First, make sure it is the column
Replace the column with a union and run the method's flow. If the pressure is still high, the blockage is in the system: an inline filter, a capillary, the autosampler needle seat. If the pressure is normal without the column, remove the guard column and test again. Replace the guard cartridge or inline filter frit before spending time on the analytical column.
| Symptom | Common cause |
|---|---|
| Pressure rising steadily | Particles on the inlet frit; precipitated buffer or sample |
| Tailing or split peaks on all compounds | Blocked frit, or a void at the column inlet |
| Tailing on basic compounds only | Loss of end-capping, or contamination exposing active sites |
| Retention falling over weeks | Loss of bonded phase, usually from low pH |
| Ghost peaks, rising baseline in gradients | Strongly retained material washing off slowly |
Cleaning a reversed-phase column
Disconnect the column from the detector and run the wash to waste. Use about 10 to 20 column volumes of each step, at half the usual flow rate for the viscous solvents.
- Water and organic solvent without buffer (the mobile phase mixture with the salt left out), to remove buffer.
- 100% acetonitrile or methanol, to remove moderately non-polar material.
- 2-Propanol, a stronger solvent that also mixes with the next step.
- Only if needed, for fats, oils and very non-polar residues: dichloromethane or heptane, always with a 2-propanol step before and after, because these do not mix with water.
- 2-Propanol, then acetonitrile or methanol, then back to the mobile phase without buffer, and finally the mobile phase.
Variations:
- Proteins and peptides stuck on the phase respond better to repeated gradients from water to acetonitrile, each with 0.1% trifluoroacetic acid.
- Precipitated buffer needs warm water with a small share of organic solvent, not pure organic solvent.
- Do not leave a conventional C18 column in 100% water for long: the phase can lose contact with the liquid and retention collapses until it is re-wetted with organic solvent.
- Mobile phase without bufferRemoves salts
- Acetonitrile or methanolRemoves moderately non-polar material
- 2-PropanolStronger, and mixes with the next step
- (Dichloromethane or heptane)Only if needed; does not mix with water
- (2-Propanol)Back to a water-miscible solvent
- Acetonitrile or methanolThen mobile phase, then test the column
Cleaning a normal-phase column
Work from non-polar to polar and back: heptane or hexane, then ethyl acetate or dichloromethane, then 2-propanol, and return in the reverse order. Keep water away from bare silica unless the manufacturer's procedure calls for it.
Back-flushing
Reversing the column and flushing to waste pushes particles off the inlet frit by the shortest route. Do this only if the manufacturer permits it for that column, and never into the detector. Some columns, particularly those with very small particles or different frits at each end, must not be reversed.
Other column types
Ion-exchange, size-exclusion, HILIC and polymer-based columns each have their own cleaning solutions, often salt, acid, base or specific organic limits. Use the procedure in the manufacturer's care sheet; a reversed-phase wash can destroy them.
What cleaning cannot fix
- A void at the head of the bed, from pressure shocks or dissolved silica.
- Lost bonded phase after long use at low pH.
- Dissolved silica after use above the pH limit.
- Irreversibly bound material that no permitted solvent removes.
If the test chromatogram is still out of range after a full wash, the column has reached the end of its life.
Afterwards
Equilibrate, run the test mixture and compare plate number, symmetry, retention and pressure with the values recorded when the column was new (testing a column). Then find the cause, otherwise the replacement will go the same way: better sample filtration, a clean-up step or a guard column (column care).
General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.