharmaceutical manufacturing. Monitoring does not make a product sterile. It demonstrates that the controls which do, the air filtration, the cleaning, the gowning and the aseptic technique, are working.

What is monitored

SampleMethodWhat it shows
Air, passiveSettle plates: open 90 mm agar plates exposed for up to 4 hoursOrganisms settling out of the air onto surfaces
Air, activeAn air sampler draws a known volume of air onto agarOrganisms per cubic metre of air
SurfacesContact plates (55 mm) pressed onto flat surfaces; swabs for uneven or hard-to-reach surfacesCleanliness of work surfaces and equipment
PersonnelGlove prints: the five fingertips touched onto an agar plate; gown contact platesAseptic technique and gowning
ParticlesA particle counterNon-viable airborne particles
OtherPressure differentials, airflow, temperature, humidityThe physical state of the room

See Petri dishes for the plate types.

Cleanroom grades and limits

Pharmaceutical cleanrooms are classified in four grades. Grade A is the critical zone where sterile product is exposed, for example inside an isolator or a laminar airflow cabinet. Grade B is the background for a grade A zone in open systems. Grades C and D are for less critical steps.

The EU and UK guidance on the manufacture of sterile medicines (Annex 1 of good manufacturing practice) gives maximum action limits for microbial contamination:

GradeAir sample (cfu per m³)Settle plate, 90 mm (cfu per 4 hours)Contact plate, 55 mm (cfu per plate)Glove print, 5 fingers (cfu per glove)
ANo growthNo growthNo growthNo growth
B10555
C1005025Not specified
D20010050Not specified

Each unit sets its own alert and action levels at or below these figures, based on its own history. Pharmacy aseptic services also follow national standards for their sector, and in the United States the compounding chapters of the USP. Work to the current version of the standard that applies to you.

Cleanroom grades A to D nested from the critical zone outwards, with the air sample limits for eachGrade D200 cfu per m³Grade C100 cfu per m³Grade B10 cfu per m³Grade ANo growthCritical zone: sterile product is exposed,for example inside an isolatorMaximum action limits for active air samples (EU and UK good manufacturing practice, Annex 1)
Cleanroom grades surround one another: each cleaner zone is protected by the one outside it.

Culture media and incubation

  • A general medium such as tryptone soya agar grows a wide range of bacteria and fungi and is the usual choice.
  • A selective medium for yeasts and moulds, such as Sabouraud dextrose agar, may be added.
  • Media for surfaces and gloves contain neutralisers, such as lecithin and polysorbate 80, which inactivate residues of disinfectant that would otherwise prevent growth and give a falsely clean result.
  • Plates are incubated at the temperatures and for the times set in the unit's validated procedure, chosen to recover both bacteria and fungi.

Media must be shown to support growth, and plates used inside grade A and B areas are normally irradiated and multiply wrapped so that they can be brought in without introducing contamination.

Doing it properly

  • Follow a written plan that fixes the locations, frequency, timing and methods. Choose locations by risk: nearest the exposed product, where staff work and where contamination would matter most.
  • Monitor during real work ("in operation"), not only when the room is at rest.
  • Sample surfaces and gloves at the end of a session, so that the agar residue is not left on a working surface, and clean the sampled area afterwards.
  • Handle plates aseptically, label them with location, date, time and operator, and transport them closed.
  • Expose settle plates for the defined time, and not so long that the agar dries out.
  • Record everything, including the sessions with no growth.

Reading the results

  1. Count the colonies and compare them with the alert and action levels.
  2. Identify organisms recovered from grade A and B areas, and those above the limits elsewhere. The type of organism points to the source: skin organisms suggest personnel, moulds suggest materials or building fabric, water organisms suggest wet areas.
  3. Trend the results over time. A slow rise, a change in the kinds of organism, or repeated low counts at one location can be more significant than a single high count.
  4. Investigate every excursion: what was being prepared, by whom, the cleaning record, the equipment and the pressure and airflow data. Record the cause and the corrective action.

Supporting measures

  • Cleaning and disinfection to a validated schedule, with rotation of disinfectants including a sporicidal agent.
  • Hand hygiene and gowning, with training and regular assessment of each operator.
  • Operator qualification by process simulation (media fill) tests.
  • Rapid checks such as ATP bioluminescence swabs give immediate feedback on cleaning, although they do not replace culture-based monitoring.
  • Water quality monitoring where water is used.

Consumables for the work include culture plates, swabs, gloves and microbiology equipment such as incubators and air samplers. Used plates are live cultures and must be decontaminated before disposal (autoclave).

General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.