What the liner does
The liner is the glass tube inside the GC inlet. The sample is injected into it, vaporises, mixes with carrier gas and passes to the column. It gives the vapour room to expand, keeps the sample away from hot metal, and traps non-volatile residue so it does not reach the column. Because every sample touches it, the liner has more influence on peak shape, discrimination and repeatability than any other inlet part.
Liners are made to fit specific inlets. Outer diameter, length and the type of seal differ between instrument makers, so always choose a liner made for your inlet.
Match the liner to the injection mode
| Injection mode | Typical liner | Why |
|---|---|---|
| Split | Straight liner with glass wool, or a design with a baffle or cup | Promotes fast, complete vaporisation and mixing before the split point |
| Splitless | Single taper (gooseneck) at the bottom, with or without wool | The taper directs the vapour onto the column and limits contact with the metal seal |
| Splitless, active compounds | Double taper, no wool | Encloses the vapour and minimises surface contact |
| Direct or on-column type | Liner that holds or guides the column | Transfers the whole sample to wide-bore columns |
| Programmed temperature (PTV) | Narrow, often baffled or sintered | Low thermal mass and a surface to hold the liquid during solvent venting |
Glass wool: yes or no
For it: wool gives a large hot surface for quick vaporisation, wipes the needle tip (better repeatability), improves mixing for split injection and catches septum fragments and non-volatile residue.
Against it: wool is the most active surface in the inlet. Sensitive compounds, such as some pesticides, phenols and free acids, can adsorb or break down on it.
Use only deactivated wool, positioned as the liner maker intends. For labile analytes try a liner without wool, or one with a built-in frit or baffle instead.
Deactivation
Bare glass carries active sites that adsorb polar compounds, causing tailing and loss of response at low levels. Deactivated liners have a treated surface. Deactivation wears off with use and with dirty samples, which is one reason liners are replaced, not cleaned.
Liner volume and backflash
A liquid expands enormously when it vaporises. If the vapour cloud is larger than the liner, it overflows into the cold gas lines above (backflash), causing carry-over, ghost peaks, poor repeatability and tailing solvent peaks.
Approximate vapour volume from 1 µL of liquid at 250 °C and an inlet pressure of 20 psi above atmospheric (about 1.4 bar):
| Solvent | Vapour volume |
|---|---|
| Water | about 1,000 µL |
| Methanol | about 450 µL |
| Acetonitrile | about 350 µL |
| Dichloromethane | about 285 µL |
| Ethyl acetate | about 185 µL |
| Toluene | about 170 µL |
| Hexane | about 140 µL |
A liner of 4 mm inner diameter holds roughly 900 µL; one of 2 mm holds roughly 250 µL. So 1 µL of hexane fits easily, while 1 µL of water or 2 µL of methanol does not fit a standard 4 mm liner under these conditions.
To avoid backflash: inject less, choose a solvent that expands less, use a liner of larger volume, raise the inlet pressure during injection (a pressure pulse), or lower the inlet temperature. Calculate the vapour volume for your own conditions when developing a method.
Inner diameter
- About 4 mm: the general choice for split and splitless injection of liquid samples.
- About 2 mm or less: for headspace, gas and solid phase microextraction injections, where there is no solvent to expand and a narrow liner gives sharper peaks.
When to replace the liner
There is no fixed interval: it depends on how dirty the samples are. Replace it when you see tailing of active compounds, falling response, poor repeatability, ghost peaks or visible deposits, and whenever you install a column. Change the O-ring and the septum at the same time, and handle the new liner with clean gloves or tweezers, never bare fingers.
Keeping a record of injections per liner for each method lets you replace it before the chromatography suffers.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.