Lancetech Limited
Laboratory practice

Nucleic acid decontamination: removing DNA, RNA and nucleases

How to prevent and remove DNA, RNA and nuclease contamination in a PCR laboratory: bleach, UV, what ethanol and autoclaving cannot do, and workflow design.

  • 3 min read
  • 1 figure
  • Reviewed 25 November 2026
On this page 8 sections
  1. Why it matters
  2. Where contamination comes from
  3. Prevention comes first
  4. What removes nucleic acids from surfaces
  5. Dealing with RNases
  6. Preventing carry-over inside the reaction
  7. A routine that works
  8. If contamination is found

Why it matters

PCR and related methods can detect a handful of copies of a target sequence. The same sensitivity makes them vulnerable: a trace of stray DNA gives a false positive. And a single PCR produces billions of copies of the product (the amplicon), which is by far the most dangerous contaminant in the laboratory, because every copy is a perfect template for the next reaction.

There are two different problems:

  • Contaminating nucleic acids (amplicons, plasmids, genomic DNA, RNA) that cause false positive results.
  • Nucleases (DNases and especially RNases) that destroy the sample and cause false negative or failed results. RNases are everywhere, on skin and dust, and are very stable.

Where contamination comes from

  • Opening tubes that contain amplified product.
  • Aerosols created by pipetting, vortexing and centrifuging.
  • Gloves, sleeves, door handles, pens, phones and notebooks moving between areas.
  • Pipettes, racks and centrifuges shared between pre- and post-amplification work.
  • Reagents and water contaminated in earlier use.
  • Skin, hair and saliva, for human targets and for RNases.

Prevention comes first

Removing contamination is harder than preventing it.

  • Separate the work in space. Use different rooms or at least dedicated, separated areas for reagent preparation, sample preparation, amplification and post-amplification analysis.
  • Work in one direction only, from clean to dirty: reagents, then samples, then amplification, then analysis. Never bring anything back from the post-amplification area.
  • Dedicate equipment to each area: pipettes, racks, centrifuges, lab coats, pens.
  • Use filter tips to keep aerosols out of pipette shafts (types of pipette).
  • Change gloves often, and always when moving between areas.
  • Aliquot reagents into single-use portions.
  • Open tubes carefully and briefly; spin them down first.
  • Use consumables certified free of DNA, DNase and RNase (tubes and PCR plastics).
  • Include controls in every run: a no-template control to reveal contamination, and positive controls to reveal inhibition or degradation.
  1. Reagent preparationCleanest area; no samples, no product
  2. Sample preparationExtraction of nucleic acids
  3. AmplificationClosed tubes
  4. Post-amplification analysisThe only place tubes of product are opened
A one-way workflow keeps amplified product away from clean reagents. Nothing travels back against the arrows.

What removes nucleic acids from surfaces

MethodEffect on DNA and RNANotes
Sodium hypochlorite (bleach)Destroys nucleic acids by oxidationThe standard for surfaces. Use freshly diluted, typically about 0.5% sodium hypochlorite (a 1 in 10 dilution of household bleach), allow contact time, then rinse with water. Corrodes metals.
Commercial DNA-removing solutionsDestroy or remove nucleic acidsFormulated to be less corrosive; follow the stated contact time
UV light (254 nm)Damages DNA so it can no longer be amplifiedFor cabinets and exposed surfaces. Less effective on short fragments, dried material and anything in shadow; lamps lose output with age
70% ethanolDoes not destroy DNAA disinfectant, not a decontaminant for nucleic acids. Useful for rinsing off bleach residue
AutoclavingKills organisms but leaves DNA fragments that can still be amplifiedNot a reliable way to remove DNA from PCR materials

Never mix bleach with acids or with ammonia-containing products, and never autoclave it.

Dealing with RNases

RNases survive autoclaving and regain activity on cooling, so sterilising is not enough.

  • Glassware and metal: bake at high temperature, for example 180 °C or more for several hours.
  • Water and solutions: use certified nuclease-free water and reagents. Water can be treated with diethyl pyrocarbonate (DEPC) and then autoclaved to destroy the DEPC; DEPC cannot be used with solutions containing amines such as Tris.
  • Surfaces and equipment: use an RNase-decontaminating solution, then rinse with nuclease-free water.
  • Plastics: use disposable, certified RNase-free items.
  • In the reaction: RNase inhibitors protect RNA during handling.
  • Always wear gloves, and change them after touching skin, hair or shared surfaces.

Preventing carry-over inside the reaction

A widely used enzymatic safeguard replaces one of the four building blocks in the PCR (dTTP) with dUTP, so that every amplicon contains uracil. Before each new reaction, the enzyme uracil-DNA glycosylase destroys any uracil-containing DNA carried over from earlier runs, while the natural sample DNA is untouched. The enzyme is then inactivated by the first heating step.

A routine that works

  1. Before and after each session, wipe benches, pipettes and racks with diluted bleach or a DNA-removing solution, then with water or 70% ethanol.
  2. Run the UV lamp in PCR cabinets between sessions.
  3. Clean centrifuges, door handles and other touch points on a schedule.
  4. Service and decontaminate pipettes regularly (pipetting techniques).
  5. Track the no-template controls. A positive control blank means stop, decontaminate, and replace reagents before continuing.

If contamination is found

Discard open reagents and working stocks, decontaminate the area and equipment thoroughly, use fresh aliquots, and repeat the controls before resuming work. If the same contamination returns, look for the route: a shared pipette, a piece of equipment moved between areas, or amplified product being opened in a clean area.

General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.