How the pharmacopoeias name a column

A pharmacopoeial monograph does not name a brand. It describes the packing, and it is the laboratory's job to pick a column that fits the description and passes the system suitability test of the monograph.

  • The United States Pharmacopeia (USP) gives each type of packing an L number.
  • The European Pharmacopoeia (Ph. Eur.) describes the stationary phase in words, as a reagent, for example "octadecylsilyl silica gel for chromatography R" or its end-capped and base-deactivated variants.

Common USP L numbers

USP codePackingUsual name
L1Octadecyl silane bonded to porous or non-porous silica or ceramic particlesC18 (ODS)
L3Porous silica particlesSilica
L7Octylsilane bonded to silicaC8
L8Aminopropylsilane bonded to silicaAmino (NH2)
L9Silica with a strongly acidic cation-exchange coatingStrong cation exchange (SCX)
L10Nitrile groups bonded to silicaCyano (CN)
L11Phenyl groups bonded to silicaPhenyl
L13Trimethylsilane bonded to silicaC1
L14Silica with a strongly basic quaternary ammonium anion-exchange coatingStrong anion exchange (SAX)
L16Dimethylsilane bonded to silicaC2
L20Dihydroxypropane groups bonded to silicaDiol
L26Butyl silane bonded to silicaC4
L43Pentafluorophenyl groups bonded to silicaPFP
L62C30 silane bonded to silicaC30

The list in the current USP is much longer and is revised from time to time. Always read the definition in the edition you work to.

Matching ranges: C18, silica, C8, amino, cyano, phenyl, diol, C4, PFP and C30.

Bonded phases on a silica surface: C18, C8, phenyl, cyano and amino groups comparedSilica surfaceC18USP L1C8USP L7PhenylUSP L11C≡NCyanoUSP L10NH2AminoUSP L8
Common bonded phases on silica, drawn schematically, with their USP codes. Chain length and end group decide retention and selectivity.

An L number is a class, not a guarantee

Hundreds of columns qualify as L1. They differ in base silica, pore size, carbon load, bonding and end-capping, and so in selectivity. Two L1 columns can give a different elution order for the same sample. A column that fits the description is only a candidate until it passes the monograph's system suitability criteria: resolution, tailing factor, plate number and repeatability, as the monograph requires.

How to choose within a class

  1. Match everything the monograph specifies: L number or description, particle size, length and inner diameter. The permitted changes to these are set out in method adjustments under USP 621 and Ph. Eur. 2.2.46.
  2. Read the wording closely. "End-capped", "base-deactivated", "polar-embedded" or "for use with highly aqueous mobile phases" each narrow the choice.
  3. Look for the column used during development. The Ph. Eur. Knowledge database and USP's chromatographic column resources name columns found suitable. They are given for information, not as a requirement, but they are the surest starting point.
  4. Check the practical limits: pH range, pressure limit and compatibility with your instrument.
  5. Run the system suitability test with the prescribed reference standards before any sample.

Keep the evidence

Record the column make, part number, serial number and batch with each analysis. If a replacement column from a new batch fails system suitability, that record is what lets you find out why.

General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.