Before you start
- Cool the oven, inlet and detector, and wear clean gloves and eye protection. Fused silica is sharp when broken.
- Check that the carrier gas is of the right purity and that the moisture, oxygen and hydrocarbon traps are in good condition.
- Take the opportunity to replace the inlet liner, O-ring and septum, and inspect the inlet seal (choosing a liner).
- Have the instrument manual to hand: insertion depths differ for each inlet and detector.
Choosing the ferrule
| Ferrule material | Upper temperature | Notes |
|---|---|---|
| Graphite | About 450 °C | Soft, seals easily, reusable; can shed particles; not for mass spectrometer interfaces |
| Graphite/polyimide blend | About 350 °C | Hard, does not flake; standard for mass spectrometer connections; needs retightening after the first heating cycles |
| Polyimide | About 280 °C | Hard; for isothermal use |
Use the ferrule bore that matches the outer diameter of the column.
Installing the column
- Mount the column on the oven hanger so that it does not touch the oven walls.
- Slide the nut and ferrule onto the column first. Then cut a few centimetres off the end. This removes any ferrule fragments pushed into the bore.
- Cut cleanly. Score the polyimide coating lightly with a ceramic wafer or column cutter and snap the end off. Inspect the cut with a magnifier: it must be square, with no jagged edge or shards. A poor cut causes tailing peaks.
- Set the insertion depth for the inlet exactly as the manual states, mark it (for example with a septum slid onto the column), insert and tighten the nut finger-tight plus a part turn.
- Turn on the carrier gas and set the head pressure or flow. Confirm gas is flowing by dipping the free end of the column in a vial of solvent: a steady stream of bubbles should appear.
- Purge the column with carrier gas for 10 to 15 minutes with the oven still at room temperature, before any heating. This flushes out air. Heating a column that still contains oxygen damages the stationary phase permanently.
- Connect the detector end in the same way: nut, ferrule, fresh cut, correct depth.
- Check for leaks at both fittings with an electronic leak detector. Do not use soap solution on capillary fittings: it can be drawn into the system.
Conditioning
- With carrier gas flowing, heat the oven in a programme to the conditioning temperature: about 20 °C above the highest temperature of your method, but never above the column's maximum temperature.
- Hold until the baseline is steady. One to two hours is typical; thick films and polar phases take longer.
- For sensitive detectors, follow the manufacturer's advice on whether to condition with the column connected to the detector.
- Cool down and retighten the fittings: ferrules, especially polyimide blends, relax after their first heating.
Checking the installation
- Inject an unretained compound (methane or butane for a flame ionisation detector) to measure the dead time and set the carrier gas velocity. The peak must be sharp and symmetrical; a tailing unretained peak means a poor cut, wrong insertion depth or a leak.
- Run a test mixture and compare with the column's test chromatogram.
- Record the result, date and column serial number in the instrument log.
Common faults and their causes
| Symptom | Likely cause |
|---|---|
| Tailing on all peaks, including the unretained one | Poor column cut, wrong insertion depth, unswept volume |
| Tailing on polar compounds only | Active sites: dirty liner, contaminated column inlet |
| Rising baseline at high temperature | Normal column bleed if moderate; oxygen leak or contamination if severe |
| Retention times shifting | Leak, or changed head pressure or flow |
| Broad solvent peak, poor early peaks | Wrong initial oven temperature or splitless settings |
Trimming 10 to 50 cm from the inlet end often restores a column whose first section is contaminated.
Storing a column
Remove the column, seal both ends (push each into a piece of septum or use end caps) and store it in its box, away from light and solvents.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.