Lancetech Limited

Sample preparation and filtration

Tips for sample application in thin-layer chromatography

How to apply samples in TLC: preparing the plate, spot size and spacing, spots or bands, manual and spray-on application, then development and calculating Rf.

Why application decides the result

In TLC the spot never becomes narrower than it starts. A sample applied as a small, even zone gives sharp, separated spots; a large or uneven one gives blurred, overlapping ones, whatever the layer and solvent. Good application costs nothing and improves every plate.

Preparing the plate

  1. Handle the plate by its edges, with gloves. Fingerprints show up under UV light and with many reagents.
  2. With a soft pencil, lightly mark the start line 10 to 15 mm from the bottom edge. It must sit above the level of the solvent in the chamber. Do not score the layer.
  3. Mark the application positions at least 10 mm from the side edges, where the solvent front runs unevenly, and about 10 mm apart (about 5 mm on HPTLC plates).
  4. Never use ink: it runs with the solvent.

The sample solution

  • Dissolve the sample in a volatile solvent of low eluting strength for the layer, so the spot does not spread as it is applied. For silica that means a fairly non-polar solvent where solubility allows.
  • Concentrations of roughly 0.1 to 1% are typical for qualitative work.
  • Apply too much and the spot tails or streaks; too little and it cannot be seen. If unsure, apply two or three amounts side by side.

Manual application

Use a glass capillary or a micropipette, typically 0.5 to 5 µL.

  • Touch the tip to the layer lightly, without digging into it. A damaged layer distorts the spot.
  • Keep the spot small: about 2 to 3 mm across on TLC, and about 1 mm on HPTLC.
  • To apply a larger volume, apply it in small portions on the same point, drying between each with a gentle stream of cool air.
  • Let the spots dry completely before development, so that the application solvent does not alter the separation.
  • Apply reference standards on the same plate, beside the samples.

Spots or bands

A sample can be applied as a round spot or as a narrow band a few millimetres long. Bands give better resolution than spots, because the zone is narrow in the direction of travel across its whole width, and they are better suited to quantitative evaluation with a scanner.

Manual versus instrumental application

Manual (capillary or pipette)Instrumental (spray-on)
How it worksThe liquid is transferred by contactThe sample is sprayed onto the layer by a gas stream, without touching it
Zone shapeSpotsBands or spots, of exact size and position
VolumeSmall, limited by spreadingLarger volumes possible, because the solvent evaporates as it is sprayed
Layer damagePossibleNone
RepeatabilityDepends on the operatorHigh
Best forQuick checks, reaction monitoringQuantitative work, HPTLC, large numbers of samples

The spray-on technique builds up the band in fine passes, leaving a very narrow, evenly loaded starting zone. That is the main reason instrumental TLC outperforms manual spotting.

Plates with a concentrating zone offer a middle way: a sample applied by hand anywhere in the zone is focused into a sharp line as the solvent rises (choosing the layer).

Developing the plate

  1. Line the chamber with filter paper, add solvent to a depth below the start line and close the lid. Allow time for the vapour to saturate the chamber, unless the method specifies an unsaturated one.
  2. Stand the plate in the chamber, close it and leave it undisturbed.
  3. Remove the plate when the solvent front has travelled the intended distance (often 10 to 15 cm on TLC, about 5 to 7 cm on HPTLC) and mark the front at once with a pencil.
  4. Dry the plate in a fume cupboard.

Seeing and measuring the spots

Examine the plate under UV light at 254 nm and 366 nm, then use a spray or dip reagent if needed. Circle the spots in pencil.

The Rf value is the distance travelled by the centre of the spot divided by the distance travelled by the solvent front, both measured from the start line. It lies between 0 and 1. Rf values depend on layer activity, chamber saturation, temperature and solvent composition, so identify a compound by comparing it with a reference standard on the same plate, not with a published figure.

Common faults

FaultUsual cause
Streaks or tailingToo much sample; sample too polar or ionised for the system
Large, diffuse spotsApplication solvent too strong; spot too big; spot not dried
Curved or uneven solvent frontPlate touching the paper or chamber wall; chamber not level or not saturated
Spots at the front or at the startSolvent too strong or too weak
Spots dissolve into the solventStart line below the solvent level

General guidance from Lancetech Limited. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.