Why passivate
Stainless steel resists corrosion because of a very thin, chromium-rich oxide layer on its surface. In an HPLC system that layer is attacked over time by acidic mobile phases, by halide salts such as chloride, and by mechanical wear. Where it is damaged, the steel releases iron and other metal ions and exposes active metal surface.
For many analyses this goes unnoticed. For others it matters a great deal:
- Compounds that bind metals, such as phosphorylated compounds, some antibiotics and other chelating molecules, give tailing peaks or low recovery.
- Proteins and other biomolecules can adsorb to metal surfaces or be altered by metal ions.
- Trace metal and ion analysis suffers from the background.
- The column can collect metal ions that then interact with later samples.
Passivation is a controlled acid treatment that removes free iron from the surface and rebuilds the protective oxide layer.
When to consider it
- On a new system or after replacing stainless steel parts, if the manufacturer advises it.
- After working with corrosive mobile phases, particularly halide salts at low pH.
- When metal-sensitive compounds start to tail or recoveries fall, and the column has been ruled out.
- At regular intervals for methods known to be sensitive.
The procedure in outline
The exact acid, concentration, time and the parts that must be bypassed differ between instruments. Follow the instrument manufacturer's procedure. The general pattern is:
- Remove the column and replace it with a union. The column must never see the passivation acid. Bypass or remove anything the manufacturer says is not compatible, which can include the detector flow cell and some degassers.
- Flush out the mobile phase with water, so that no buffer or organic solvent meets the acid.
- Flush with the acid for the specified time. Nitric acid of about 6 mol/L (roughly 30%) is widely used; some procedures use phosphoric or citric acid instead.
- Rinse with water until the outflow is neutral, checking with indicator paper.
- Flush with the storage solvent or mobile phase, then refit the column and equilibrate.
Never mix organic solvent with nitric acid: remove one completely before introducing the other. Wear acid-resistant gloves and eye protection, work with good ventilation, and collect the waste as acid waste.
Repeat regularly
Passivation is not permanent. The layer is worn down again by the same conditions that damaged it. For metal-sensitive methods, build a repeat into the maintenance schedule, at the interval the manufacturer suggests or that your own results show is needed.
Reducing the need for it
- Flush salts out at the end of each day and do not leave halide-containing mobile phases standing in the system.
- Avoid chloride at low pH where the method allows another buffer.
- Add a metal chelator to the mobile phase only if the method permits it.
PEEK and other alternatives
Where metal contact must be avoided altogether, flow paths are built from other materials: PEEK tubing and fittings, titanium, nickel-cobalt alloys, or steel with an inert lining. These "biocompatible" or "bio-inert" systems are common for protein work (monoclonal antibodies).
PEEK has limits of its own. Its pressure rating is lower than steel and falls with wider bores and higher temperature. It is not suitable for concentrated nitric or sulfuric acid, and it swells in tetrahydrofuran, dichloromethane and dimethyl sulfoxide. Makers differ on how much nitric acid PEEK tolerates, so check the tubing and instrument maker's limit before passivating a PEEK flow path. See choosing tubing and chemical resistance of plastics.
General guidance from Lancetech Limited. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.