Pipetting techniques: forward, reverse and how to avoid errors

How to pipette accurately: forward and reverse technique, pre-wetting, immersion depth and angle, viscous and volatile liquids, common errors and checks.

  • 4 min read
  • 1 figure
  • Reviewed 25 November 2026
On this page 9 sections
  1. Why technique matters
  2. Proper forward pipetting
  3. Reverse pipetting
  4. The habits that decide accuracy
  5. Liquids that need special care
  6. Glass and serological pipettes
  7. Common errors
  8. Checking your pipetting
  9. Looking after the pipette and yourself

Why technique matters

A calibrated pipette delivers its stated volume only when it is used the way it was calibrated. With air-displacement micropipettes, differences in technique between two people can easily exceed the tolerance of the instrument. Consistent technique improves the precision of any method at no cost.

Proper forward pipetting

Forward pipetting is the standard technique for aqueous solutions.

  1. Fit the tip firmly.
  2. Press the plunger to the first stop.
  3. Hold the pipette vertical, immerse the tip just below the surface, and release the plunger slowly and smoothly. Wait about a second for the liquid to finish rising.
  4. Withdraw the tip, touching it against the wall of the vessel to remove liquid from the outside.
  5. To dispense, hold the tip against the wall of the receiving vessel at an angle, press slowly to the first stop, pause, then press to the second stop to blow out the remainder.
  6. Withdraw the tip along the wall, then release the plunger.

Reverse pipetting

Reverse pipetting takes up more than the set volume and leaves the excess in the tip. It is more precise for difficult liquids.

  1. Press the plunger to the second stop.
  2. Immerse the tip and release the plunger slowly, drawing up the set volume plus an excess.
  3. Dispense by pressing to the first stop only. The excess stays in the tip.
  4. Discard the remainder with the tip, or return it to the source.

Use reverse pipetting for viscous liquids such as glycerol and protein solutions, foaming liquids such as detergents and culture media, volatile liquids and very small volumes. For the most difficult liquids, a positive-displacement pipette is better still (types of pipette).

Plunger positions in forward and reverse pipettingRestFirst stopSecond stopForward pipettingPressdownRelease:draws liquidDispenseBlow outthe restReverse pipettingPressfurtherRelease: drawsextra liquidDispenseExcess staysin the tipThe height of each dot is the plunger position at that moment
Plunger positions. Forward pipetting blows out the last liquid; reverse pipetting draws an excess and leaves it in the tip.

The habits that decide accuracy

  • Pre-wet the tip. Draw up and dispense the liquid two or three times before taking the measured volume. This humidifies the air cushion and coats the inside of the tip; without it, the first volumes are low. It matters most for volatile liquids.
  • Immersion depth. Immerse only as deep as needed:
VolumeImmersion depth
Up to about 1 µLAbout 1 mm
1 to 100 µL2 to 3 mm
100 to 1,000 µL2 to 4 mm
1 to 10 mL3 to 6 mm

Too deep, and liquid clings to the outside of the tip and the extra pressure forces in more; too shallow, and air is drawn in.

  • Angle. Aspirate with the pipette vertical, or within about 20 degrees of it. Dispense against the vessel wall at an angle.
  • Rhythm and speed. Use the same smooth, unhurried plunger movement every time. Letting the plunger snap back draws in air and aerosols, and can pull liquid into the shaft.
  • Temperature. Let the pipette, the tips and the liquid reach the same temperature. Cold liquid is delivered in excess and warm liquid in deficit with an air-displacement pipette. Avoid holding the pipette in a warm hand for long periods.
  • Volume range. Work in the upper part of the pipette's range, and set the volume by turning down to the value from above.
  • Tips. Use tips that fit, and change the tip whenever there is any risk of carry-over.

Liquids that need special care

LiquidProblemWhat to do
Viscous (glycerol, oils, serum)Slow to move; clings to the tipReverse pipetting; slow movement; wide-bore or low-retention tips; positive displacement
Volatile (acetone, ethanol, chloroform)Vapour expands the air cushion; liquid dripsPre-wet several times; work quickly; positive displacement
Foaming (detergents, protein solutions)BubblesReverse pipetting; slow movement
Dense or very low surface tensionVolume error; drippingPositive displacement; check gravimetrically
Hot or coldVolume errorBring to room temperature where possible

Glass and serological pipettes

  • Use a pipette controller, never the mouth.
  • Read the bottom of the meniscus at eye level, with the pipette vertical.
  • Wipe the outside of the tip before adjusting to the mark.
  • Let the liquid run out freely with the tip touching the wall of the vessel, and wait the specified time.
  • Do not blow out a pipette calibrated "to deliver", unless it is marked as a blow-out type.

Common errors

ErrorEffect
No pre-wettingFirst volumes too low
Tip immersed too deepVolume too high
Pipette held at an angle while aspiratingVolume too high
Plunger released too fastAir bubbles; liquid in the shaft; volume too low
Tip not sealingLeak; volume too low
Using the second stop to aspirate in forward modeVolume too high
Wiping the tip opening with tissueLiquid drawn out; volume too low
Working at the bottom of the rangePoor accuracy and precision

Checking your pipetting

A simple gravimetric check shows both the state of the pipette and the quality of the technique. Dispense the set volume of pure water ten times onto a suitable balance, in a draught-free place at a stable temperature, and convert each weight to volume using the density of water at that temperature. The mean shows the accuracy; the spread shows the precision. ISO 8655 gives the full procedure and the tolerances.

Looking after the pipette and yourself

  • Never lay a pipette down with liquid in the tip, and never turn it upside down: liquid reaches the piston and corrodes it.
  • Store pipettes upright on a stand.
  • Clean and decontaminate the shaft as the manufacturer instructs; after any liquid has entered it, take the pipette out of use and have it serviced.
  • Do not turn the volume setting beyond the range.
  • Pipetting for long periods strains the hand and wrist. Use light, well-balanced pipettes, low-force tips, relaxed posture and regular breaks; electronic pipettes help for repetitive work.

General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.