What a fitting has to do
A fitting joins tubing to a port (on a column, valve, pump or detector) so that the joint holds the system pressure without leaking and adds no unswept volume. It has three parts: the nut, which provides the force; the ferrule, which grips the tubing and seals against the port; and the tubing itself. In one-piece fittings the nut and ferrule are combined.
Stainless steel fittings
A steel nut and ferrule give the highest pressure rating. When first tightened, the ferrule is permanently squeezed (swaged) onto the tubing at a fixed distance from the tube end. That distance was set by the depth of the port it was made up in.
Port depths differ between manufacturers. A steel ferrule swaged in one maker's port and then used in another's will either leave a gap in front of the tube (unswept volume and tailing peaks) or stop the ferrule from seating (a leak). Keep steel fittings with the port they were made for, or cut the tubing and fit a new ferrule.
PEEK finger-tight fittings
Polymer fittings, usually PEEK, are tightened by hand and do not grip the tube permanently. Each time they are loosened the tubing can slide to suit the new port, so they adapt to any standard port. They are the usual choice for column connections on conventional HPLC. Their pressure rating is lower than steel and varies with the design; fittings sold for UHPLC are built and rated for higher pressures. Check the rating against your system pressure, and do not use tools on fittings meant to be finger-tight.
Threads and port shapes
| Thread | Used with | Typical use |
|---|---|---|
| 10-32 | 1/16 inch outer diameter tubing, coned port | High-pressure connections: columns, injectors, detectors |
| 1/4-28 | 1/16 or 1/8 inch tubing, flat-bottom port | Low-pressure connections: solvent lines, valves, waste |
| M6 | 1/16 or 1/8 inch tubing, flat-bottom port | Low-pressure connections on some instruments |
Two rules follow:
- Do not mix port shapes. A coned ferrule seals in a coned port; a flangeless ferrule seals on a flat bottom. Forcing one into the other damages the port.
- Do not mix threads. 1/4-28 (6.35 mm) and M6 (6 mm) are close enough in size to start and then strip.
The inch sizes are explained in inches to mm.
Making the connection
- Cut the tubing square and clean, with the proper cutter, and check the bore is open.
- Slide on the nut and ferrule. In a coned port (10-32) the ferrule's taper faces the port. With a flangeless ferrule for a flat-bottom port (1/4-28, M6) the taper faces the nut, and the flat face seals on the bottom of the port.
- Push the tubing all the way into the port and hold it there.
- Tighten the nut while still pushing: finger-tight for polymer fittings, and for steel fittings as the manufacturer instructs (commonly finger-tight plus a part turn with a spanner for a new ferrule).
- Pressurise and check for leaks. If it weeps, tighten slightly. If it still leaks, take it apart and inspect instead of forcing it.
Overtightening is the commonest way to ruin a fitting: it deforms the ferrule, narrows or crushes polymer tubing and damages the port threads.
Unions, adapters and tees
Use zero-dead-volume unions to join two capillaries, with a bore matched to the tubing. Adapters join different thread types. Every extra joint is a possible leak and a little added volume, so use only what the method needs (preventing dead volume).
Finding and curing leaks
- A slow leak of buffered mobile phase leaves a white crust of salt around the fitting.
- A leak on the high-pressure side shows as falling pressure, drifting retention times and noisy baselines.
- A fitting on the low-pressure side can draw air in without dripping (air bubbles).
- Rinse salt away with water before retightening, so crystals do not score the seal.
- Replace any fitting with a damaged thread, a cracked nut or a deformed ferrule from HPLC replacement parts.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.