Why technique matters
A calibrated pipette delivers its stated volume only when it is used the way it was calibrated. With air-displacement micropipettes, differences in technique between two people can easily exceed the tolerance of the instrument. Consistent technique improves the precision of any method at no cost.
Proper forward pipetting
Forward pipetting is the standard technique for aqueous solutions.
- Fit the tip firmly.
- Press the plunger to the first stop.
- Hold the pipette vertical, immerse the tip just below the surface, and release the plunger slowly and smoothly. Wait about a second for the liquid to finish rising.
- Withdraw the tip, touching it against the wall of the vessel to remove liquid from the outside.
- To dispense, hold the tip against the wall of the receiving vessel at an angle, press slowly to the first stop, pause, then press to the second stop to blow out the remainder.
- Withdraw the tip along the wall, then release the plunger.
Reverse pipetting
Reverse pipetting takes up more than the set volume and leaves the excess in the tip. It is more precise for difficult liquids.
- Press the plunger to the second stop.
- Immerse the tip and release the plunger slowly, drawing up the set volume plus an excess.
- Dispense by pressing to the first stop only. The excess stays in the tip.
- Discard the remainder with the tip, or return it to the source.
Use reverse pipetting for viscous liquids such as glycerol and protein solutions, foaming liquids such as detergents and culture media, volatile liquids and very small volumes. For the most difficult liquids, a positive-displacement pipette is better still (types of pipette).
The habits that decide accuracy
- Pre-wet the tip. Draw up and dispense the liquid two or three times before taking the measured volume. This humidifies the air cushion and coats the inside of the tip; without it, the first volumes are low. It matters most for volatile liquids.
- Immersion depth. Immerse only as deep as needed:
| Volume | Immersion depth |
|---|---|
| Up to about 1 µL | About 1 mm |
| 1 to 100 µL | 2 to 3 mm |
| 100 to 1,000 µL | 2 to 4 mm |
| 1 to 10 mL | 3 to 6 mm |
Too deep, and liquid clings to the outside of the tip and the extra pressure forces in more; too shallow, and air is drawn in.
- Angle. Aspirate with the pipette vertical, or within about 20 degrees of it. Dispense against the vessel wall at an angle.
- Rhythm and speed. Use the same smooth, unhurried plunger movement every time. Letting the plunger snap back draws in air and aerosols, and can pull liquid into the shaft.
- Temperature. Let the pipette, the tips and the liquid reach the same temperature. Cold liquid is delivered in excess and warm liquid in deficit with an air-displacement pipette. Avoid holding the pipette in a warm hand for long periods.
- Volume range. Work in the upper part of the pipette's range, and set the volume by turning down to the value from above.
- Tips. Use tips that fit, and change the tip whenever there is any risk of carry-over.
Liquids that need special care
| Liquid | Problem | What to do |
|---|---|---|
| Viscous (glycerol, oils, serum) | Slow to move; clings to the tip | Reverse pipetting; slow movement; wide-bore or low-retention tips; positive displacement |
| Volatile (acetone, ethanol, chloroform) | Vapour expands the air cushion; liquid drips | Pre-wet several times; work quickly; positive displacement |
| Foaming (detergents, protein solutions) | Bubbles | Reverse pipetting; slow movement |
| Dense or very low surface tension | Volume error; dripping | Positive displacement; check gravimetrically |
| Hot or cold | Volume error | Bring to room temperature where possible |
Glass and serological pipettes
- Use a pipette controller, never the mouth.
- Read the bottom of the meniscus at eye level, with the pipette vertical.
- Wipe the outside of the tip before adjusting to the mark.
- Let the liquid run out freely with the tip touching the wall of the vessel, and wait the specified time.
- Do not blow out a pipette calibrated "to deliver", unless it is marked as a blow-out type.
Common errors
| Error | Effect |
|---|---|
| No pre-wetting | First volumes too low |
| Tip immersed too deep | Volume too high |
| Pipette held at an angle while aspirating | Volume too high |
| Plunger released too fast | Air bubbles; liquid in the shaft; volume too low |
| Tip not sealing | Leak; volume too low |
| Using the second stop to aspirate in forward mode | Volume too high |
| Wiping the tip opening with tissue | Liquid drawn out; volume too low |
| Working at the bottom of the range | Poor accuracy and precision |
Checking your pipetting
A simple gravimetric check shows both the state of the pipette and the quality of the technique. Dispense the set volume of pure water ten times onto a suitable balance, in a draught-free place at a stable temperature, and convert each weight to volume using the density of water at that temperature. The mean shows the accuracy; the spread shows the precision. ISO 8655 gives the full procedure and the tolerances.
Looking after the pipette and yourself
- Never lay a pipette down with liquid in the tip, and never turn it upside down: liquid reaches the piston and corrodes it.
- Store pipettes upright on a stand.
- Clean and decontaminate the shaft as the manufacturer instructs; after any liquid has entered it, take the pipette out of use and have it serviced.
- Do not turn the volume setting beyond the range.
- Pipetting for long periods strains the hand and wrist. Use light, well-balanced pipettes, low-force tips, relaxed posture and regular breaks; electronic pipettes help for repetitive work.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.