Two numbers on every cartridge
An SPE cartridge is described by its sorbent mass (the packing weight, in mg or g) and its tube volume (in mL), for example 500 mg / 6 mL. They are chosen separately:
- the sorbent mass from how much material has to be retained;
- the tube volume from how much liquid has to be handled.
How much sorbent: capacity
Capacity is the total mass of compounds a sorbent can hold before they start to pass through. As a working rule:
- Silica-based sorbents retain up to about 5% of their own mass. A 500 mg bed holds up to roughly 25 mg.
- Polymeric sorbents have a larger surface area and hold noticeably more per gram.
- Ion exchangers are rated by exchange capacity, in milliequivalents per gram; use the figure given for the sorbent.
Capacity is shared by everything that binds: the analytes and all matrix compounds retained with them. With a clean sample such as drinking water, almost all of it is available to the analytes. With plasma, a plant extract or waste water, most of it is taken by the matrix, so allow a generous margin.
Bed volume and elution volume
The bed volume is the liquid volume needed to fill the packed bed. For silica-based sorbents it is about 120 µL per 100 mg.
- Elution needs at least about two bed volumes, and often more, to recover the analyte completely.
- Conditioning and washing normally use several bed volumes each.
This is why the smallest cartridge that has enough capacity is the best choice: less sorbent means a smaller elution volume, a more concentrated extract and less solvent to evaporate.
Typical loading sizes
| Sorbent mass | Typical tube volume | Bed volume (approx.) | Capacity at 5% (approx.) | Minimum elution volume (approx.) |
|---|---|---|---|---|
| 50 mg | 1 mL | 60 µL | 2.5 mg | 0.12 mL |
| 100 mg | 1 mL | 120 µL | 5 mg | 0.25 mL |
| 200 mg | 3 mL | 240 µL | 10 mg | 0.5 mL |
| 500 mg | 3 or 6 mL | 600 µL | 25 mg | 1.2 mL |
| 1 g | 6 mL | 1.2 mL | 50 mg | 2.4 mL |
| 2 g | 12 or 15 mL | 2.4 mL | 100 mg | 5 mL |
| 5 g | 20 or 25 mL | 6 mL | 250 mg | 12 mL |
| 10 g | 60 or 70 mL | 12 mL | 500 mg | 24 mL |
These are guide values for silica-based sorbents. The figures for a particular product are given by its manufacturer.
How large a tube: sample volume
The tube above the sorbent is the reservoir. A 500 mg bed is offered in a 3 mL tube and in a 6 mL tube; the 6 mL version simply takes more liquid per filling.
- For small samples (up to a few millilitres), choose a tube that takes the sample in one or two fillings.
- For large water samples (100 mL to 1 L), fit a reservoir or a large-volume sampling adapter on top of the cartridge, and choose enough sorbent that the analytes do not break through during the long loading step.
Breakthrough
Loading a very large volume can wash weakly retained analytes through the bed even though the capacity in milligrams is far from used up. The more polar the analyte, the smaller the volume it tolerates. Test it by loading increasing volumes of spiked sample, or by stacking a second cartridge under the first and checking whether it catches any analyte. If there is breakthrough, use more sorbent or a more retentive one (selecting the sorbent).
Flow rate
Larger beds tolerate higher flow, but retention still takes time. As a guide, load at about one drop a second through typical 3 and 6 mL cartridges, and more slowly with ion exchangers, where binding is slower. Elute slowly, letting the solvent soak into the bed.
Sizing step by step
- Estimate the total mass of analytes plus retained matrix in the sample.
- Choose a sorbent mass whose capacity is comfortably above it.
- Check the elution volume that follows, against the concentration you need.
- Choose the tube volume, with a reservoir if required, for the sample volume.
- Verify by experiment: recovery at the highest expected load, and a breakthrough test at the largest sample volume.
Other formats
For many samples of small volume, 96-well plates with 10 to 100 mg per well apply the same rules on a smaller scale. For dirty extracts where the aim is only to remove matrix, dispersive clean-up tubes are sized by the volume of extract. Hardware such as manifolds, adapters and reservoirs is listed under SPE accessories. The procedure is described in solid phase extraction.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.