Lancetech Limited
HPLC system care

Preventing air bubbles in the HPLC system

Why air bubbles form in an HPLC system, how to recognise them (pressure ripple, baseline spikes, drifting retention) and how to remove and prevent them.

  • 2 min read
  • 1 figure
  • Reviewed 25 November 2026
On this page 5 sections
  1. Why bubbles form
  2. How to recognise them
  3. Getting rid of them
  4. Preventing them
  5. When it is not air

Why bubbles form

Solvents standing in air hold dissolved gas. That gas comes out of solution when:

  • Two solvents are mixed. A mixture of water with methanol or acetonitrile holds less air than the two liquids held separately, so mixing releases gas. This is why bubbles appear at the mixing point of a gradient pump.
  • The pressure drops. Liquid leaving the column falls from high pressure to almost none in the detector cell.
  • The temperature rises, in a warm laboratory or a heated detector cell.

Air can also be drawn in directly through a loose fitting on the low-pressure side of the pump, or when a reservoir runs low.

How to recognise them

Where the bubble isWhat you see
Pump head or check valvesPressure that pulses, falls or will not build; retention times that drift or jump; unstable flow
Mixer or proportioning valveWrong or unstable composition, shifting retention in gradients
Detector flow cellSharp spikes, sudden baseline steps, a noisy or wandering baseline
Autosampler syringe or loopPoor area repeatability, small or missing peaks
  1. Reservoir and inlet filterLow level or a blocked filter starves the pump
  2. DegasserMust be on and flushed
  3. Pump and check valvesPressure pulses, unstable flow
  4. Mixing pointGas is released when solvents mix
  5. Detector cellPressure drops: spikes and noise
Where air appears in an HPLC system, and what each place does to the result.

Getting rid of them

  1. Check the reservoirs: enough solvent, inlet filters fully submerged and not blocked.
  2. Purge the pump with the purge valve open at high flow, each channel in turn, until no bubbles pass. A water-miscible alcohol such as methanol or 2-propanol wets the surfaces well and helps to dislodge a trapped bubble from a pump head or check valve.
  3. Prime the autosampler and its wash lines.
  4. Clear the detector cell by flushing with a degassed, low-surface-tension solvent such as methanol or 2-propanol. Applying a little back pressure after the cell helps to push the bubble through. Never exceed the pressure rating of the cell.

Before changing between water-based buffers and an alcohol, check that nothing will precipitate.

Preventing them

  • Degas the mobile phase. An inline vacuum degasser is the standard solution; make sure it is switched on and working, and that its chambers are flushed when solvents are changed. Helium sparging is an effective alternative. Sonication or vacuum filtration alone give only brief, partial degassing.
  • Let solvents reach room temperature before use. Cold solvent from a store warms up in the tubing and releases gas.
  • Premix isocratic mobile phases where practical; a premixed solvent has already given up most of the gas that mixing releases.
  • Keep the inlet filters clean. A blocked filter makes the pump pull a partial vacuum on the solvent, and gas comes out of solution in the inlet line.
  • Check the low-pressure fittings between reservoir, degasser and pump. A fitting that leaks air in may show no liquid leaking out.
  • Keep reservoirs above the pump, with enough solvent that the inlet lines stay full.
  • Use a back-pressure restrictor after the detector if the manufacturer recommends one, within the rating of the flow cell.
  • Store the system wet with a suitable solvent, never dry.

When it is not air

Pressure fluctuation can also come from a worn piston seal, a dirty or sticking check valve, or a leak. If degassing and purging do not cure it, inspect those parts and replace as needed from HPLC replacement parts. Preparation of the mobile phase itself is covered in mobile phase preparation.

General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.

New products and price changes, once a month

No daily emails. Unsubscribe in one click.

  • New arrivals
  • Price changes
  • Brand offers

About once a month. Unsubscribe any time. See our privacy policy.

Air bubbles in HPLC: causes, symptoms and how to prevent them | Lancetech Limited