Three questions
Tubing for a liquid chromatography system is chosen by answering three questions: what pressure must it hold, what liquids will pass through it, and how much volume can that part of the flow path afford?
Materials
| Material | Pressure | Strengths | Limits |
|---|---|---|---|
| Stainless steel (316) | Highest; standard for HPLC and UHPLC | Compatible with nearly all HPLC solvents; rigid and durable | Attacked by halide salts and strong acids over time; can release metal ions; needs a tube cutter |
| PEEK | Moderate to high, depending on bore and temperature | Metal-free, flexible, easy to cut, finger-tight fittings | Swells in tetrahydrofuran, dichloromethane and dimethyl sulfoxide; not for concentrated nitric or sulfuric acid |
| PEEK-lined or fused-silica-lined steel | High | Metal-free wetted surface with the strength of steel | Cannot be cut to length by the user; minimum bend radius |
| PTFE, FEP, PFA | Low | Chemically inert to almost everything; flexible; translucent | Low-pressure lines only; PTFE lets gases through its wall |
| ETFE | Low to moderate | Stronger than PTFE with good chemical resistance | Not for the high-pressure side |
Pressure ratings depend on bore, wall thickness and temperature. Use the rating stated for the specific tubing, and remember it falls as temperature rises. For solvent compatibility see chemical resistance of plastics.
Where each is used
- Reservoir to pump (low pressure): PTFE, FEP or PFA, usually 1/8 inch outer diameter with a wide bore so the pump is never starved.
- Pump to injector (high pressure, before the sample): stainless steel or PEEK. Volume here does not broaden peaks, though it adds to the gradient delay.
- Injector to column and column to detector (high pressure, carrying the sample): the narrowest and shortest capillaries in the system, in steel, PEEK or lined steel.
- Detector to waste: wide-bore PTFE or PEEK, sized so it adds no back pressure beyond what the flow cell tolerates.
The standard outer diameter for capillaries is 1/16 inch (1.59 mm).
Choosing the inner diameter
Between injector and detector, a wider bore adds extra-column volume and broadens peaks; a narrower bore adds back pressure and blocks more easily.
| Inner diameter | Volume per metre | Typical use |
|---|---|---|
| 0.005 inch (0.127 mm) | 13 µL | UHPLC; columns of 2.1 mm inner diameter and small particles |
| 0.007 inch (0.178 mm) | 25 µL | General HPLC; columns of 3.0 to 4.6 mm inner diameter |
| 0.010 inch (0.254 mm) | 51 µL | Conventional HPLC with 4.6 mm columns and 5 µm particles |
| 0.020 inch (0.508 mm) | 203 µL | Pump outlet lines, preparative work |
| 0.030 to 0.040 inch (0.76 to 1.02 mm) | 456 to 811 µL | Waste and low-pressure transfer lines |
Manufacturers colour-code PEEK tubing by bore, but the schemes are not universal: read the printed size.
Pressure drop
For a given flow, the pressure lost along a tube is proportional to its length and to the viscosity of the liquid, and inversely proportional to the fourth power of its bore. Halving the inner diameter multiplies the pressure drop sixteen times. This is why very narrow capillaries are kept short, and why a partly blocked capillary raises the pressure so sharply.
Cutting and fitting
- Cut square and burr-free: a rotary cutter and deburring for steel, a guillotine-type cutter for polymer tubing. Do not use scissors or side-cutters, which crush the end.
- Check the bore is open and flush the piece before connecting it to a column.
- Keep bends gentle. A kink is a restriction and a weak point.
- Make each joint with the tubing pushed fully home (HPLC fittings, preventing dead volume).
Beyond HPLC
For peristaltic pumps, gas lines and general liquid transfer, the same three questions apply, with the addition of flexibility, transparency, gas permeability and, where relevant, suitability for sterilisation. Match the material to the liquid first, then the size to the flow.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.