What is tablet dissolution?
Before a medicine taken as a tablet or capsule can be absorbed, the active substance has to dissolve in the fluids of the gut. Dissolution is that process: the drug passing from the solid dosage form into solution. How fast and how completely it happens depends on the drug's solubility and particle size, on the excipients, and on how the tablet was made.
Dissolution is not the same as disintegration, which is only the break-up of the tablet into smaller particles. A tablet can disintegrate quickly and still release its drug slowly.
Tablet dissolution testing
A dissolution test measures, under standardised conditions, how much of the labelled amount of drug dissolves in a given time. It is used to:
- check batch-to-batch consistency in quality control;
- guide formulation development;
- detect changes during stability studies;
- support the equivalence of products after changes in manufacture, and in some cases in place of studies in people.
The general methods are in Ph. Eur. chapter 2.9.3 and USP chapter 711, which are harmonised. The monograph or the registered specification of the product gives the exact conditions.
The apparatus
| Apparatus | Design | Typical use |
|---|---|---|
| Apparatus 1 (basket) | The dosage form sits in a rotating mesh basket | Capsules; tablets that float |
| Apparatus 2 (paddle) | A paddle stirs the medium; the tablet sinks to the bottom of the vessel | Most tablets; the most widely used |
| Apparatus 3 (reciprocating cylinder) | The dosage form moves up and down through a series of media | Modified-release products |
| Apparatus 4 (flow-through cell) | Medium is pumped through a cell holding the dosage form | Poorly soluble drugs; implants; suppositories |
Standard conditions
- Temperature: 37 °C, held within 0.5 °C.
- Volume: commonly 500, 900 or 1,000 mL per vessel.
- Speed: commonly 50 or 75 rpm for the paddle and 100 rpm for the basket.
- Medium: chosen to reflect conditions in the gut and to dissolve the dose: dilute hydrochloric acid (about pH 1.2), acetate buffer (pH 4.5), phosphate buffer (pH 6.8) or water, sometimes with a surfactant for poorly soluble drugs.
- Deaeration: dissolved air forms bubbles on the tablet and the basket and changes the result, so the medium is deaerated where it affects the product.
Ideally the test runs under sink conditions: the volume of medium is at least about three times that needed to dissolve the whole dose, so that dissolved drug does not slow further release.
Prepare media accurately: a buffer made to the wrong pH can change the result considerably. Calibrate the pH meter with fresh buffers (pH measurement).
Sampling and filtration
At each time point a sample is withdrawn from a defined position in the vessel and filtered immediately. Filtration stops the test for that sample: any undissolved particles carried over would continue to dissolve and inflate the result.
The filter is part of the validated method:
- Choose a membrane and pore size that retain the particles, commonly 0.45 µm, or a coarser filter where the method specifies one.
- Check for adsorption by filtering a standard solution of known concentration and comparing it with the unfiltered solution.
- Discard the first few millilitres of filtrate, as determined in validation.
- Use the same filter type every time.
See the syringe filter selection guide.
Measuring the dissolved drug
The filtered samples are measured by UV spectrophotometry, when the drug absorbs strongly and nothing interferes, or by HPLC, when excipients, several active substances or degradation products make a separation necessary. Results are calculated against a reference standard and expressed as a percentage of the labelled content.
Acceptance criteria
For immediate-release products the specification gives a quantity Q, the percentage of the labelled content that must dissolve within the stated time, for example Q = 80% in 30 or 45 minutes. Testing proceeds in up to three stages:
| Stage | Units tested | Criterion |
|---|---|---|
| S1 | 6 | Each unit is not less than Q + 5% |
| S2 | A further 6 (12 in total) | The average of 12 is not less than Q, and no unit is less than Q - 15% |
| S3 | A further 12 (24 in total) | The average of 24 is not less than Q; not more than 2 units are less than Q - 15%; no unit is less than Q - 25% |
Modified-release and gastro-resistant products have their own criteria, with several time points or an acid stage followed by a buffer stage.
Keeping the equipment in control
Results are sensitive to the geometry and movement of the apparatus. Laboratories verify it regularly by mechanical qualification (vessel and shaft alignment, centring, wobble, rotation speed, temperature and the dimensions of vessels, paddles and baskets) and, where required, by a performance verification test with reference tablets. The documents that come with vessels, paddles and baskets are explained in COC and COA certificates.
Common causes of failing or variable results
- Medium not deaerated, at the wrong pH or the wrong temperature.
- Vibration from nearby equipment.
- Tablets sticking to the vessel wall or floating; a sinker may be specified.
- Worn or out-of-tolerance baskets, paddles or vessels.
- Sampling position or timing not as specified.
- Filter adsorption, or filtration delayed.
General guidance from Lancetech Limited. Always follow your validated method, the instrument manual and the manufacturer's instructions for the product in your hands.