How to size an SPE cartridge: sorbent mass from the load, bed volume, minimum elution volume, tube size from the sample volume, with a table of typical sizes.

  • 3 min read
  • 1 figure
  • Reviewed 25 November 2026
On this page 9 sections
  1. Two numbers on every cartridge
  2. How much sorbent: capacity
  3. Bed volume and elution volume
  4. Typical loading sizes
  5. How large a tube: sample volume
  6. Breakthrough
  7. Flow rate
  8. Sizing step by step
  9. Other formats

Two numbers on every cartridge

An SPE cartridge is described by its sorbent mass (the packing weight, in mg or g) and its tube volume (in mL), for example 500 mg / 6 mL. They are chosen separately:

  • the sorbent mass from how much material has to be retained;
  • the tube volume from how much liquid has to be handled.
The parts of an SPE cartridge: reservoir, frits, sorbent bed and Luer tipReservoir: the tube volume,for example 6 mLUpper fritSorbent bed: the packing weight,for example 500 mgBed volume about 120 µL per 100 mgLower fritLuer tip: fits the vacuum manifold
An SPE cartridge. The packing weight sets how much can be retained; the tube volume sets how much liquid one filling takes.

How much sorbent: capacity

Capacity is the total mass of compounds a sorbent can hold before they start to pass through. As a working rule:

  • Silica-based sorbents retain up to about 5% of their own mass. A 500 mg bed holds up to roughly 25 mg.
  • Polymeric sorbents have a larger surface area and hold noticeably more per gram.
  • Ion exchangers are rated by exchange capacity, in milliequivalents per gram; use the figure given for the sorbent.

Capacity is shared by everything that binds: the analytes and all matrix compounds retained with them. With a clean sample such as drinking water, almost all of it is available to the analytes. With plasma, a plant extract or waste water, most of it is taken by the matrix, so allow a generous margin.

Bed volume and elution volume

The bed volume is the liquid volume needed to fill the packed bed. For silica-based sorbents it is about 120 µL per 100 mg.

  • Elution needs at least about two bed volumes, and often more, to recover the analyte completely.
  • Conditioning and washing normally use several bed volumes each.

This is why the smallest cartridge that has enough capacity is the best choice: less sorbent means a smaller elution volume, a more concentrated extract and less solvent to evaporate.

Typical loading sizes

Sorbent massTypical tube volumeBed volume (approx.)Capacity at 5% (approx.)Minimum elution volume (approx.)
50 mg1 mL60 µL2.5 mg0.12 mL
100 mg1 mL120 µL5 mg0.25 mL
200 mg3 mL240 µL10 mg0.5 mL
500 mg3 or 6 mL600 µL25 mg1.2 mL
1 g6 mL1.2 mL50 mg2.4 mL
2 g12 or 15 mL2.4 mL100 mg5 mL
5 g20 or 25 mL6 mL250 mg12 mL
10 g60 or 70 mL12 mL500 mg24 mL

These are guide values for silica-based sorbents. The figures for a particular product are given by its manufacturer.

How large a tube: sample volume

The tube above the sorbent is the reservoir. A 500 mg bed is offered in a 3 mL tube and in a 6 mL tube; the 6 mL version simply takes more liquid per filling.

  • For small samples (up to a few millilitres), choose a tube that takes the sample in one or two fillings.
  • For large water samples (100 mL to 1 L), fit a reservoir or a large-volume sampling adapter on top of the cartridge, and choose enough sorbent that the analytes do not break through during the long loading step.

Breakthrough

Loading a very large volume can wash weakly retained analytes through the bed even though the capacity in milligrams is far from used up. The more polar the analyte, the smaller the volume it tolerates. Test it by loading increasing volumes of spiked sample, or by stacking a second cartridge under the first and checking whether it catches any analyte. If there is breakthrough, use more sorbent or a more retentive one (selecting the sorbent).

Flow rate

Larger beds tolerate higher flow, but retention still takes time. As a guide, load at about one drop a second through typical 3 and 6 mL cartridges, and more slowly with ion exchangers, where binding is slower. Elute slowly, letting the solvent soak into the bed.

Sizing step by step

  1. Estimate the total mass of analytes plus retained matrix in the sample.
  2. Choose a sorbent mass whose capacity is comfortably above it.
  3. Check the elution volume that follows, against the concentration you need.
  4. Choose the tube volume, with a reservoir if required, for the sample volume.
  5. Verify by experiment: recovery at the highest expected load, and a breakthrough test at the largest sample volume.

Other formats

For many samples of small volume, 96-well plates with 10 to 100 mg per well apply the same rules on a smaller scale. For dirty extracts where the aim is only to remove matrix, dispersive clean-up tubes are sized by the volume of extract. Hardware such as manifolds, adapters and reservoirs is listed under SPE accessories. The procedure is described in solid phase extraction.

General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.