What the eluent controls
The eluent (mobile phase) decides how long each compound is retained and, with the column, the order and spacing of the peaks. Three properties are adjusted: strength (how quickly everything elutes), selectivity (the spacing between peaks) and, for ionisable compounds, pH.
Reversed phase: solvent strength
In reversed phase, water is the weak solvent and the organic solvent is the strong one. More organic means less retention. Elution strength increases in the order:
water < methanol < acetonitrile < tetrahydrofuran
As a rule of thumb, raising the organic content by 10 percentage points cuts the retention factor by a factor of roughly two to three. Aim for retention factors between about 2 and 10 (terms and equations).
Mixtures of similar strength are, very roughly, 50% methanol, 40% acetonitrile and 30% tetrahydrofuran in water. Use these only as a starting point when swapping solvents.
Reversed phase: selectivity
The organic solvents do not differ in strength alone. Methanol can donate and accept hydrogen bonds; acetonitrile interacts through its dipole; tetrahydrofuran is a strong hydrogen-bond acceptor. Swapping one for another often changes the spacing, and sometimes the order, of the peaks. If two peaks will not separate by changing strength, change the solvent before changing the column.
Practical differences: acetonitrile gives lower back pressure and a lower UV cut-off; methanol is cheaper and less toxic but more viscous when mixed with water. See solvents for HPLC.
pH for acids and bases
An ionised compound is more polar and is retained less in reversed phase.
- Acids are neutral, and best retained, at a pH below their pKa.
- Bases are neutral, and best retained, at a pH above their pKa.
Set the pH at least about two units away from the analyte's pKa so that it is essentially all in one form. Close to the pKa a small pH error causes a large shift in retention and the method is not robust. Many methods for basic drugs run at low pH (2 to 3), where residual silanol groups on the silica are not ionised, which improves peak shape. Stay within the pH range of the column and use a buffer suited to the target pH (mobile phase preparation).
Isocratic or gradient
Run a scouting gradient, for example 5 to 95% organic over 20 minutes.
- If all the peaks elute within a narrow part of the gradient (about a quarter of the gradient time or less), an isocratic method is usually possible. The composition at which the peaks elute suggests the starting isocratic mixture.
- If the peaks are spread across the run, keep the gradient and optimise its slope.
Normal phase
On bare silica, cyano, diol or amino phases the order is reversed: a non-polar solvent such as heptane or hexane is weak, and a polar modifier such as 2-propanol, ethanol or ethyl acetate is strong. Water content has a large effect on retention on silica, so control it by using solvents of consistent quality.
HILIC
HILIC uses a polar column with a mobile phase high in acetonitrile, typically 60 to 95%, with the rest water or aqueous buffer. Here water is the strong solvent: more water means less retention. A buffer such as ammonium formate or ammonium acetate is normally needed, and columns take longer to equilibrate than in reversed phase.
Match the sample solvent
Dissolve the sample in the mobile phase or in something weaker. Injecting a sample dissolved in a stronger solvent distorts or splits early peaks, especially on narrow columns and with large injection volumes.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.