Definition

Gas chromatography (GC) separates compounds that can be turned into vapour without breaking down. The sample is vaporised in a heated inlet and carried by an inert gas, the mobile phase, through a column that holds the stationary phase. Compounds that are more volatile and interact less with the stationary phase reach the detector first.

GC suits volatile and semi-volatile compounds: solvents, fuels, essential oils, fatty acid methyl esters, pesticides, many environmental pollutants and gases. Compounds that are too polar or not volatile enough can often be analysed after derivatisation; those that cannot be vaporised belong to HPLC.

The parts of a gas chromatograph

  1. Carrier gas supply. Helium, hydrogen or nitrogen of high purity, with traps to remove moisture, oxygen and hydrocarbons.
  2. Inlet. Vaporises the sample and transfers it to the column. The split/splitless inlet is the most common. Its liner and septum are consumables.
  3. Column, inside an oven. Where the separation happens.
  4. Detector. Produces a signal for each compound that leaves the column.
  5. Data system. Records the chromatogram.
  1. Carrier gasHelium, hydrogen or nitrogen, through traps
  2. InletVaporises the sample in the liner
  3. Column in the ovenSeparation, controlled by temperature
  4. DetectorFor example FID or mass spectrometer
  5. Data system
The flow path of a gas chromatograph.

Injection modes

  • Split: only a set fraction of the vaporised sample enters the column; the rest is vented. Used for concentrated samples, giving sharp peaks.
  • Splitless: nearly all the sample is transferred. Used for trace analysis.
  • On-column and programmed-temperature inlets: gentler techniques for heat-sensitive or high-boiling compounds.
  • Headspace: only the vapour above the sample is injected. See headspace analysis in GC.

Columns

Most GC today uses capillary columns: fused silica tubing with the stationary phase coated as a thin film on the inner wall.

DimensionCommon rangeEffect of increasing it
Length15 to 60 m (30 m is typical)More resolution, longer run
Inner diameter0.18 to 0.53 mm (0.25 mm is typical)More sample capacity, less efficiency
Film thickness0.1 to 5 µm (0.25 µm is typical)More retention and capacity; suits volatile compounds

Packed columns, wider tubes filled with coated particles, are still used for gases and some older methods.

Stationary phases

Choose the phase by the polarity of the analytes: like dissolves like.

PhasePolarityTypical use
100% dimethylpolysiloxaneNon-polarHydrocarbons, solvents; separation roughly by boiling point
5% phenyl, 95% dimethylpolysiloxaneLow polarityGeneral purpose; semi-volatiles, pesticides
35 to 50% phenyl polysiloxaneMid polarityDrugs, pesticides, confirmation
Cyanopropylphenyl polysiloxane (6% or 14%)Mid polarityResidual solvents, volatile organics
Polyethylene glycol (wax)PolarAlcohols, flavours, fatty acid methyl esters
Porous layer (PLOT)AdsorbentPermanent gases, light hydrocarbons

Browse GC columns.

The oven programme

Temperature is the main control of retention in GC. An isothermal run holds one temperature. A temperature programme starts low and ramps up, so that volatile compounds are separated first and heavier ones are then driven off in reasonable time with sharp peaks. Stay below the column's maximum temperature. Columns usually state two limits: a lower isothermal limit that can be held for long periods, and a higher programmed limit that may be held only for a few minutes at the end of a ramp.

Detectors

DetectorResponds to
Flame ionisation (FID)Almost all organic compounds; the general-purpose detector
Thermal conductivity (TCD)Everything, including permanent gases; less sensitive
Electron capture (ECD)Halogenated and other electron-capturing compounds, at trace levels
Nitrogen-phosphorus (NPD)Compounds containing nitrogen or phosphorus
Flame photometric (FPD)Sulfur and phosphorus compounds
Mass spectrometer (MS)All compounds, with a spectrum for identification

What makes GC results reliable

General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.

New products and price changes, once a month

No daily emails. Unsubscribe in one click.

  • New arrivals
  • Price changes
  • Brand offers

About once a month. Unsubscribe any time. See our privacy policy.

What is gas chromatography? GC explained: columns, inlets, detectors | Lancetech Limited