What has to be chosen
A TLC plate is a thin, even layer of sorbent on a flat support. Four things define it: the sorbent, any indicator in the layer, the grade (TLC, HPTLC or preparative) and the backing. Browse TLC plates and sheets.
Silica gel in normal-phase TLC
Silica gel 60 (mean pore size 60 Å) is the standard layer and is used for the great majority of separations. It is polar: with a less polar solvent mixture, non-polar compounds travel furthest and polar compounds stay near the start. It suits most organic compounds of low to medium polarity, including pharmaceuticals, natural products, lipids, dyes and reaction mixtures, and is the layer on which solvent systems for column chromatography are worked out.
Silica holds water from the air, and its activity falls as it does. For reproducible results store plates dry, and where the method requires it activate them by heating (commonly around 100 to 120 °C for about half an hour) and let them cool in a desiccator.
Other polar layers:
- Aluminium oxide: basic, neutral or acidic; an alternative where compounds are unstable or too strongly held on silica.
- Cellulose: separates very polar compounds such as amino acids and sugars by partition.
- Polyamide: for phenols, flavonoids and other compounds that form hydrogen bonds.
- Amino, cyano and diol modified silica: intermediate polarity, usable in normal- or reversed-phase mode.
Phases in reversed-phase TLC
In reversed-phase layers the silica is bonded with alkyl chains (RP-2, RP-8, RP-18), making it non-polar. The solvent is polar, typically water with methanol or acetonitrile, and the order of migration is reversed: polar compounds travel furthest.
Use reversed-phase TLC for polar or ionisable compounds that streak on silica, and to scout conditions for reversed-phase HPLC. Fully covered RP-18 layers repel water, so mixtures rich in water may not wet them; choose layers described as water-wettable for such solvents.
Fluorescent indicator
Layers marked F254 or UV254 contain an indicator that glows green under short-wave UV light (254 nm). Compounds that absorb at that wavelength show as dark spots against the bright background, without any reagent and without altering the sample. Some layers also carry an indicator for long-wave UV (366 nm). Choose a layer without indicator if it would interfere with your detection reagent or with recovering the compound.
TLC, HPTLC or preparative
| Grade | Particle size | Layer thickness | Use |
|---|---|---|---|
| TLC | About 10 to 12 µm | About 0.2 to 0.25 mm | Routine qualitative work |
| HPTLC | About 5 to 6 µm, narrow distribution | About 0.1 to 0.2 mm | Sharper spots, shorter runs, better sensitivity; quantitative work with a scanner |
| Preparative | Coarser | 0.5 to 2 mm | Isolating milligram amounts by scraping off the band |
HPTLC layers made from spherical silica particles, instead of the usual irregular ones, give still more compact spots and a lower background, which helps in quantitative and trace work.
The backing
| Backing | Advantages | Limits |
|---|---|---|
| Glass | Rigid, inert, resists heat and aggressive spray reagents, including charring | Heavy, breakable; cannot be cut without a glass cutter |
| Aluminium sheet | Light, cut with scissors, good heat resistance | Attacked by strong acids and bases in solvents or reagents |
| Plastic (polyester) sheet | Flexible, cut with scissors, resists most solvents | Limited heat resistance: not for charring |
Useful extras
- Concentrating zone: a strip of inert material at the bottom of the plate. Samples applied there, even as large or irregular spots, are swept into a sharp line at the start of the separating layer.
- Pre-scored glass plates snap into smaller pieces.
- Layers with binder vary in hardness: harder layers stand up better to handling and writing.
A short guide
| Your sample | Start with |
|---|---|
| General organic compounds, reaction control | Silica gel 60 F254 on aluminium or plastic |
| Detection by charring or aggressive reagents | Silica gel 60 on glass |
| Polar compounds streaking on silica | RP-18, water-wettable |
| Amino acids, sugars | Cellulose |
| Quantitative work, many samples per plate | HPTLC silica gel 60 F254 |
Applying the sample well matters as much as the layer: see sample application in TLC.
General guidance from Lancetech Limited. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.