A relative technique
An HPLC detector does not measure amount directly. It gives a signal, and the signal becomes a concentration only by comparison with a standard of known content run under the same conditions. Any error in the standard, in its stated purity, in how it was weighed or in how it was stored, passes straight into every result calculated from it. The standard is as important as the column.
Types of standard
| Type | What it is | Typical use |
|---|---|---|
| Pharmacopoeial (primary) reference standard | Established by a pharmacopoeia for the tests in its monographs | Identification, assay and impurity tests to that pharmacopoeia |
| Certified reference material (CRM) | Value certified with a stated uncertainty and traceability, from an accredited producer | Calibration and validation where traceability is required |
| Secondary (working) standard | Characterised against a primary standard | Routine work, to conserve the primary standard |
| Analytical standard | High-purity substance with a certificate of analysis | Method development, identification, general calibration |
| Standard mixture | Several compounds at stated concentrations in solution | Multi-component methods, system checks |
Browse reference standards, including primary reference standards.
Read the certificate
- Use the assigned content stated for the intended use, not a nominal 100%. Note whether it is given "as is" or on a dried or anhydrous basis.
- Follow the stated storage conditions and any instruction to dry or not to dry before use.
- Check the validity. Pharmacopoeial standards are valid for the current batch as listed by the pharmacopoeia, and are intended for the use described in the monograph.
- Keep the certificate with the batch record. See certificates of conformance and analysis.
Calibration methods
- External standard. Standards and samples are injected separately and compared. Simple, and the usual choice when injection precision is good.
- Internal standard. A known amount of a different, well-separated compound is added to every standard and sample. The ratio of analyte to internal standard corrects for losses in sample preparation and variation in injection. Common in bioanalysis and after extraction steps.
- Standard addition. Known amounts of the analyte are added to portions of the sample itself. It compensates for matrix effects when no blank matrix is available.
Calibrate over the range the samples will cover, with enough levels to show the response is linear, and bracket long sequences with repeat standards to detect drift.
Handling that protects the result
- Let a refrigerated or frozen container reach room temperature before opening, so moisture does not condense on the substance.
- Weigh on a calibrated balance suited to the amount; very small weighings carry large relative errors. Record the actual weight.
- Dissolve completely, in a solvent compatible with the mobile phase, using volumetric glassware.
- Protect light-sensitive solutions and note how long solutions are stable. Do not assume a stock solution keeps.
- Prepare two independent weighings where the procedure calls for it, and check that they agree.
- Never return unused material to the container.
Standards for system suitability
System suitability solutions, such as a mixture of the analyte and a closely eluting impurity, show that the system can deliver the required resolution, peak shape and repeatability on the day. They are run before the samples and are part of the evidence that the result is valid (testing a column).
Standards for identification
A compound is identified by matching its retention time, and ideally its spectrum, with a reference standard run in the same sequence. Retention time alone is weak evidence; see peak purity.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.