What pH is

pH expresses how acidic or basic an aqueous solution is. It is the negative logarithm (base 10) of the hydrogen ion activity. Because the scale is logarithmic, a change of one pH unit is a tenfold change in acidity.

In dilute aqueous solutions at 25 °C the scale runs from about 0 (strongly acidic) to 14 (strongly basic), and pure water is neutral at pH 7.00. The neutral point depends on temperature: at 37 °C it is about 6.8.

pH meters

A pH measuring system has three parts.

  • The electrode. Most are combination electrodes containing two half-cells in one body. The glass membrane at the tip develops a voltage that depends on the pH of the sample. The reference electrode provides a constant voltage and makes electrical contact with the sample through a small porous junction, via an electrolyte, usually potassium chloride solution.
  • The meter. It measures the tiny voltage between the two half-cells and converts it to a pH value using the calibration.
  • A temperature sensor, built into the electrode or separate.

The electrode's response follows the Nernst equation. At 25 °C the ideal slope is 59.16 mV per pH unit, and an ideal electrode reads 0 mV at pH 7. Real electrodes deviate a little, and change with age, which is why calibration is necessary.

Choose the electrode for the sample: a general-purpose electrode for clear aqueous solutions; special designs for low-conductivity water, viscous or dirty samples, small volumes, surfaces and samples at extreme pH or temperature. Browse pH meters and electrodes and probes.

The parts of a combination pH electrode: glass membrane, reference electrode, junction and filling holeCap and cable to the meterFilling hole: open while measuringReference electrode(silver / silver chloride)Reference electrolyte(potassium chloride solution)Inner electrode in its own bufferJunction: contact with the sampleGlass membrane: responds to pH.Keep it wet and never rub it
A combination pH electrode: the measuring and the reference electrode in one body.

Buffer solutions

Calibration uses buffer solutions of accurately known pH. Common sets:

Buffer setNominal values at 25 °C
Technical buffers4.01, 7.00, 10.01
Standard (NIST and DIN) buffers1.68, 4.01, 6.86, 9.18, 12.45
  • The pH of a buffer changes with temperature. The values for each temperature are printed on the bottle or its certificate. Alkaline buffers change the most.
  • Use fresh buffer each time, poured into a clean beaker. Never calibrate in the bottle, and never pour used buffer back.
  • Alkaline buffers absorb carbon dioxide from the air and drift downwards. Keep the bottle closed and discard opened pH 10 buffer sooner than the others.
  • Respect the expiry date, and use buffers with a certificate where traceability is required (certificates).

Browse buffers.

Calibrating

  1. Use at least two buffers that bracket the expected pH of the samples: for example 7 and 4 for acidic samples, 7 and 10 for basic ones. A third point improves accuracy over a wide range.
  2. Bring buffers and samples to the same temperature.
  3. Rinse the electrode with deionised water and blot it gently. Do not rub: wiping the glass creates static charge and slows the response.
  4. Immerse the electrode so that the membrane and the junction are covered, stir gently, and accept the reading when it is stable.
  5. Check the calibration result:
ValueWhat it meansUsual acceptable range
SlopeSensitivity of the electrode, as a percentage of the idealAbout 95 to 102%
Offset (zero point)Reading at pH 7Within about plus or minus 30 mV

A low slope or a large offset indicates a dirty, aged or damaged electrode, or bad buffers. Calibrate at least daily, and more often for demanding work.

A pH calibration line: electrode voltage against pH through buffers of pH 4, 7 and 10Voltage, mV-300-150+150+30002471012pHBuffer pH 4.01: about +177 mVBuffer pH 7.00: 0 mV (offset)Buffer pH 10.01: about -178 mVIdeal slope at 25 °C: 59.16 mV for each pH unit (100%)
Calibration fits a line through the buffer readings. Its slope and its offset at pH 7 show the condition of the electrode.

Temperature

Temperature affects pH measurement in three separate ways:

  1. The electrode slope changes with temperature. Automatic temperature compensation corrects this, and only this.
  2. The pH of the buffers changes. Meters with buffer recognition apply the correct value for the temperature.
  3. The pH of the sample itself changes with temperature. This is real chemistry and is not corrected by the meter.

For that reason always report the temperature with the pH value, and compare results only when they were measured at the same temperature.

Measuring

  • Rinse between samples with deionised water, or with a little of the next sample.
  • For refillable electrodes, keep the filling hole open during measurement and the electrolyte level above the sample level, so that electrolyte flows slowly outwards through the junction.
  • Stir gently and consistently, and wait for a stable reading.
  • Very pure water and other samples of low ionic strength give slow, drifting readings: use an electrode made for them, and protect the sample from the air.

For HPLC buffers, measure the pH of the aqueous portion before adding organic solvent (mobile phase preparation).

Storage and cleaning

  • Store the electrode wet, in the storage solution recommended by its manufacturer, normally potassium chloride solution. Never store it in deionised or distilled water, which leaches ions from the glass and the reference, and never let it dry out.
  • If it has dried, soak it for several hours before use.
  • Clean according to the contamination: a mild detergent for grease, a pepsin and hydrochloric acid cleaning solution for protein, a suitable cleaning solution for a blocked or blackened junction. Rinse and recondition in storage solution afterwards.
  • Refill or replace the electrolyte as instructed.
  • Electrodes are consumables. With good care a general-purpose electrode typically lasts one to three years.

Sources of error

SymptomLikely cause
Slow or drifting readingDirty membrane; blocked junction; dry or old electrode; low-conductivity sample
Low slopeAged or contaminated membrane; stale or contaminated buffers
Different result on repeatTemperature differences; carry-over between samples; carbon dioxide absorption
Readings low above about pH 12Alkaline (sodium) error of the glass membrane: use an electrode rated for high pH
Unstable readingAir bubble at the membrane or junction; poor cable or connector

Indicator papers

Where an approximate value is enough, pH indicator papers and test strips give a quick reading without calibration. They are no substitute for a calibrated meter when a method specifies a pH.

General guidance from Lancetech Limited, reviewed 25 November 2026. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.