Definition
Gas chromatography (GC) separates compounds that can be turned into vapour without breaking down. The sample is vaporised in a heated inlet and carried by an inert gas, the mobile phase, through a column that holds the stationary phase. Compounds that are more volatile and interact less with the stationary phase reach the detector first.
GC suits volatile and semi-volatile compounds: solvents, fuels, essential oils, fatty acid methyl esters, pesticides, many environmental pollutants and gases. Compounds that are too polar or not volatile enough can often be analysed after derivatisation; those that cannot be vaporised belong to HPLC.
The parts of a gas chromatograph
- Carrier gas supply. Helium, hydrogen or nitrogen of high purity, with traps to remove moisture, oxygen and hydrocarbons.
- Inlet. Vaporises the sample and transfers it to the column. The split/splitless inlet is the most common. Its liner and septum are consumables.
- Column, inside an oven. Where the separation happens.
- Detector. Produces a signal for each compound that leaves the column.
- Data system. Records the chromatogram.
- Carrier gasHelium, hydrogen or nitrogen, through traps
- InletVaporises the sample in the liner
- Column in the ovenSeparation, controlled by temperature
- DetectorFor example FID or mass spectrometer
- Data system
Injection modes
- Split: only a set fraction of the vaporised sample enters the column; the rest is vented. Used for concentrated samples, giving sharp peaks.
- Splitless: nearly all the sample is transferred. Used for trace analysis.
- On-column and programmed-temperature inlets: gentler techniques for heat-sensitive or high-boiling compounds.
- Headspace: only the vapour above the sample is injected. See headspace analysis in GC.
Columns
Most GC today uses capillary columns: fused silica tubing with the stationary phase coated as a thin film on the inner wall.
| Dimension | Common range | Effect of increasing it |
|---|---|---|
| Length | 15 to 60 m (30 m is typical) | More resolution, longer run |
| Inner diameter | 0.18 to 0.53 mm (0.25 mm is typical) | More sample capacity, less efficiency |
| Film thickness | 0.1 to 5 µm (0.25 µm is typical) | More retention and capacity; suits volatile compounds |
Packed columns, wider tubes filled with coated particles, are still used for gases and some older methods.
Stationary phases
Choose the phase by the polarity of the analytes: like dissolves like.
| Phase | Polarity | Typical use |
|---|---|---|
| 100% dimethylpolysiloxane | Non-polar | Hydrocarbons, solvents; separation roughly by boiling point |
| 5% phenyl, 95% dimethylpolysiloxane | Low polarity | General purpose; semi-volatiles, pesticides |
| 35 to 50% phenyl polysiloxane | Mid polarity | Drugs, pesticides, confirmation |
| Cyanopropylphenyl polysiloxane (6% or 14%) | Mid polarity | Residual solvents, volatile organics |
| Polyethylene glycol (wax) | Polar | Alcohols, flavours, fatty acid methyl esters |
| Porous layer (PLOT) | Adsorbent | Permanent gases, light hydrocarbons |
Browse GC columns.
The oven programme
Temperature is the main control of retention in GC. An isothermal run holds one temperature. A temperature programme starts low and ramps up, so that volatile compounds are separated first and heavier ones are then driven off in reasonable time with sharp peaks. Stay below the column's maximum temperature. Columns usually state two limits: a lower isothermal limit that can be held for long periods, and a higher programmed limit that may be held only for a few minutes at the end of a ramp.
Detectors
| Detector | Responds to |
|---|---|
| Flame ionisation (FID) | Almost all organic compounds; the general-purpose detector |
| Thermal conductivity (TCD) | Everything, including permanent gases; less sensitive |
| Electron capture (ECD) | Halogenated and other electron-capturing compounds, at trace levels |
| Nitrogen-phosphorus (NPD) | Compounds containing nitrogen or phosphorus |
| Flame photometric (FPD) | Sulfur and phosphorus compounds |
| Mass spectrometer (MS) | All compounds, with a spectrum for identification |
What makes GC results reliable
- A leak-free system and clean carrier gas. Oxygen at high temperature destroys the stationary phase.
- A correctly installed and conditioned column.
- A clean, suitable inlet liner and a fresh septum.
- Sound vials, caps and septa, so volatile analytes do not escape.
- Calibration with suitable reference standards.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.