Lancetech Limited
HPLC system care

Installing an HPLC column and getting started

How to install a new HPLC column: check the shipping solvent, flush the system, connect in the flow direction, make fittings correctly, equilibrate and test.

  • 2 min read
  • 2 figures
  • Reviewed 25 November 2026
On this page 6 sections
  1. Before you connect anything
  2. Avoid two classic mistakes
  3. Connecting the column
  4. Make the fittings properly
  5. Equilibrate and test
  6. Protect it from day one

Before you connect anything

  1. Read the paperwork. The certificate or test report states the shipping solvent, the flow direction, the test conditions and the limits for pH, pressure and temperature. Keep it: it is the reference for every later test.
  2. Check the shipping solvent against your mobile phase. Reversed-phase columns are usually shipped in a mixture of water and acetonitrile or methanol. Normal-phase columns are shipped in a non-polar solvent such as heptane or hexane. The first liquid you pump must mix with it.
  3. Flush the system without the column. Replace the old column with a union and flush pump, autosampler and tubing with the new mobile phase, so that no buffer, incompatible solvent or debris from the previous method reaches the new column.

Avoid two classic mistakes

  • Buffer precipitation. If the mobile phase contains a buffer, first flush the column with the same water and organic mixture without the salt, then switch to the buffered mobile phase.
  • Immiscible solvents. Never go straight from a non-polar solvent to water. Use an intermediate that mixes with both, such as 2-propanol, at a reduced flow rate because it is viscous.

Connecting the column

  1. Fit the column so the flow follows the arrow on the label.
  2. Connect the inlet first and leave the outlet open, with a beaker underneath.
  3. Start the pump at a low flow, for example 0.1 to 0.2 mL/min for a 4.6 mm column, and wait until liquid drips steadily from the outlet. This pushes out air instead of sending it into the detector.
  4. Connect the outlet to the detector and raise the flow to the working value in steps. Sudden pressure changes disturb the packed bed.
  5. Check each fitting for leaks. Note the back pressure at the working flow: it is your baseline for troubleshooting later.
  1. Flush the systemWithout the column, with the new mobile phase
  2. Check the solvents mixShipping solvent and mobile phase; no buffer at first
  3. Connect the inletIn the direction of the arrow
  4. Start at low flowOutlet open until liquid drips
  5. Connect the detectorThen raise the flow in steps
  6. Equilibrate and testRecord the first result as the baseline
Installing a new HPLC column, in order.

Make the fittings properly

The tubing must sit fully home in the column port before the nut is tightened. A gap in front of the tubing is a small mixing chamber that broadens and tails every peak; a ferrule set too far along for the port will not seal. Fittings swaged on one manufacturer's column may not fit another's. Finger-tight PEEK fittings adjust to each port and avoid the problem. See HPLC fittings and preventing dead volume.

A correctly seated HPLC fitting beside one with a gap in front of the tubingTubing pushed fully homeNutFerrulePortTube end meets the bottom of the port:no unswept volumeTubing not fully homeNutFerrulePortGap in front of the tube: sample lingers,peaks broaden and tail
Left: the tubing sits against the bottom of the port. Right: a gap in front of the tubing is unswept volume, the commonest cause of sudden peak tailing.

Equilibrate and test

Pump 10 to 20 column volumes of mobile phase and wait for a steady baseline and pressure (column equilibration). Before the first samples:

  • Run the manufacturer's test mixture, or your own, and record plate number, peak symmetry, retention and pressure (testing a column).
  • Run the system suitability test of your method.

These first results are the fingerprint of the healthy column. Without them you cannot tell later whether the column has changed.

Protect it from day one

Fit a guard column of the same packing or an inline filter, filter samples, and start a log of injections, pressure and storage. Day-to-day care is covered in maintaining and storing HPLC columns.

General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.