What it is
Column chromatography is the classical preparative form of liquid chromatography. A glass column is filled with a sorbent, the mixture is placed on top and a solvent (the eluent) is run through. The components travel down at different speeds and leave the column one after another, to be collected as fractions.
It is the standard way to purify reaction products and natural extracts on a scale from milligrams to many grams. When the solvent moves by gravity it is called gravity chromatography; when it is pushed through with gentle gas or pump pressure it is flash chromatography, which is faster and gives sharper bands.
The sorbent
- Silica gel is the usual choice. It is polar and slightly acidic. "Silica gel 60" has a mean pore size of 60 Å. A particle size of 0.063 to 0.200 mm (70 to 230 mesh) suits gravity columns; 0.040 to 0.063 mm (230 to 400 mesh) is the standard flash grade.
- Aluminium oxide (alumina) is supplied as basic, neutral or acidic. It is useful where a compound is unstable on silica. Its activity falls as water is added, described by the Brockmann grades I (most active) to V.
- Reversed-phase silica (C18), polyamide, cellulose and Florisil cover special cases such as very polar compounds, phenols or pesticide clean-up.
On silica and alumina the separation is normal phase: the least polar compound comes off first and polar compounds are held longest. Browse LC sorbents and bulk packings.
Choosing the solvent by TLC
Run the mixture on a TLC plate coated with the same sorbent. Adjust the solvent mixture until the compound you want has an Rf of roughly 0.2 to 0.35 and is clear of its neighbours. That mixture is the starting eluent for the column. Typical systems are hexane or heptane with ethyl acetate, and dichloromethane with methanol for more polar compounds. See choosing the right TLC layer.
How much sorbent
A common working range is 20 to 100 times the weight of the sample: nearer 20 to 30 times for an easy separation, 50 to 100 times when spots sit close together on TLC. A taller, narrower bed resolves better; a shorter, wider bed runs faster.
Packing and loading
- Plug the column outlet with cotton or use a column with a frit, and add a thin layer of sand.
- Wet (slurry) packing: stir the sorbent into the starting solvent, pour the slurry in and let it settle while tapping the column. Dry packing: pour in the dry sorbent, then run solvent through until the bed is evenly wetted.
- Never let the solvent level fall below the top of the bed. A bed that runs dry cracks and the bands smear.
- Load the sample as a narrow band: dissolved in the minimum of solvent, or adsorbed on a little sorbent and added as a dry powder.
- Add a protective layer of sand, then elute. Raise the solvent strength in steps if the later compounds move too slowly.
- Collect fractions and check them by TLC before combining.
Flash cartridges
Pre-packed flash cartridges replace the hand-packed glass column on automated systems. They give reproducible beds and remove the dust and effort of packing. Choose the cartridge size from the sample load in the same way, using the supplier's loading guidance for the sorbent.
Safety
Fine silica dust is a respiratory hazard: weigh and pour it in a fume cupboard. Glass columns under pressure must be rated for it and used behind a shield or with tape or mesh protection.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.