Retention
- Retention time (tR): the time from injection to the top of a peak.
- Dead time (t0 or tM): the time an unretained compound takes to pass through the column.
- Elution (retention) volume (VR): retention time multiplied by the flow rate, VR = tR x F.
- Retention factor (k), formerly the capacity factor k': how long a compound spends in the stationary phase relative to the mobile phase.
k = (tR - t0) / t0
A k between about 2 and 10 is a good target. Below 1 the peak sits too close to the unretained material; far above 10 the run is long and the peaks broad.
Selectivity
The selectivity (separation) factor compares two neighbouring peaks, with the later one on top:
alpha = k2 / k1
Alpha is always 1 or more. At exactly 1 the two compounds cannot be separated on that phase with that mobile phase, whatever the efficiency.
Efficiency
The plate number (N) describes how narrow a peak is for its retention time.
- From the width at the base (W): N = 16 x (tR / W)²
- From the width at half height (W½): N = 5.54 x (tR / W½)²
The height equivalent to a theoretical plate is the column length divided by the plate number: HETP (H) = L / N. A smaller H means a more efficient column. The reduced plate height h = H / dp divides out the particle size (dp), so columns of different particle size can be compared; a well-packed column reaches about 2 to 3.
Resolution
Resolution (Rs) measures how well two peaks are separated:
- Rs = 2 x (tR2 - tR1) / (W1 + W2), using widths at the base
- Rs = 1.18 x (tR2 - tR1) / (W½1 + W½2), using widths at half height
Peaks of similar size are separated to the baseline at Rs of 1.5 or more.
The three things that set resolution appear together in one equation:
Rs = (√N / 4) x ((alpha - 1) / alpha) x (k / (1 + k))
Because N sits under a square root, doubling the column length improves resolution by only about 40%. A small gain in selectivity does far more, which is why changing the phase or the mobile phase is usually the best way to pull two peaks apart.
Peak shape
- Tailing (symmetry) factor, the pharmacopoeial measure: T = W0.05 / (2 x f), where W0.05 is the peak width at 5% of its height and f is the distance from the front edge to the peak maximum at that height. A perfectly symmetrical peak has T = 1. The pharmacopoeias call this the symmetry factor and write it As.
- Asymmetry factor, as given on many column certificates: b / a, the back half-width divided by the front half-width at 10% of peak height.
The two are measured at different heights, so their numbers are not interchangeable.
Speed, pressure and the van Deemter equation
- Linear velocity (u): column length divided by dead time, u = L / t0.
- van Deemter equation: H = A + B / u + C x u. The A term is flow-path differences in the packed bed, B is diffusion along the column and C is the slowness of transfer between the phases. The curve has a minimum: the velocity at which the column is most efficient. Smaller particles lower the curve and flatten it at high speed.
- Back pressure rises in proportion to column length, flow velocity and mobile phase viscosity, and with the inverse square of the particle size. Halving the particle size at the same length and velocity quadruples the pressure.
Volumes
- Empty column volume: pi x r² x L. A 150 x 4.6 mm column has a volume of about 2.5 mL.
- Void (dead) volume, V0: the liquid volume inside the packed column, roughly 60 to 70% of the empty volume for fully porous particles: about 1.5 to 1.7 mL for a 150 x 4.6 mm column.
- Extra-column volume: everything between injector and detector that is not the column. It broadens peaks. See preventing dead volume.
- Dwell volume: the volume between the point where the gradient is mixed and the head of the column. It delays the gradient and differs between instruments.
Sensitivity
The limit of detection is commonly taken as a signal-to-noise ratio of 3, and the limit of quantitation as a ratio of 10.
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.